TABLE 1.

Troubleshooting table.

Problem Possible reason Solution
Preparation drifting Explant is not properly pinned into the dish Pin the explant stably and flat into the dish; make sure that no parts touch the objective
Superfusion or suction lines touch the rim of the dish Place the superfusion lines in a way that their tips are in the fluid, but do not touch the dish
Temperature has not yet stabilized Wait until the temperature is stable; avoid temperature spikes; turn off auto-feedback control
Preparation twitching Muscle fibers have been damaged during the preparation Discard preparation
Tissue is suffering from phototoxicity Reduce excitation light
Blurry images Air bubble is trapped in front of the objective Flush with physiological solution, taking care to avoid a temperature spike due to increased flow rate during flush. Have oxygenated room temperature solution ready for ‘rescue operations’
Consider heating solution in reservoir to avoid ‘degassing’ as temperature rises in in-line heater
Superfusion or suction lines touch the rim of the dish and transmit vibrations—or air table is not floating Place the superfusion lines in a way that their tips are in the fluid but do not touch the dish. Check that all four corners of table are floating
Absence of cellular dynamics Tissue has been mechanically damaged, suffered hypoxia or has been in chamber for too long Discard the preparation
Organelles look abnormal (e.g., rounded mitochondria) Tissue has been damaged or suffered hypoxia Discard the preparation
Slow warping of imaged synapse ‘Contraction clots’ in muscle fibers are forming as tissue has been mechanically damaged, suffered hypoxia or was exposed to phototoxicity Discard the preparation. Reduce light exposure for new sample
Bleaching and phototoxicity Light levels too high or exposure time too long Reduce excitation light to lowest possible level, especially during search for suitable axons (dark-adapt your eyes before starting the experiment)
Undesired wavelengths (UV, infrared) present in excitation light Check the presence of heat filters. Consider inserting extra UV-filters