Published in final edited form as: Magn Reson Med. 2016 Mar 26;77(3):1184–1192. doi: 10.1002/mrm.26207

Table 2.

Driving Frequency, Magnitude of the Complex Shear Modulus (|CM|) for DMA and MRE DI estimate, MRE LFE Shear Stiffness, Average OSS-SNR for the Included Volume, Percentage of Pixels with OSS-SNR > 3 for all 8 PVC samples and Percent Difference ((MRE DI – DMA)/DMA) Between MRE DI and DMA Estimates and LFE Shear Stiffness Estimates for Eight PVC Samples

PVC
Sample
Frequency
(Hz)
DMA|CM|
(kPa)
MRE DI|CM|
(kPa)
MRE LFE
(kPa)
Average
OSS-SNR
% pixels with
OSS-SNR > 3
DI %
Difference
LFE %
Difference
50-50 81.96 3.59 3.95 3.89 22.65 100 10.03 8.36
60-40 105.09 5.92 6.56 6.47 18.22 100 10.82 9.29
70-30 130.89 9.13 9.60 9.42 16.93 100 5.15 3.18
75-25 139.17 10.37 11.64 11.57 8.41 99 12.25 11.57
80-20 147.06 11.51 11.20 11.11 26.73 100 −2.69 −3.47
85-15 176.29 16.47 18.05 17.95 6.43 93 9.60 8.98
90-10 175.76 16.58 18.53 18.52 4.73 80 12.44 11.71
95-05 203.24 22.05 22.74 22.44 4.03 70 3.13 1.77

Data are for effective wavelength (λeff) = 6.6 with a corresponding pixels per wavelength of 9.24.