Skip to main content
. Author manuscript; available in PMC: 2009 Jul 1.
Published in final edited form as: J Struct Biol. 2008 Apr 14;163(1):45–52. doi: 10.1016/j.jsb.2008.04.001

Table 1.

Examples of cryo-EM maps of various specimens studied at different electron accelerating voltages achieved at subnanometer resolutions expressed in terms of a fraction of Nyquist frequency using the Gatan 4k × 4k CCD camera.

Specimen Voltage (kV) Effective Magnification Å/pixel Map Resolution (Å) Resolution as a fraction of Nyquist frequency
Adenovirus (Saban et al., 2005) 300 253,654 × 0.59 8.9 0.13
Adenovirus (Saban et al., 2006) 300 ≥253,654 × 0.59 6.9 0.17
Cytoplasmic polyhedrosis virus (Booth et al., 2004) 200 ~83,100 × 1.81 9.0 0.40
Flagellar filament (Yonekura et al., 2006) 300 ~110,00 × 1.35 8.0 0.34
GroEL (Chiu et al., 2006) 200 ~110,800 × 1.34 9.0 0.30
GroEL (Stagg et al., 2006) 120 66,400 × 2.26 7.8 0.58
Pyruvate dehydrogenase complex (Yu et al., 2008) 200 69,220 × 2.17 8.8 0.49
Ryanodine receptor (Ludtke et al., 2005) 200 ~83,100 × 1.81 9.6 0.38