Table 1.

Suggested sequences for DWIBS at 1.5 T (++ very good, + good, +/− moderate, − poor, −− very poor; FOV field of view, RFOV rectangular FOV, NA not applicable, TE echo time, TI inversion time, TR repetition time)

Name DWIBS-STIR DWIBS-CHESS
Main characteristics Most suitable for organs and body regions suffering from large magnetic field inhomogeneities Most suitable for organs and body regions without significant magnetic field inhomogeneities
Useful in suppressing bowel signal Relatively higher SNR than DWIBS-STIR
Relatively lower SNR than DWIBS-CHESS Relatively less time-consuming than DWIBS-STIR
Relatively more time-consuming than DWIBS-CHESS
Suitability
Neck and shoulder ++ −
Chest ++ +
Abdomen and pelvis
Liver, pancreas, spleen + ++
Colon ++ +
Kidneys ++ +
Uterus, bladder + ++
Prostate + +
Extremities ++ +
Imaging parameters
Fat suppression STIR CHESS
b-value (s/mm2) 1,000a, b 1,000a, b
Direction of MPGs Phase, frequency, and slice Phase, frequency, and slice
TR (ms) 10,205 7,800
TE (ms) 70 70
TI (ms) 180 NA
Parallel imaging factor 2 2
Halfscan factor 0.6 0.6
EPI factor 47 47
Slice orientation Axial Axial
No. of slices 60 60
Slice thickness/gap (mm) 4/0 4/0
FOV (mm) 400 400
RFOV (%) 70 70
Matrix 160 160
Scan percentage 70% 70%
Actual voxel size (mm3) 2.5 × 3.6 × 4.0 2.5 × 3.6 × 4.0
Calculated voxel size (mm3) 1.6 × 1.6 × 4.0 1.6 × 1.6 × 4.0
No. of signals averaged 10 6
Scan time 7 min 8 s 3 min 2 s

aFor imaging of the upper abdomen (including liver, spleen, and pancreas), a b-value of 500 s/mm2 can be applied, if necessary combined with respiratory triggered scanning

bFor tumor detection in the prostate, b-values in the order of 1,500–2,000 s/mm2 can be applied