Table 1.
Optimal MR parameters derived under the small variation assumption.
| Optimal TR | Optimal TE | Optimal θ | |
|---|---|---|---|
| T1 contrast efficiencya | TR < T1 | TE ≪T2* | θ = θT1 > θErnst |
| T2* contrast efficiencya | TR < T1 | TE = T2* | θ = θErnst |
| IONP contrast efficiency | TR < T10 | TE ≪ T20* | θ = θT1 > θErnst |
| Positive contrast regime (CNRT1) | |||
| IONP contrast efficiency | TR < T10b | TE = TET2* | θ = θT2* < θErnst |
| Negative contrast regime (CNRTT2*) | |||
| IONP contrast efficiency | TR < T10 | TE = TET1–T2* | θ = θT1–T2* |
| Synergistic T1-T2* (CNRT1–T2*) | θT2* < θT1–T2* < θT1 | ||
TR < T1 means that the optimum is essentially flat, as opposed to TE ≪ T2* here.
T1 and T2* variations treated separately. See (37) for an analytical expression of θT1.
When β < 1, a low flip angle does not reduce T1 effects enough at short TR, and a longer TR (> T1) may slightly improve CNRT2*. However in such cases T1 contrast is the dominant mechanism and thus is a better strategy than T2* contrast.