Dose-Escalation Study of Amygdalar Transcranial Focused Ultrasound in Healthy Volunteers

Norman M Spivak 1,2,3,‡, Andrew AED Bishay 1, Jonathan Haroon 1,4, Amber R Hopkins 5, Jody Tanabe 6, Sabrina Halavi 1, Bianca Hoang-Dang 1,7, Andrew J Swenson 1,8, Samantha F Schafer 9, Mark E Schafer 9,10, Alexander Bystritsky 1, Susan Y Bookheimer 1, Martin M Monti 1,2,5,*
PMCID: PMC13693848  NIHMSID: NIHMS2214545  PMID: 42171549

Abstract

Background:

Transcranial focused ultrasound (tFUS) is a promising noninvasive technique for modulating deep brain structures, but the optimal and safe intensity range for neuromodulation remains unclear. Current FDA intensity limits, designed for diagnostic use rather than therapy, may limit the potential effectiveness of tFUS.

Objective:

This study aimed to evaluate the safety and preliminary efficacy of escalating tFUS intensities targeting the right amygdala in healthy volunteers, including intensity levels exceeding current FDA diagnostic ultrasound limits.

Materials and Methods:

Ten healthy adults (mean age = 19.5 ± 1.4 years; 70% female) participated in a within-subject, repeated-measures design. Each received four randomized tFUS stimulation conditions (ISPTA.3 = 0.72–10.08 W/cm2) using the BX Pulsar 1002 system. Structural MRI was performed before each stimulation to monitor for safety. Affective modulation was assessed pre- and post-stimulation via the International Affective Picture Set (IAPS) task, measuring changes in valence and arousal ratings. Acoustic modeling was conducted to estimate intracranial energy deposition.

Results:

No adverse events or MRI-detectable structural changes (e.g., edema, hemorrhage, or microstructural abnormalities) occurred under any stimulation condition. Behavioral analyses revealed no significant overall effects of intensity in response to affective stimuli. Comparisons between high and low duty cycle paradigms showed no significant differences.

Conclusions:

tFUS of the right amygdala at intensities up to 10.08 W/cm2 derated spatial-peak temporal-average intensity (ISPTA.3), an FDA-recognized metric used to estimate ultrasound intensity at the target, substantially exceeding the FDA diagnostic limit, was well tolerated and produced no structural or subjective adverse effects. These findings support the safety of higher-intensity tFUS and justify further investigation in larger and clinical populations to explore potential dose–response effects on emotion regulation.

Introduction

Determining the optimal intensity of transcranial focused ultrasound (tFUS) needed to effectuate change in neural activity is a critical step in moving towards using tFUS as a safe and effective neuromodulatory tool. While prior studies have established the feasibility of modulating deep brain structures such as the amygdala, thalamus, and entorhinal cortex, these all operated within the Food and Drug Administration’s (FDA) output limit for diagnostic ultrasound [1–4].

In 1985, the United States FDA established exposure limits for diagnostic ultrasound established application-specific intensity limits for diagnostic ultrasound based on the highest output intensities being currently used by devices already in clinical use when the FDA gained regulatory control. These limits were expressed as derated spatial-peak temporal-average intensity (ISPTA.3), where derating applies an assumed soft-tissue attenuation of 0.3 dB/cm/MHz to better estimate in situ exposure. The FDA limits were set as follows: peripheral vascular, 720 mW/cm2; cardiac, 430 mW/cm2; fetal and other, 94 mW/cm2; and ophthalmic, 17 mW/cm2 (all are derated intensity spatial peak temporal average (Ispta.3)). Derated acoustic metrics refer to ultrasound output parameters that have been adjusted using a standardized attenuation coefficient (0.3 dB/cm/MHz) to approximate in situ exposure within tissue, rather than free-field measurements obtained in water. In this study, exposure was quantified using FDA-recognized derated intensity metrics, including the spatial-peak temporal-average intensity (Ispta.3), which reflects time-averaged energy deposition and is primarily associated with thermal effects, and the spatial-peak pulse-average intensity (Isppa.3), which reflects peak energy during pulse delivery and is more relevant to mechanical bioeffects. In cooperation with the ultrasonic community, these values were revised in 1992, establishing a new limit of Ispta.3 = 720 mW/cm2 for all applications except those involving ophthalmic applications. The vast majority of tFUS work uses the FDA “Track 3” requirements, which require the following: Ispta.3 must not exceed 720 mW/cm2, and either the Mechanical Index (MI) must be below 1.9 or the intensity spatial peak pulse average (Isppa.3) must be below 190 W/cm2. The Mechanical Index (MI) is a dimensionless parameter defined as the ratio of the derated peak rarefactional pressure to the square root of frequency (MI = Pr.3/√f), and serves as an indicator of the likelihood of cavitation-related mechanical effects. However, these limits apply to diagnostic ultrasound, not therapeutic ultrasound. Nevertheless, many researchers abide by the FDA limits for ease of gaining approval from oversight bodies (such as Institutional Review Boards (IRB)) to conduct their research. Furthermore, many publications over the last two decades have underscored both the safety record of tFUS [5–8] and its ability to non-invasively exert neuromodulatory effects[9, 10]. It is important to note that these guidelines do not imply that these are the safety limits, above which stimulation would be unsafe and that stimulation below these intensities is automatically safe.

While the FDA limit has allowed for ease in the obtaining of regulatory approval from an IRB to proceed with these studies, neuromodulation requires different considerations, as ultrasound delivered at intensities compliant with the limit may evoke non-uniform responses attributable to the stochastic nature of neuronal firing. Accordingly, variability at FDA-compliant outputs should not be taken to imply a lack of efficacy; rather, it suggests that achieving reliable neuromodulatory engagement may require operation within a parameter space not fully encompassed by diagnostic ultrasound limits. In fact, multiple studies have shown the safety of intensities exceeding the FDA limits[11–14]. Our group conducted an open-label study of ultrasound administration to the epileptogenic focus in pre-surgical temporal lobe epilepsy patients who subsequently underwent resection of the epileptogenic zone to treat their refractory epilepsy. In that work, we did not see histological changes in brain tissue when sonicating up to Ispta.3 of 5.76 W/cm2 (eight times above the FDA limit)[13]. However, since in-vivo sonication requires transmission through the skin and skull, which inherently causes attenuation and some loss of energy[15], we aimed to first understand the minimum energy required to cause damage in tissue directly. An ex vivo brain tissue sample study was designed to determine safety thresholds for ultrasonic brain stimulation. In that work, pieces of brain tissue from recently deceased cadavers and surgical resections were directly sonicated in a petri dish at varying intensity levels to determine the threshold at which damage starts to appear on histology[11]. No signs of damage were observed in any samples that underwent sonication at or below 14 W/cm2 Ispta.0, while inconsistent damage above this intensity included changes such as extravasation of blood vessels and spongiosis. Again, there was no skull to provide for any attenuation of ultrasonic energy. While the aforementioned study involved direct ex-vivo sonication of brain tissue, studies in living humans involve transcranial administration through skin, skull, and hair. It was therefore important to understand if the increased energy deposition associated with higher intensity stimulation (and subsequent energy absorption by the skull) would result in appreciable heating of the skin and skull. A small study in healthy volunteers involved sonication of the right scapula, administered from the dorsal surface[16]. With thermoprobes placed between the ultrasound transducer and the skin, the temperature was measured during stimulation with five different intensity levels up to 14 W/cm2 Ispta.3. Even at the maximum stimulation intensity, only 1.5° C of heating were seen, which is a reasonable rise in temperature that is not expected to produce adverse events and which is below the FDA cutoff of 2° C. These findings thus suggest the presence of a safe window of increased energy deposition well above the current FDA energy limit, implying a region of ultrasound parametrization capable of increased neuromodulation and, putatively, increased therapeutic effectiveness, without a concomitant increase in risk profile.

The possibility of employing intensities above the FDA limit is particularly important, since the skull substantially attenuates ultrasound[17], which may greatly limit the effect sizes induced by ultrasonic neuromodulation. As such, at intensities compliant with the FDA limit, there is limited energy delivered to the tissue itself, which limits the neuromodulatory effect. This raises a central question: does an intensity range above the FDA limit permit stronger neuromodulatory effects without compromising safety, and is there a dose–response such that increasing intensity yields more reliable or larger changes in brain activity?

Of course, there needs to be safe, systematic, and incremental explorations of these FDA limit-exceeding intensities. Higher intensities carry the risk of ablative tissue damage or cavitation effects[18]. Perhaps most notable for the field was the likely-cavitation event that occurred at West Virginia University Rockefeller Neuroscience Institute (WVU/RNI), in which a subject enrolled in a study using tFUS for neuromodulation in severe substance use disorder was rendered comatose during treatment[19]. Notably, this study was a “significant-risk” designated study, carried out at an FDA-limit exceeding intensity and mechanical index (MI)[20]. While report of this event was released after our study and analysis had concluded, it nonetheless points to the importance of safely exploring new parameter spaces in controlled, non-significant risk studies.

Consequently, the present study helps to fill this gap through a carefully designed, dose-escalation study, conducted in healthy volunteers. Specifically, using four carefully chosen stimulation paradigms, we tested whether tFUS of the right amygdala at intensities above the FDA limit had any differential effect on a behavioral task previously validated with amygdalar tFUS[1]. Throughout this manuscript, FDA diagnostic limits are referenced in relation to free-field derated transducer output (ISPTA.3), consistent with regulatory definitions, whereas intracranial intensity estimates are provided solely for contextual interpretation and are not used as safety thresholds.

It is known that the amygdala plays a role in various emotional and behavioral responses, and that amygdalar circuits are crucial for anxiety and fear processing and play an essential role in regulating emotions[21, 22]. Given the significant role of the amygdala in emotional processing and its links to psychiatric disorders, investigating clinical interventions aimed at modulating amygdalar activity is key to making forward therapeutic progress. We also acquired clinical MR sequences to monitor for changes indicative of injury caused by tFUS[23]. We hypothesized that higher stimulation intensities would lead to more significant emotional shifts as assessed by changes in the behavioral tasks, while ensuring that the increased intensities would not result in tissue injury as assessed by structural MRI.

Materials and Methods

Participants

Ten healthy volunteers (age (SD): 19.5 (1.4) years; 70% female) were recruited and screened for eligibility. Inclusion required normal or corrected-to-normal vision and hearing. Participants with any DSM-5 diagnosis, current use of psychiatric medications, MRI contraindications (such as implants), or pregnancy were excluded. The study was reviewed by the UCLA Medical IRB # 3 (IRB-23-0612). All subjects signed a written informed consent form prior to participating in any study activities. Subjects were compensated $50 per study visit with an additional $50 completion bonus, for a total compensation of $300 for completing all five study visits.

Study Design

This was a within-subjects, repeated-measures design in which each subject received four different tFUS stimulation parametrizations—randomized and counterbalanced across visits—to the right amygdala. Each stimulation visit also required subjects to complete a behavioral task twice, once before and once after stimulation, as well as undergo a structural MR acquisition, which was always performed before tFUS administration. There was a fifth follow-up visit with imaging only. Adverse effects were assessed using a prospective structured verbal query administered by study staff, in which participants were systematically asked about the presence or absence of each predefined symptom reported by Legon et al., rather than relying on spontaneous self-report or a validated side-effect instrument [24]. All study visits were spaced one week apart. The tFUS parameters used are shown in Table 1.

Table 1: tFUS Stimulation Parameters.

All parameter sets used the device fundamental 650 kHz carrier frequency and a 10 minute sonication duration, using a 30s on / 30s off pulse paradigm.

Parameter Set ISPTA.3 (W/cm2) ISPPA.3 (W/cm2) Pr.3 (MPa) Mechanical Index Duty Cycle (%) Pulse Repetition Frequency (Hz) Pulse Duration (ms)
A 0.72 1.03 0.177 0.220 70 1000 0.7
B 5.76 8.23 0.477 0.592 70 1000 0.7
C 10.08 14.4 0.635 0.788 70 1000 0.7
D 10.08 201.6 2.26 2.803 4.97 71 0.7

Parameters were chosen as follows: In all paradigms, the pulse length was held constant at 0.7 ms. Parameter Set A uses the FDA output limit for the derated spatial-peak temporal-average intensity, with a 70% duty cycle [12, 25]. Parameter Set A was then modified to have a derated spatial-peak temporal-average intensity that matches the maximum intensity used in a safety study by Stern and colleagues to yield Parameter Set B[13]. Parameter Set C was a modification of Parameter Set B with a derated spatial-peak pulse-average intensity of 14.4 W/cm2 that matches that of Cain and colleagues[26]. Lastly, Parameter Set D was made by modifying Parameter Set C to maintain the same derated spatial-peak temporal-average intensity but with a lower duty cycle at 5%, which Dell’Italia suggests as having different directionality of effects than the higher 70% duty cycle. In this case, Parameter Set 4 was at 4.97% because of the need to maintain a 0.7 ms pulse length. (N.B., We do not believe that a difference of 0.03% is biologically meaningful.)

Focused Ultrasound Device

The BX Pulsar 1002 Focused Ultrasound Device (BrainSonix Corp, Sherman Oaks, CA) was used to deliver the stimulation in this study[27, 28]. The focused ultrasound device operates at a fundamental frequency of 650 kHz and uses transducers with a 61.5 mm aperture diameter, with varying radii of curvature to allow for different focal depths of stimulation. In this study, we used either the 55 or 65mm focal depth transducer, depending on individual neuroanatomy, which varied per subject. The transducer was coupled to the subject’s temporal window using the device manufacturer’s transmit acoustic standoff pads and sufficient ultrasound gel (Aquasonic 100, Parker Laboratories) to wet the hair and ensure no air bubbles would be present that could potentially attenuate ultrasound.

Imaging and Neuronavigation

All MRI scans were acquired on a Siemens 3T Prisma scanner using a 32 channel coil at UCLA’s Staglin Center for Cognitive Neuroscience. Each session included a high-resolution T1-weighted MPRAGE scan for anatomical localization (TR/TE = 2000/2.52 ms, TI = 1100 ms, flip angle = 12°, voxel size = 1 mm3 isotropic, GRAPPA = 2, 192 slices), a T2-weighted turbo spin echo (TR/TE = 5000/104 ms, flip angle = 120°, voxel size = 0.5 × 0.5 × 4.0 mm3, GRAPPA = 2, 25 slices), and a 3D susceptibility-weighted imaging (SWI) scan sensitive to venous vasculature and microhemorrhages (TR/TE = 27/20 ms, flip angle = 15°, voxel size = 0.9 × 0.9 × 1.5 mm3, GRAPPA = 2, 80 slices). Diffusion imaging was acquired with a RESOLVE sequence for distortion-reduced diffusion imaging (TR = 4050 ms, TE1/TE2 = 63/111 ms, flip angle = 180°, voxel size = 0.5 × 0.5 × 4.0 mm3, 4 directions, 27 slices). Apparent diffusion coefficient (ADC) maps were then derived from the diffusion imaging and ultimately used for the clinical review, rather than the raw diffusion images. All four sequences were repeated at each session to monitor for structural changes potentially resulting from the prior week’s ultrasound stimulation. An expert neuroradiologist reviewed the first and final session scans, with the intermediate scans from sessions two, three, and four scheduled for review only if changes were observed.

Additionally, the T1-weighted image from the first session was then subsequently used at every session for neuronavigation with BrainSight (Rogue Research, Montreal, QC), which allowed for positioning of the transducer aimed at the right amygdala without the need for real-time MR guidance.

Acoustic Modeling

Acoustic modeling was performed using BabelBrain to confirm targeting accuracy and to provide contextual estimates of intracranial exposure [29]. Importantly, simulation-derived metrics were not intended as safety thresholds; safety assessment was based on empirical neuroradiological evaluation of multimodal MRI, consistent with clinical focused ultrasound practice. The modeling software allows for the simulation of acoustic effects achieved with an ultrasound transducer trajectory (location and orientation of the transducer toward a target region) for an individual brain, considering the type of transducer, the distance from the transducer surface to the scalp, the sonication parameters, and distortion effects caused by the skull (estimated from the individual T1-weighted data). The 65-mm fixed-focal-length transducer used in this work is natively supported by the software was used in all modeling. We selected one participant per parameter set for modeling to illustrate how the simulated acoustic effects differed between the parameter sets and the participants, considering differences in individual head characteristics and transducer trajectory. Simulated acoustic effects achieved with each parameter set are presented in Fig. 1 and Table 2. Transducer trajectory defined, estimated skull and tissue characteristics amongst participants, and estimated thermal effects of each parameter set from the modeling are presented in Supplementary Fig. 1. As modeling can be perfomed using solely T1-weighted imaging, and our endpoint is safety changes as assessed by clinician review, we did not acquire CT images, nor did we acquire ultrashort echo MR images to generate pseudoCT. A comparison between BabelBrain results and acoustic effects estimated using another software, SMART_FUS [30], in an example participant is provided in Supplementary Figure 2, given that different acoustic modeling approaches can yield very different results [31].

Figure 1: Simulated acoustic effects of sonication with each parameter set.

Figure 1:

One participant was selected for modeling each parameter set. Green crosshairs intersect at the single voxel coordinate defined in the neuronavigational system that tFUS was aimed at for each participant. A target region (3x3x3 mm white cube) was generated around this single coordinate to better compare the simulated acoustic effects against the target region. The focus of the simulated ultrasound beam is colored by the estimated acoustic pressure achieved, ranging least (dark blue) to greatest (dark red) pressure achieved for the individual brain.

Table 2:

Estimated acoustic effects

Set Sub Sess Isppa.0 (W/cm2) Isppa brain max (W/cm2) Isppa target (W/cm2) Ispta.0 (W/cm2) Ispta brain max (W/cm2) Ispta target (W/cm2) MI ΔT (°C)
A 5 1 1.33 0.21 0.12 0.93 0.15 0.09 0.10 0.03
B 11 2 10.63 1.40 0.47 7.44 0.98 0.33 0.26 0.26
C 3 1 18.60 3.34 0.88 13.02 2.34 0.61 0.41 0.53
D 8 1 261.97 38.85 12.98 13.02 1.94 0.65 1.39 0.42

Isppa.0 and Ispta.0 are the non-derated values output by the drive system. “Isppa target” and “Ispta target” are the values at the target coordinate (a single voxel-space coordinate) defined in the individual T1-weighted image using the neuronavigation system for tFUS aiming. “Isppa max brain” and “Ispta max brain” are the maximum values achieved within the focal point of the simulated ultrasound beam. “MI” and “ΔT” are the maximum MI and temperature change from baseline in the brain.

Abbreviations: MI, mechanical index; req, required; Sess, session.

*

We report the Isppa and Isppa achieved at the target voxel, which is the single coordinate in voxel that was defined and aimed at during tFUS administration.

Behavioral Task

Before and after each stimulation, subjects completed a computer task using the International Affective Picture Set (IAPS)[1, 32]. Here, the subjects rated the valence and arousal of emotionally salient images, and changes in arousal and valence were used as a proxy for modulation of amygdalar-driven affective activity. There were eight different forms of the task, for which the images were randomized and counterbalanced among the sessions, and the forms themselves were designed to have balanced normative means for arousal and valence. Furthermore, each form of the task had balanced normative means for valence of three subcategories of images: positive, neutral, and negative[33, 34].

To assess stimulation effects on affective responses, we calculated within-subject pre-/post-task differences in positive and negative valence ratings for each stimulation parameter set. Additionally, a directional valence index (DVI) was computed for each condition as the change in positive valence minus the absolute change in negative valence (ΔPositive – |ΔNegative|). Likewise, a directional arousal Index (DAI) was computed for each condition as the change in positive arousal minus the absolute change in negative arousal (ΔPositive – |ΔNegative|). Normality of the arousal and valence difference distributions, DAI, and DVI distribution was assessed via the Shapiro–Wilk test. To assess differences in the arousal and valence differences, DAI, and DVI across stimulation paradigms, we used a repeated measures ANOVA with post-hoc pairwise comparisons with Holm correction for multiple testing as needed. For deviations from normality, a Wilcoxon signed-rank test or a Friedman test was used as appropriate. As Parameter Set D is low duty cycle, arousal and valence changes, DAI, and DVI were analyzed against Parameter Set C, as they have equivalent Ispta.3 levels to see if there is a directional effect of duty cycle using a paired t-test. Arousal and valence differences, DAI, and DVI from parameter Sets A-C were analyzed together using a polynomial contrast to assess for dose-response. One subject completed one of the same IAPS forms twice inadvertently; the second administration is not included in the analysis.

Results

Safety

Subjects were prospectively queried at each session for any discomfort or pain due to the stimulation and were asked about whether they had any of the adverse events reported by Legon and colleagues since their last stimulation session [24]. Across all 10 subjects and stimulation paradigms, there were no reports of any subjective distress or noticeable changes in cognition, emotional regulation, pain, or physical comfort[24]. In addition to a lack of subjective safety findings, subjects showed no objective safety concerns either. They did not report any thermal changes at the stimulation site.

Across all study sessions, structural MRI demonstrated no significant interval changes between the first and last sessions. As no such changes were detected, only the initial and final scans were evaluated. In several cases, one ADC map per session showed an artifact that is typically associated with multiband acquisition. Incidental findings are listed in the supplementary materials (Supplement 5).

Modeling

Acoustic modeling indicates that the right amygdala was reached with tFUS amongst the participants (see Fig. 1 and Table 2). Distortion effects from the skull and other tissue types in the head shifted the focus of the ultrasound beam from the target region (usually towards the skull), however, this was only by a few millimeters in any direction. Additional modeling results are presented in Supplementary Table 1, Supplementary Fig. 1, and Supplementary Fig. 2.

Behavioral Outcomes

The IAPS behavioral task was used to measure changes in emotion as a result of stimulation. The eight forms of the task that were created were randomized and counterbalanced across sessions. A form of the task was administered before and after each stimulation session.

Change in Directional Valence Index

While there was a general dose-dependent decrease in mean DVI across stimulation paradigms, the Friedman test revealed no significant overall effect of the stimulation paradigm on the DVI (χ2(2) = 2.00, p = 0.368), with a small effect size indicated by Kendall’s W = 0.111. Post-hoc Conover tests with Holm correction showed no significant pairwise differences between any stimulus pairs (all p > 0.50). The changes in DVI are plotted in Figure 2a.

Figure 2: Behavioral Task Changes by Parameter Set.

Figure 2:

(a) Changes in Directional Valence Index (DVI) by stimulation paradigm. (b) Changes in Directional Arousal Index (DAI) by stimulation paradigm. (c) Changes in positive valence by stimulation paradigm. (d) Changes in negative valence by stimulation paradigm. (e) Changes in negative arousal by stimulation paradigm. (f) Changes in positive arousal by stimulation paradigm.

Change in Directional Arousal Index

The Friedman test revealed no significant overall effect of the stimulation paradigm on the DAI (χ2(2) = 0.941, p = 0.625), with a very small effect size indicated by Kendall’s W = 0.052. Post-hoc Conover tests with Holm correction showed no significant pairwise differences between any stimulus pairs (all p > 0.50). The changes in DAI are plotted in Figure 2b.

Changes in Positive Valence

A repeated-measures ANOVA found no significant differences in positive valence changes across stimulation paradigms, F(2,16) = 0.585, p = 0.569, with a small effect size of η2p= 0.068. Post-hoc comparisons with Holm correction showed no significant differences between stimulus pairs (all p = 1.000), although the comparison between paradigm A and C revealed a moderate effect size (Cohen’s d = 0.53) despite being non-significant. Polynomial contrast did not show a significant linear (p= 0.352) or quadratic (p= 0.809) effect. The changes in positive valence by stimulation paradigm are plotted in Figure 2c.

Changes in Negative Valence

A repeated-measures ANOVA found no significant differences in negative valence differences across stimulation paradigms, F(2,16) = 2.04, p = 0.163, η2p = 0.203. Polynomial contrast showed a significant linear effect was observed across paradigms A to C (p = 0.025); however, this effect was not supported by significant omnibus or pairwise comparisons. The post-hoc Conover test with Holm correction between paradigm A and paradigm C was not significant (p = 0.075) and yielded a large effect size (d = 0.97). The changes in negative valence by stimulation paradigm are plotted in Figure 2d.

Changes in Positive Arousal

The repeated-measures ANOVA did not show a significant effect of stimulation paradigm on the change in positive arousal (F(2,16)=0.16, p=0.858) with a small effect size of ηp2= 0.019. Polynomial contrast did not show a significant linear (p=0.838) or quadratic (p=0.623) effect. Post-hoc pairwise comparisons with Holm correction showed no significant differences between any condition pairs (all p = 1.000). The changes in positive arousal by stimulation paradigm are plotted in Figure 2e.

Changes in Negative Arousal

The repeated-measures ANOVA showed no significant changes in negative arousal across stimulation paradigms (F(2,16)=3.28, p=0.064) with a moderate to large effect size of n2p = .29. Post-hoc comparisons using Holm correction revealed no significant pairwise differences between conditions (all p > .170). Polynomial contrast analysis indicated a moderate effect size but no significant quadratic effect across conditions (p = .053, d = 0.80). The changes in negative arousal by stimulation paradigm are plotted in Figure 2f.

Directional Effect of Low Duty Cycle versus High Duty Cycle Stimulation

Directional Valence Index

The results revealed no significant difference in DVI between the two conditions, t((9) = −0.452, p = .662). These results suggest that, despite opposite theoretical neuromodulatory directions as a result of high versus low duty cycle, there was no statistically detectable difference in directional valence modulation between paradigm C and D. The DVI plotted against the Stimulation Paradigm is seen in Figure 3a.

Figure 3: Directional Effect of Low versus High Duty Cycle Stimulation on Behavior Task.

Figure 3:

(a) DVI by stimulation paradigm. (b) DAI by stimulation paradigm. (c) Changes in positive valence by stimulation paradigm. (d) Changes in negative valence by stimulation paradigm. (e) Changes in positive arousal by stimulation paradigm. (f) Changes in negative arousal by stimulation paradigm.

Directional Arousal Index

The DAI did not show significance on a paired t-test between Stimulation Paradigm C and Stimulation Paradigm D (t(9)= 0.387, p= 0.708). The DAI plotted against stimulation paradigm is seen in Figure 3b.

Change in Positive Valence

The change in positive valence did not show significance on a paired t-test between Stimulation Paradigm C and Stimulation Paradigm D (t(9)= -0.828, p=0.429). The change in positive valence plotted against Stimulation Paradigms C and D is seen in Figure 3c.

Change in Negative Valence

The change in negative valence did not show significance on a paired t-test between Stimulation Paradigm C and Stimulation Paradigm D (t(9)= 0.993, p= 0.347). The change in negative valence plotted against Stimulation Paradigms C and D is seen in Figure 3d.

Change in Positive Arousal

The change in positive arousal did not show significance on a paired t-test between Stimulation Paradigm C and Stimulation Paradigm D (t(9)= 1.106, p= 0.336). The change in positive arousal plotted against Stimulation Paradigms C and D is seen in Figure 3e.

Change in Negative Arousal

The change in negative arousal did not show significance on a paired t-test between Stimulation Paradigm C and Stimulation Paradigm D (t(9)= 1.454, p= 0.180). The change in negative arousal plotted against Stimulation Paradigms C and D is seen in Figure 3f.

Within-Paradigm Changes

To evaluate whether any stimulation paradigm independently produced behavioral changes, exploratory within-paradigm pre-post comparisons were conducted using Wilcoxon signed-rank tests on changes in valence, arousal, and composite indices (DAI and DVI). Across stimulation paradigms, several uncorrected tests demonstrated nominal significance (Table 3), particularly for composite arousal stimuli; however, given the number of exploratory comparisons performed (24 total tests), a conservative Bonferroni correction was applied across all behavioral outcomes (corrected α = 0.002). Under this global correction, no within-paradigm behavioral effect remained statistically significant. Effect sizes, expressed as rank-biserial correlations, were in some cases large, indicating consistent directional changes across subsets of participants; nevertheless, these findings were variable across outcomes and paradigms and did not survive correction for multiple comparisons.

Table 3.

Within-paradigm Pre-Post Behavioral Changes Across Stimulation Paradigms

Outcome-Stimulation Condition p Rank-Biserial Correlation (r)
StimA DVI 0.039 −0.778
StimB DVI 0.076 −0.689
StimC DVI 0.193 −0.491
StimD DVI 0.084 −0.636
StimA DAI 0.039 −0.778
StimB DAI 0.009 −1.000
StimC DAI 0.009 −0.945
StimD DAI 0.009 −0.945
StimA Negative Valence 0.123 −0.600
StimB Negative Valence 0.770 0.127
StimC Negative Valence 0.084 0.636
StimD Negative Valence 0.333 0.364
StimA Positive Arousal 0.407 −0.333
StimB Positive Arousal 0.124 −0.600
StimC Positive Arousal 0.407 −0.333
StimD Positive Arousal 0.014 −0.855
StimA Negative Arousal 0.012 0.956
StimB Negative Arousal 0.722 −0.156
StimC Negative Arousal 0.284 0.400
StimD Negative Arousal 0.221 −0.455

For each stimulation paradigm (A–D), within-subject pre–post change scores were tested against zero using Wilcoxon signed-rank tests. Reported values include uncorrected p-values and rank-biserial correlations (r), which quantify the consistency and direction of change across participants. These analyses were exploratory; after applying a conservative global Bonferroni correction across all behavioral comparisons (α = 0.002), no outcome remained statistically significant.

Discussion

This dose-escalation study demonstrated the initial safety and practicality of applying tFUS to the right amygdala in healthy volunteers, using intensities well above the FDA output limit threshold for diagnostic ultrasound of Ispta.3= 720 mW/cm2. Our primary safety aim was evaluated through clinical review of structural MRI, specifically, whether any of the stimulation parameter sets would create changes on structural MRI. This means of safety evaluation was employed to mirror methods used clinically during high-intensity focused ultrasound (HIFU) procedures, during which structural MRI is used to monitor for edema, hemorrhage, or unintended tissue damage during and after sonication[35]. Importantly, across all stimulation sessions, there were no structural changes observed on any MR scans. This held true even for the previously unused in-vivo stimulation conditions (Parameter Set C and D) with an Ispta.3 of 10.08 W/cm2, as they did not lead to any adverse effects, despite the delivered intensity corresponding to a free-field derated Ispta.3 approximately 14-fold higher than the FDA diagnostic ultrasound output limit, which is defined at the transducer face and does not account for transcranial attenuation. Additionally, MRI scans of patients who received Parameter Set D, which has a Mechanical Index of 2.8 (exceeding the iTRUSST value of 1.9 [36] ), did not show any of the structural changes, such as edema or microhemorrhage, that are typical with HIFU-like stimulation. Furthermore, there was a lack of any adverse events or serious adverse events during any stimulation session across subjects.

We acknowledge that because CT or pseudo-CT from ultrashort-echo MR imaging was not acquired, skull acoustic properties were estimated using T1-weighted MRI, which may reduce the quantitative accuracy of modeled intracranial intensities. However, because safety conclusions in this study were based on serial neuroradiological MRI assessment rather than simulation-derived exposure metrics, this limitation does not affect the primary safety conclusions. Acoustic modeling was performed solely to demonstrate accurate targeting and obtain a crude estimate of the intracranial effects of stimulation. Additionally, we note the lack of confidence intervals for the estimated stimulation intensities, which is a limitation of the modeling software itself. Despite the lack of pseudo-CT, we note that the estimated ISPTA at the target in all instances falls within the FDA regulatory limit. This suggests that investigators may increase free-field intensities, even substantially beyond current FDA limits as done in this study, safely while continuing to deliver safe stimulation paradigms. Together, these results show a level of safety within the parameter space used in our study.

Although our sample size was small, with ten subjects, these findings still suggest the safety and practicality of these parameters for in vivo neuromodulation studies. They also raise questions about the current FDA intensity guidelines for Ispta, set at 720 mW/cm2. Based on the evidence presented in this study and other research indicating that safe tFUS intensities exist beyond this limit, additional systematic investigation into higher-output parameter spaces is desperately needed to determine true safety margins prior to any regulatory reassessment of the existing limits. Such adjustments might also help facilitate the approval process for future studies within this parameter space, possibly making IRB approval easier to obtain, and allowing for faster study initiation and smoother translation of innovative tFUS protocols to clinical applications. Additionally, understanding ultrasound targeting using techniques such as magnetic resonance acoustic radiation force impulse imaging (MR-ARFI) also requires higher stimulation intensities[37].

Of course, further exploration of this parameter space must be done in a safe, controlled, and incremental manner. The WVU/RNI event serves as an unfortunate example of the possible danger when this is not done. In a followup to their initial publication, the authors provide some estimates of the parameters they used[38]. Their highest stimulation pressure in water was 3.5 MPa, which at a frequency of 220 KHz, corresponds to an MI of 7.46. Parameter set D in this study had an in-water MI of 2.8, placing it well below that of the WVU study. As such, these findings and the unfortunate incident underscore the need for systematic safety data at all stimulation intensities, with careful stepwise escalation, rigorous monitoring, and careful interpretation of studies approaching or exceeding these values. We also note that while recent iTRUSST guidance provides guidance on stimulation at biophysical levels that have never shown adverse events, these stimulations may be inadequate to elicit robust neuromodulatory effects. Further, the iTRUSST values should not be assumed to classify intensities above their stated values as inherently dangerous or unsafe. Accordingly, these guidelines should be interpreted as conservative reference points rather than definitive safety thresholds, particularly in the context of emerging neuromodulation applications where effective parameter ranges may extend beyond those historically studied. The present findings, together with prior work, suggest the existence of a broader therapeutic window that remains incompletely characterized, emphasizing the need for carefully designed dose-escalation studies that balance the pursuit of efficacy with rigorous safety oversight. Moving forward, the field would benefit from systematic mapping of this parameter space, integrating empirical safety outcomes, acoustic modeling, and real-time monitoring to more precisely define exposure limits that are both biologically informed and clinically meaningful.

In addition to supporting the structural and regulatory safety of higher-intensity tFUS, our findings also offer preliminary insights into its temporal profile, raising important questions about the duration and persistence of neuromodulatory effects following stimulation. Even with sessions spaced one week apart, there was no cumulative or residual effect on emotional reactivity or baseline emotional state, as measured by pre-stimulation behavioral task results, indicating that, even with significantly higher-intensity tFUS, neuromodulatory effects wore off between stimulation sessions. It is likely that with 10 subjects, we did not have sufficient power to detect an effect. While this further supports the temporal safety of this parameter space, it also points to an incomplete understanding of the temporal effects of tFUS neuromodulation. The precise duration of neuromodulatory effects induced by tFUS remains to be fully characterized, although prior studies in rodents and humans have demonstrated that such effects can persist for days to weeks[26, 39], underscoring the need for continued longitudinal evaluation within this new parameter space. Furthermore, much has been written about the acoustic artifact and its potential as a confounder in any analyses[40]. Unfortunately, this study was not designed to assess for the acoustic artifact, so this is another limitation to the generalization of these results. Lastly, subjective tolerability was not captured using a formal standardized questionnaire. Thus, while these results do demonstrate a preliminary level of safety of this parameter space and point towards further clinical trial testing, we are still unsure of the relationship, if any, between tFUS intensity and transient neuromodulatory effects.

The primary efficacy aim was to test whether increasing the intensity of ultrasound would produce a dose-dependent modulation of affective responses. Regarding our dose-response hypothesis, the results were rather mixed and inconclusive. Notably, no stimulation condition produced a statistically significant pre–post change relative to baseline after correction for multiple comparisons. A significant linear effect was observed in changes in negative valence across stimulation paradigms A to C; however, this effect was not supported by significant omnibus or pairwise comparisons, did not survive correction for multiple comparisons, and was not consistently reflected across related behavioral measures. Accordingly, this finding should be interpreted cautiously and does not provide robust evidence of a dose–response relationship. While the direction of this effect may be consistent with the known role of the right amygdala in negative emotional processing, the lack of statistical robustness and internal consistency precludes a reliable interpretation[41, 42]. Importantly, no participants reported subjective distress or adverse emotional effects. Given the small sample size and use of a healthy cohort, this finding should be considered exploratory and not indicative of maladaptive or countertherapeutic outcomes.

With respect to changes in arousal across paradigms A to C, the DAI and positive arousal changes did not reach statistical significance. Notably, a quadratic effect did not reach statistical significance, although a moderate effect size was observed across changes in negative arousal. While non-linear dose–response curves have been described in other neuromodulation contexts, particularly in studies targeting subcortical structures like the amygdala[12, 43], the present findings are inconclusive and likely reflect variability related to small sample size and limitations of the behavioral task.

When comparing Stimulation Paradigm C, with a high duty cycle (10.08 W/cm2 Ispta.3, 70% DC) and Stimulation Paradigm D, with a low duty cycle (10.08 W/cm2 Ispta.3, 5% DC), no metric reached a level of significance. It’s likely then that while these paradigms differ theoretically in their predicted neuromodulatory direction (based on high versus low duty cycle), the task used here was not sensitive enough to detect these specific effects, especially given the small sample size in this pilot study. Additionally, it’s possible that the modulation occurred at a neurophysiological level that did not lead to immediate directional changes in arousal or valence. As such, future work should ensure that concurrent neuroimaging or electrophysiology is collected to assess the polarity of the neuromodulation on a more granular level.

When conducting within-paradigm analyses, while several metrics showed nominal significance prior to correction, none survived conservative correction for multiple comparisons (Table 3), with a corrected significance threshold of α= 0.002. Given the exploratory nature of behavioral outcomes and the primary focus on safety, all statistical tests were treated as a single family and corrected using a Bonferroni procedure. Thus, combined with the small sample size, these results should be interpreted cautiously and may reflect limited power to detect small effects. Importantly, the absence of robust effects across paradigms is consistent with the intended safety properties of the system.

Collectively, these results do not provide robust evidence for a dose–response effect of tFUS intensity. Beyond the limitations imposed by the small sample size, it is also important to note that this trial was conducted in healthy subjects, and as such, the primary endpoint was safety rather than therapeutic efficacy. The tasks used are primarily meant for anxious individuals, not for healthy volunteers. Subsequently, the lack of clear and robust neuromodulatory effects was not unexpected. In fact, the lack of significant effects on affective stimuli might be considered a positive, since it suggests that the intensities used in the study did not have adverse effects on the subject’s behavior, adding another successful safety metric to this work. We anticipate that installing these higher intensity tFUS parameters in clinical populations with dysregulated affective circuitry, such as in mood or anxiety disorders, could result in clearer effects, especially in the context of targeting abnormally hyperactive or hypoactive circuits. This is especially relevant given that prior studies were constrained by the FDA’s diagnostic ultrasound intensity limit of Ispta.3 = 720 mW/cm2 may have underdelivered energy relative to what is needed for clinically meaningful engagement. Ultimately, replication of these findings in patient populations, with symptom-specific targets and outcomes, will be essential to determine the true translational value of higher-intensity tFUS.

Conclusion

This dose-escalation study provides preliminary evidence that low-intensity neuromodulatory tFUS delivered using free-field outputs above current FDA diagnostic output limits can be applied safely to the right amygdala in healthy volunteers. Across all stimulation sessions, structural MRI and behavioral monitoring revealed no adverse events, edema, hemorrhage, or other tissue disruption, even at intensities exceeding the current FDA diagnostic ultrasound output limit and the iTRUSST recommended maximum. These findings provide empirical data relevant to ongoing discussions about whether diagnostic ultrasound output limits are optimally calibrated for neuromodulation research, while underscoring the need for larger, multi-site studies before any regulatory reconsideration. While exploratory analyses suggested possible dose-dependent effects on affective processing, particularly in negative valence and arousal, these results were inconsistent and likely limited by small sample size and the use of healthy controls. The complete absence of adverse effects, together with preliminary signals of neuromodulatory engagement, supports the continued evaluation of higher-intensity parameters in clinical populations. Future studies with larger samples, patient cohorts, and concurrent functional neuroimaging will be essential for clarifying both the therapeutic potential and mechanistic specificity of tFUS beyond current FDA and iTRUSST limits.

Supplementary Material

1
2
3
4

Funding

Funding Sources: NIH T32GM008042 (to NMS), NIH F30MH136802 (to NMS), DOD CDMRP HT9425-24-1-1081 (to MMM), NIH T32GM141938 (to BHD)

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflict of Interest Disclosures

Norman Spivak reports financial support was provided by National Institute of Mental Health. Norman Spivak reports a relationship with BrainSonix Corporation that includes: consulting or advisory and travel reimbursement. Andrew Bishay reports a relationship with BrainSonix Corporation that includes: consulting or advisory. Alexander Bystritsky reports a relationship with BrainSonix Corporation that includes: board membership and equity or stocks. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

References

  • [1].Hoang-Dang B, Halavi SE, Rotstein NM, Spivak NM, Dang NH, Cvijanovic L, et al. Transcranial Focused Ultrasound Targeting the Amygdala May Increase Psychophysiological and Subjective Negative Emotional Reactivity in Healthy Older Adults. Biol Psychiatry Glob Open Sci 2024;4(5):100342. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [2].Spivak NM, Swenson AJ, Haroon J, Holland B, Dancy M, Collins E, et al. Corrigendum to “Evaluating transcranial focused ultrasound stimulation (tFUS) for targeted neuromodulation in generalized anxiety disorder: A double-blind feasibility study”[Brain Stimul 18 (1)(2025) 415–416]. Brain stimulation 2025;18(4):1244. [Google Scholar]
  • [3].Cain JA, Spivak NM, Coetzee JP, Crone JS, Johnson MA, Lutkenhoff ES, et al. Ultrasonic Deep Brain Neuromodulation in Acute Disorders of Consciousness: A Proof-of-Concept. Brain Sci 2022;12(4). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Nicodemus NE, Becerra S, Kuhn TP, Packham HR, Duncan J, Mahdavi K, et al. Focused transcranial ultrasound for treatment of neurodegenerative dementia. Alzheimers Dement (N Y) 2019;5:374–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Bystritsky A, Korb AS, Douglas PK, Cohen MS, Melega WP, Mulgaonkar AP, et al. A review of low-intensity focused ultrasound pulsation. Brain stimulation 2011;4(3):125–36. [DOI] [PubMed] [Google Scholar]
  • [6].Blackmore J, Shrivastava S, Sallet J, Butler CR, Cleveland RO. Ultrasound Neuromodulation: A Review of Results, Mechanisms and Safety. Ultrasound in medicine & biology 2019;45(7):1509–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Pasquinelli C, Hanson LG, Siebner HR, Lee HJ, Thielscher A. Safety of transcranial focused ultrasound stimulation: A systematic review of the state of knowledge from both human and animal studies. Brain stimulation 2019;12(6):1367–80. [DOI] [PubMed] [Google Scholar]
  • [8].Toccaceli G, Barbagallo G, Peschillo S. Low-intensity focused ultrasound for the treatment of brain diseases: safety and feasibility. Theranostics 2019;9(2):537–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Yoo SS, Kim H, Min BK, Franck E, Park S. Transcranial focused ultrasound to the thalamus alters anesthesia time in rats. Neuroreport 2011;22(15):783–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Yoo SS, Lee JH, Fischer K, Zhang Y, Bystritsky A, McDannold N, et al. FUS-mediated reversible modulation of region-specific brain function. Washington, D.C.: Proceedings of MRgFUS; 2008:10. [Google Scholar]
  • [11].Spivak NM, Korb AS, Reyes SD, Bych BP, Schafer SF, Khanlou N, et al. Histological examination of focused ultrasound effects on human brain tissue. Brain stimulation 2021;14(6):1486–8. [DOI] [PubMed] [Google Scholar]
  • [12].Yoon K, Lee W, Lee JE, Xu L, Croce P, Foley L, et al. Effects of sonication parameters on transcranial focused ultrasound brain stimulation in an ovine model. Plos One 2019;14(10):e0224311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Stern JM, Spivak NM, Becerra SA, Kuhn TP, Korb AS, Kronemyer D, et al. Safety of focused ultrasound neuromodulation in humans with temporal lobe epilepsy. Brain stimulation 2021;14(4):1022–31. [DOI] [PubMed] [Google Scholar]
  • [14].Dallapiazza RF, Timbie KF, Holmberg S, Gatesman J, Lopes MB, Price RJ, et al. Noninvasive neuromodulation and thalamic mapping with low-intensity focused ultrasound. J Neurosurg 2018;128(3):875–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Attali D, Tiennot T, Schafer M, Fouragnan E, Sallet J, Caskey CF, et al. Three-layer model with absorption for conservative estimation of the maximum acoustic transmission coefficient through the human skull for transcranial ultrasound stimulation. Brain stimulation 2023;16(1):48–55. [DOI] [PubMed] [Google Scholar]
  • [16].Bishay A, Swenson AJ, Spivak NM, Schafer S, Bych BP, Gilles SD, et al. Preliminary Examination of the Effects of Focused Ultrasound on Living Skin and Temperature at the Skin-Transducer Interface. Bioengineering (Basel) 2024;11(11). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Krokhmal A, Simcock IC, Treeby BE, Martin E. A comparative study of experimental and simulated ultrasound beam propagation through cranial bones. Phys Med Biol 2025;70(2). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Izadifar Z, Babyn P, Chapman D. Mechanical and Biological Effects of Ultrasound: A Review of Present Knowledge. Ultrasound in medicine & biology 2017;43(6):1085–104. [DOI] [PubMed] [Google Scholar]
  • [19].Rezai A, Ranjan M, Bhagwat A, Arsiwala T, Carpenter J, Schafer M, et al. Brain injury during focused ultrasound neuromodulation for substance use disorder. Brain Stimulation: Basic, Translational, and Clinical Research in Neuromodulation 2025;18(6):2050–3. [DOI] [PubMed] [Google Scholar]
  • [20].Klein-Flugge MC, Airan RD, Attali D, Aubry J-F, Bubrick EJ, Caskey CF, et al. Open letter on intervention regimes and adverse events in focused ultrasound for neuromodulation. Brain Stimulation: Basic, Translational, and Clinical Research in Neuromodulation 2026;19(1). [DOI] [PubMed] [Google Scholar]
  • [21].Davis M. The role of the amygdala in fear and anxiety. Annu Rev Neurosci 1992;15:353–75. [DOI] [PubMed] [Google Scholar]
  • [22].Rauch SL, Shin LM, Wright CI. Neuroimaging studies of amygdala function in anxiety disorders. Ann N Y Acad Sci 2003;985(1):389–410. [DOI] [PubMed] [Google Scholar]
  • [23].Keil VC, Borger V, Purrer V, Groetz SF, Scheef L, Boecker H, et al. MRI follow-up after magnetic resonance-guided focused ultrasound for non-invasive thalamotomy: the neuroradiologist’s perspective. Neuroradiology 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Legon W, Adams S, Bansal P, Patel PD, Hobbs L, Ai L, et al. A retrospective qualitative report of symptoms and safety from transcranial focused ultrasound for neuromodulation in humans. Sci Rep 2020;10(1):5573. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Dell’Italia J, Sanguinetti JL, Monti MM, Bystritsky A, Reggente N. Current State of Potential Mechanisms Supporting Low Intensity Focused Ultrasound for Neuromodulation. Front Hum Neurosci 2022;16:872639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Cain JA, Spivak NM, Coetzee JP, Crone JS, Johnson MA, Lutkenhoff ES, et al. Ultrasonic thalamic stimulation in chronic disorders of consciousness. Brain stimulation 2021;14(2):301–3. [DOI] [PubMed] [Google Scholar]
  • [27].Schafer ME, Spivak NM, Korb AS, Bystritsky A. Design, Development, and Operation of a Low-Intensity Focused Ultrasound Pulsation (LIFUP) System for Clinical Use. IEEE Trans Ultrason Ferroelectr Freq Control 2021;68(1):54–64. [DOI] [PubMed] [Google Scholar]
  • [28].Bystritsky A, Spivak NM, Swenson AJ, Bishay A, Haroon JM, Monti M, et al. Corrigendum to “Development and implementation of a transcranial focused ultrasound device”[Brain Stimul 18 (1)(2025) 530]. Brain Stimulation: Basic, Translational, and Clinical Research in Neuromodulation 2025;18(4):1242. [Google Scholar]
  • [29].Pichardo S. BabelBrain: An Open-Source Application for Prospective Modeling of Transcranial Focused Ultrasound for Neuromodulation Applications. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control 2023. [DOI] [PubMed] [Google Scholar]
  • [30].Cain JA, Visagan S, Monti MM. S.M.A.R.T. F.U.S: Surrogate Model of Attenuation and Refraction in Transcranial Focused Ultrasound. Plos One 2022;17(10):e0264101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Brandts SD, Staveland BR, Madhusudhan A, Ash RT, Di Ianni T, Sugrue LP, et al. Empirical limitations of current low-intensity focused ultrasound simulation platforms. medRxiv 2026. [DOI] [PubMed] [Google Scholar]
  • [32].Lang PJ, Bradley MM, Cuthbert BN. International Affective Picture System (IAPS) [Database record]. APA PsycTests 2005. [Google Scholar]
  • [33].Lang PJ, Bradley MM, Fitzsimmons JR, Cuthbert BN, Scott JD, Moulder B, et al. Emotional arousal and activation of the visual cortex: an fMRI analysis. Psychophysiology 1998;35(2):199–210. [PubMed] [Google Scholar]
  • [34].Constantinescu AC, Wolters M, Moore A, MacPherson SE. A cluster-based approach to selecting representative stimuli from the International Affective Picture System (IAPS) database. Behavior research methods 2017;49(3):896–912. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [35].Ghanouni P, Pauly KB, Elias WJ, Henderson J, Sheehan J, Monteith S, et al. Transcranial MRI-Guided Focused Ultrasound: A Review of the Technologic and Neurologic Applications. AJR Am J Roentgenol 2015;205(1):150–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [36].Aubry JF, Attali D, Schafer ME, Fouragnan E, Caskey CF, Chen R, et al. ITRUSST Consensus on Biophysical Safety for Transcranial Ultrasound Stimulation. Brain stimulation 2025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Spivak NM, Kuhn TP. Variations in targeting techniques of focused ultrasound for use in neuromodulation. Brain stimulation 2019;12(6):1595–6. [DOI] [PubMed] [Google Scholar]
  • [38].Rezai A, Ranjan M, Bhagwat A, Arsiwala T, Carpenter J, Schafer M, et al. Further clarification regarding brain injury during focused ultrasound neuromodulation for substance use disorder. Brain stimulation 2026;19(1):103014. [DOI] [PubMed] [Google Scholar]
  • [39].Chu PC, Huang CS, Ing SZ, Yu HY, Fisher RS, Liu HL. Pulsed Focused Ultrasound Reduces Hippocampal Volume Loss and Improves Behavioral Performance in the Kainic Acid Rat Model of Epilepsy. Neurotherapeutics 2023;20(2):502–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40].Guo H, Hamilton M 2nd, Offutt SJ, Gloeckner CD, Li T, Kim Y, et al. Ultrasound Produces Extensive Brain Activation via a Cochlear Pathway. Neuron 2018;98(5):1020–30 e4. [DOI] [PubMed] [Google Scholar]
  • [41].Baas D, Aleman A, Kahn RS. Lateralization of amygdala activation: a systematic review of functional neuroimaging studies. Brain Res Brain Res Rev 2004;45(2):96–103. [DOI] [PubMed] [Google Scholar]
  • [42].Lanteaume L, Khalfa S, Regis J, Marquis P, Chauvel P, Bartolomei F. Emotion induction after direct intracerebral stimulations of human amygdala. Cereb Cortex 2007;17(6):1307–13. [DOI] [PubMed] [Google Scholar]
  • [43].Kim H, Park MY, Lee SD, Lee W, Chiu A, Yoo SS. Suppression of EEG visual-evoked potentials in rats through neuromodulatory focused ultrasound. Neuroreport 2015;26(4):211–5. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1
2
3
4

RESOURCES