Abstract
This study provides a review of recent publications on different quality control (QC) techniques for radiotherapy linear accelerators (linacs). QC in radiotherapy is essential to ensure patient safety, deliver optimal quality treatments, monitor equipment performance over time, and to satisfy statutory or recommended testing. While conventional QC methods are still widely used and valuable, there have been advances in QC methods and approaches which aim to improve efficiency and keep pace with the rapidly advancing complexity of clinical equipment and treatment delivery methods. Existing reviews of QC concepts mainly focus on single QC techniques. There are limited reviews that explore and compare the variety of techniques to QC, which is the aim of this report. QC techniques were separated into 7 domains: conventional QC, automated and manufacturer integrated QC, risk, statistical, artificial intelligence (AI), end-to-end and patient-specific QC (PSQC), external dosimetry audits and clinical trial credentialing audits. An overview of each technique with example studies from literature is presented.
Keywords: radiotherapy, quality control, QC, quality assurance, QA, PSQC, PSQA, FMEA, SPC, end-to-end, AI, audits, clinical trials
Introduction
Radiotherapy technology, clinical knowledge, and treatment techniques have progressed rapidly in recent years, with complex delivery techniques such as volumetric-modulated arc therapy (VMAT), stereotactic ablative body radiotherapy (SABR), and motion-gated treatments now forming part of standard, modern practice.1, 2 Adaptive treatments and advanced patient set-up verification including image-guided radiotherapy and surface-guided radiotherapy (SGRT), are becoming increasingly established within clinical workflows.3, 4
Technical advances in radiotherapy delivery techniques has increased the complexity of required quality assurance (QA) and quality control (QC) procedures and conventional methods have limitations in what they can test.5 Performing QC is time-consuming, resource heavy, and increasingly time-constrained as clinical pressures increase.6, 7 Therefore, determining optimal QC techniques which balance capability in detecting performance degradation against efficiency is essential.
QC is required in external beam radiotherapy to support patient safety, monitor equipment performance over time, and for statutory compliance8–10 and this can be achieved via a range of techniques. Literature showcasing radiotherapy QC practices is often fragmented between these different techniques, for example, reviews are available on statistical process control (SPC),5 end-to-end (E2E) QC,11 artificial intelligence (AI),12 and patient-specific QC (PSQC).13 Kron et al14 details 3 generations of “QC approaches”; “generic and prescriptive” tests, equipment/application-specific, and risk-based, and describes the emergence of a fourth generation with the use of AI. This report is based on similar themes but considers 7 techniques to QC: conventional QC, automated and manufacturer integrated, risk-based, statistical-based, AI, E2E and PSQC, and dosimetry audits and clinical trial credentialling audits.
A conventional linac QC technique employs component testing and historically follows a defined set of tests, with tolerances and frequencies prescribed by governing or professional bodies.14 While this does encourage standardized practice and a minimum level of safety, this technique is time-consuming15, 16 and may not fully capture the intricacies of an increasingly integrated workflow,5, 14 or full risks and performance trends. Manufacturers are increasingly designing “on board” QC methods integrated into their treatment systems. Utilizing these methods can increase efficiency but may not provide sufficient independent performance verification.17–19 Risk-based and statistical techniques demonstrate a more proactive identification of vulnerabilities and provide data-driven process improvement.20, 21 AI is starting to become a useful tool to detect errors, predict performance, and automate QC tasks.12 E2E QC methods take a macroscopic approach by testing an output of the entire system potentially catching errors that component testing alone would not11, 22 but are often not able to point to the cause of an error.11, 19 The role of measured PSQC, in conjunction with machine QC, is under examination to determine the optimal combination for efficiency and clinically relevant error detection.6, 13, 14, 23 Interdepartmental dosimetry audits and clinical trials accreditation have continued importance to ensure consistency and quality of treatment between radiotherapy departments24 and to motivate departments to modernize their radiotherapy processes.25
The aim of this work is to review literature on these different techniques to linac QC, not to answer the question of which technique is “best,” but to inform how these techniques can complement each other to be used to create a comprehensive approach to radiotherapy QC.
Methods
A systematic literature search was conducted to identify publications related to QC separated into the 7 thematic techniques, as above.
Searches were conducted in 3 data bases, Scopus, Science Direct, and Medline, and were restricted to publications between January 2015 and December 2025. Search terms were identified related to the 7 techniques. Results were long listed by title and shortlisted by abstract. This information was supplemented by reviewing reference lists and the authors’ own sources and including key publications outside of the date range if they were the latest guidance available.
This report defines QA as the overarching framework for maintaining quality and safety. QC is defined as the specific tests that are performed to verify performance. However, it is acknowledged that the literature often considers these terms interchangeable. Therefore, when discussing activities that fall under this report’s definition of QC, but the literature has referred to as QA, the original wording has been kept.
Results
Conventional
Historically, the basis of a QC program is built on testing individual linac components in isolation using test methods, frequencies, and tolerances14 that are not specific to the linac manufacturer, age, or workload. These are typically prescribed by regulatory and professional bodies via guidance documents and reports, such as the those published by the Institute of Physics and Medicine (IPEM),26 the American Association of Physicists in Medicine (AAPM),27–30 the Netherlands Commission for Radiation Dosimetry (NCS),31, 32 and Australasian College of Physical Scientists and Engineers in Medicine (ACPSEM).33 An aim of these guidance documents is to standardize practice; however, a 2024 UK survey of linac QC34 demonstrated a variability in implementation of QC practices, despite 91% of responding centers stating that they used IPEM Report 81 (2nd edition, 2018)26 to inform their QC schedule.34 In addition, the rate of publication of official guidance can struggle to keep pace with rapidly advancing radiotherapy techniques.5 Furthermore, Barnes13 describes how some IPEM26 and AAPM27, 29 reports mainly detail static testing of linac components and questions their validity of testing modern dynamic radiotherapy. Barnes13 also states that while the NCS reports31, 32 detail more dynamic testing protocols, it is unclear if these methods have become widespread.13
Conventional tolerances are defined thresholds that, when used simplistically, require a performance parameter to deteriorate past a limit before a fault is identified. This can result in binary decision-making, and often does not consider the variability within the threshold which could indicate deteriorating performance prior to breakdown.5, 35
Conventional QC frequencies are fixed time intervals and do not consider the age or workload of treatment machines, which could impact upon the required safe testing frequency.36 A conventional technique is often tedious and time-consuming due to the component-based nature and manual methods employed,15, 16 and does not consider the failure modes that exist when components are combined as in treatment delivery.5, 14
Automated and manufacture integrated
Manufacturer integrated QC (MIQC) is defined by Pearson et al17 as “An automated quality control (QC) tool supplied by the treatment machine vendor.”17 MIQC utilizes existing linac components and data, such as the electronic portal imaging device (EPID) and machine log files, to automatically perform and analyze QC tests. The results of a 2022 IPEM survey into the use of MIQC17 stated that 25% of survey respondents envisaged the potential future scope of MIQC as “detecting errors ahead of time,” 21% as “alternative to daily QA,” and 21% “reduce the burden on monthly QC.”17 The follow-up IPEM topical report: guidance on the use of MIQC18 stated that benefits may include “reduced error, standardization, and big-data analysis.”18
Despite these benefits, there is still uncertainty as to whether MIQC can be solely relied upon to verify machine performance. Multiple studies have compared MIQC methods to conventional methods.37–41 The use of an EPID may not be a sufficient replacement for output measurements due to uncertainties in dose measurement compared to conventional methods.38, 41–43 To reduce this uncertainty, several guidance documents recommend comparison of chamber and Varian Machine Performance Check (MPC) (Varian Medical Systems, Palo Alto, CA, USA) output as a routine QC test.26, 34, 38, 41
The independence of MIQC to the delivery device being tested has also been raised as a possible concern.17, 18 A total of 81% of the 2022 IPEM survey respondents considered EPID measurements to meet the threshold for independence as the EPID panel is independent of the treatment beam parameters it tests.17 This principle, given in the guidance report by Pearson et al,18 defines 3 aspects of independence: financial, manufacturer, and technical. For MIQC to be implemented technical independence should be proven by the user. In addition, an overlap period of performing both MIQC and conventional QC is recommended.18
Third-party automated QC has been assessed in a number of studies.43–45 Bonanno et al45 compared DoseLab AutoQA module (Varian Medical Systems, Palo Alto, CA, USA) to conventional QC methods for monthly routine QC. Potential time savings were among the parameters evaluated. DoseLab took 80 minutes to complete monthly QC compared to 195 minutes for the conventional technique. The time savings were mainly seen in the analysis, only 20 minutes of machine time was saved compared to the 115 minutes overall time saving. Stambaugh et al43 compared the accuracy and efficiency of SunCHECK Machine (Sun Nuclear Corporation, Melbourne, FL, USA) with Varian MPC for daily QC. Output measurements were also compared to measurements acquired by their monthly standard, the IC profiler (Sun Nuclear Corporation, Melbourne, FL, USA). Stambaugh et al43 found that SunCHECK machine tests took on average 22 minutes compared to 15 minutes with MPC. MPC output saw a drift in agreement with SunCHECK of 1.4% over 3 months, whereas SunCHECK drifted in line with the IC profiler. MPC drift has been found in other studies38, 41, 42 and as a result, regular re-baselining is recommended to maintain confidence in the results.42 This approach has been taken in the UK where 21% of centers perform monthly comparisons between MPC and ion chamber to monitor for drift.34 SunCHECK machine was stated as being preferred as it could cover more AAPM TG-142 recommendations27 than MPC, yet MPC was deemed a useful addition as it provides valuable technical information for service engineers.43
This use of machine log files as a method of machine-reported performance is also regularly investigated. Log files can be used to compare intended and actual parameters during delivery.46–48 Son et al49 used comparison of multi-leaf collimator (MLC) and monitor unit (MU) log files to determine if their Halcyon performance was in compliance with AAPM TG 14227 guidelines. As a result of this analysis, Son et al49 determined log file analysis could be used to test linac performance parameters and suggested further work comparing log files with other QC methods (eg, EPID or phantom measurements) should be carried out. Wojtasik et al50 performed multivariate analysis on MLC log files to predict MLC failure before it occurred. Their model was found to have a 62% MLC failure prediction sensitivity, which allowed for prospective maintenance. The model was not able to identify which leaf required replacement. Many other studies employ the use of both log files and deep learning methods to predict MLC positions51–53 (see Artificial Intelligence section for further detail).
Risk
Risk-based methods evaluate a whole process and determine which QC should be performed to reduce the highest rated risks in a system. AAPM TG-10054 gives advice on the application of risk analysis to radiotherapy, specifically failure modes effects analysis (FMEA). Liu et al55 conducted a literature review of the use of FMEA for proactive healthcare risk assessment. While the review was not specific to radiotherapy, the majority of results returned were in this field. In addition, Liu et al55 states that FMEA in healthcare should consider more risk factors in addition to the conventional 3 (occurrence [O], severity [S], and detectability [D]). This was done by Ochi et al56 when evaluating the frequency and tolerances of standard outputs. An additional 2 indices were added: severity to output drifting and number of treatment fractionations. Different frequency and tolerance combinations were used to calculate Risk Priority Numbers (RPNs) and quantitatively determine the optimal combination for either safety or efficiency. Conversely, Rah et al57 recommends the use of a modified healthcare FMEA (mFMEA) using 2 indices (severity and frequency) as it was found to be conceptually easier to implement. Kawahara et al58 used 3 indices, but these differed from the conventional O, S, and D. Instead, the indices used were biological effect, occurrence, and cost of labor and repairs to assess the impact of treatment interruption due to machine breakdown in terms of biological effect to the patient and the economic cost to the institution. They found that MLC faults had the highest biological risk and laser guard the highest economic risk.
O’Daniel and Yin59 used FMEA to determine clinic-specific linear accelerator QC test frequencies for the tests detailed in AAPM TG-142.27 Occurrence was determined from 9 years of linac QC results, severity was determined through introduced errors in head and neck plans, and detectability was set to be equal for all TG-142 tests considered. This method claimed to reduce bias in comparison to a method based on expert opinion. The greatest severity was found for daily imaging QC and the greatest occurrence for output and laser QC. Consequently, output and laser QC was conducted daily and imaging QC weekly, reducing daily QC time by 5-10 minutes. Bonfantini et al60 used both quantitative methods, via QC records, and expert opinions, from surveys, to determine risk scores to prioritize linac QC. This allowed for reduced bias while maintaining staff experience. Pearson et al61 used a combination of QC records and scoring surveys to validate a hybrid QC approach of conventional QC and MIQC. In this case MPC, daily QA3 (Sun Nuclear Corporation, Melbourne, FL, USA), and Delta4 (ScandiDos AB, Uppsala, Sweden) were used. FMEA helped determine that a hybrid approach would be appropriate, but more MLC QC tests would be beneficial.
De Kerf et al62 used FMEA to assess the risk of a fully automated workflow (FAW). The highest risk found was staff interaction with the FAW, for example, poor judgment of the outcome and a deskilled workforce. This was of particular concern when considering the generalizability of the FAW in conjunction with the inherent variability of patient populations. The best outcome for all patients could not be guaranteed, highlighting the need for staff education and new QC tools capable of testing these automated platforms.
Statistical
Statistical Process Control (SPC) trends QC results over time and can predict when a process starts to go out of tolerance before failure occurs. In a report of the application of SPC within radiotherapy, Xiao and Li5 stated the need to move away from conventional techniques that make binary decisions on predetermined tolerances, ignoring variability with the tolerances which does not support process improvement or mitigate against equipment failure and unplanned maintenance/repairs. Instead, move toward SPC for QC such as MLC and PSQC to be able to flag deviations from normal performance which could indicate a fault. Raveendran et al35 states ignoring trends within conventional tolerances could lead to passive negligence and propose SPC to minimize variation, with the caveat that this is not always achievable in a clinical situation given human involvement. Raveendran et al35 recommends using SPC to get machine, site and technique specific tolerances and actions for better PSQC monitoring.
Ma et al63 performed SPC via the 6-sigma methodology64 to determine MPC tolerances, as they questioned the independence of vendor tolerances. Two years of MPC results were collated to determine capability indices and associated tolerances and tested with 5 months of data. MPC vendor tolerances for isocentre size, MLC maximum offset, and rotation-induced couch shift were found to be too small. This knowledge reduced unnecessary engineer call outs and helped prevent fault blindness. All other calculated SPC tolerances were smaller than the vendor’s data and could identify a fault that resulted in machine breakdown that went previously unnoticed.
Able et al65 proposed a model for preemptive maintenance of linacs by performing SPC on daily QC log files. Baseline tolerances were obtained by performing SPC on the log files of 7 linacs from 5 centers. Future log files were recorded and displayed on a dashboard against the tolerances. The dashboard was able to identify faults prior to occurrence and prompt preemptive maintenance. The authors stated that while it showed promise long-term monitoring, further evaluation will be required to establish the effectiveness of the model. Kawahara et al58 performed FMEA to assess the biological effects to patients due to treatment interruption due to machine breakdown and stated that the method proposed by Able et al65 would aid in reducing this risk.
Artificial intelligence
Research into the use of AI in radiotherapy is rapidly developing66, 67 and automated QC is a promising application of machine learning (ML).68 A common theme of investigation was the use of AI models to predict the outcome of PSQC results.69–73 Ono et al12 found that, as of December 2023, more articles had been published on the use of AI within PSQC than in machine QC. AI has been trained on existing plans to link plan complexity metrics to PSQC results,70, 71 identify PSQC dosimetric errors,74 identify planning data errors,75 and predict actual MLC positions.51–53, 76 This information can be fed back into plan optimization52, 70 and/or can be used to reduce PSQC measurements.51, 71, 72 Concerns that too much confidence would be placed on the model and limit safe implementation exist.72, 73 Therefore, some researchers went a step further and calculated the uncertainty of the predictions to aid the user in decision-making.72, 73
“Big data” of past QC results could be used to optimize future performance; by offering the opportunity to personalize QC pathways for patients with similar cases77 or predicting machine faults.78 Kalet et al79 also queries whether vendors could use the linac log files they collect to develop specific maintenance and repair schedules. Ma et al80 used daily QC results to develop a neural network model that could predict the record and trend of data variations and was robust when applied to a different radiotherapy machine and could foresee possible machine failure.
Mehrens and Kry81 used previous data obtained from Imaging and Radiation Oncology Core (IROC) thoracic audits and ML models to predict the pass/fail outcomes of dosimetry audits. Different treatment and motion-management approaches were found to influence audit performance. The results could be used to give centers personalized QC programs tailored to specific treatment setups.
E2E and patient-specific QC
An E2E technique examines a workflow and can identify potential failure modes between steps that would be missed by testing components in isolation. For example, Mertens et al82 considered the geometric uncertainty, which is increasingly important with the development of high precision treatments, in addition to the dosimetric uncertainty. Stating that it is hard to carry out these tests separately due to introduction of setup errors. Therefore, they developed an in-house 3D printed phantom that combined the 2. The use of phantoms to facilitate E2E is often employed, and literature pertaining to the design of novel phantoms for this purpose is common.83–88 Ranging from phantoms designed to interrogate the entire workflow from CT/MR to patient treatment86 to those designed for smaller sections of the pathway, such as SGRT only.87
E2E testing is recommended in guidance, as part of commissioning or an annual QC26 and as a method of testing a black-box system.19 Results of a 2024 UK linac QC survey34 suggest mixed uptake with 49% of centers stating that E2E testing is performed, of which 21% clarified only as part of commissioning or external audits, and only 7% stated routine E2E QC (5% annually, 2% monthly). Loughery et al89 produced a protocol for monthly E2E QC designed to detect complex errors not caught by conventional component-based QC. The protocol was validated by 3 centers and results were found to be consistent, suggesting that the protocol has potential use in future E2E QC initiatives. Schreiner11 also advocated for the increased use of E2E within radiotherapy in a comprehensive review article. They state that while conventional QC verifies that individual steps are technically achievable, it does not consider whether the staff are carrying out the process correctly (human factors). They suggest this is where a regular E2E program could be a beneficial training and monitoring exercise for the performance of the entire system. Schreiner11 caveats that while E2E testing catches errors conventional QC does not, E2E cannot be used in isolation as the testing is not specific enough to identify the cause of potential failures. In which case conventional QC records should be checked to troubleshoot.11
Schreiner11 presents 4 levels of QC tests, higher levels represent more system level testing: from level 1 component testing, to level 4 E2E of whole process. PSQC is defined as a level 3 QC test and thought to be a pseudo-E2E test. Despite this, Schreiner11 states that PSQC should not be carried out in place of E2E tests, as audits have found the ability to identify errors as a result of PSQC is not always sufficient.90, 91 For example, Kry et al90 compared in-house IMRT QC results with IROC QC results completed prior to clinical trials and found that in-house IMRT had a poor ability to detect a failing result. More recently, the SEAFARER study by Lehmann et al91 found centers passing their own PSQC, yet failing external audits which implied that centers could not detect clinically meaningful delivery errors. These results, alongside QC survey results, which stated PSQC was a significant burden on departments92 and plans were rarely modified as a result of PSQC results,34, 93 prompted Barnes13 to suggest a departure from measurement-based PSQC, provided a comprehensive sub-system QC program is in place.
The desire to reduce measurement-based PSQC is common. O’Daniel et al23 used FMEA to investigate combining PSQC and conventional QC methods to determine the optimal combination for detecting errors that did not require measuring every plan. TG-10054 was used to identify the failure modes that could be detected with PSQC. First, no PSQC was assumed which gave a detectability score of 0, and this was repeated for different combinations of PSQC types. Overall, a combination of log file-based QC with 3D dose calculation and MLC QC was determined to be optimal and allowed centers to consider a reduction in their measured QC. In 2025, O’Daniel et al94 queried which errors PSQC should catch as part of a joint AAPM-ESTRO TG-360 subgroup. Failure modes that could cause greatest patient harm were identified and given priority for PSQC testing. To identify errors that machine QC alone would not detect, multiple methods of PSQC were required, from secondary dose calculation, transfer checks, and in vivo dosimetry.
Conversely, Mills and Colligan,6 proposed that the introduction of PSQC allows for a reduction of prospective QC (PQC). They argued that PQC demonstrates a machine is capable of treatment delivery, but the design of these prospective tests becomes too complex with the introduction of more advanced delivery techniques. As a result, PSQC becomes unavoidable as it verifies treatment delivery. However, given limited resources, it is impractical to increase PSQC without the reduction of PQC. Therefore, Mills and Colligan6 proposed a framework that allows for the reduction of PQC by approximately 50%, provided that weekly PSQC is performed on each linac.
Audits and clinical trials
Dosimetry audits are facilitated by national and international audit groups, for example, IPEM interdepartmental dosimetry audit group (IDA) in the UK,95 IROC in the US,96 BELdART in Belgium,97 ACDS in Australia,96 and the IAEA internationally.98 Dosimetry audit credentialing is often a requirement for participation in clinical trials to ensure that a center is capable of specific clinical delivery techniques and to reduce variable center-specific practice from influencing results, and potentially underpowering clinical trials. Again, these are facilitated by approved groups, for example, RTTQA in the UK,95 JCOG in Japan,99 and DMCG in Denmark.100
There have been many review articles on dosimetry audits. Clark et al25 published a review article in 2015, concluding undertaking regular audits help modernize practice and ensure prescribed dose is delivered. Ibbott and Thwaites101 published a review regarding dosimetry audits for more complex techniques. Muhammad et al102 wanted to ensure that for accuracy in radiation dose delivery, external audits had an equal importance with routine in-house QC. Pasler et al103 wrote a review article on novel methodologies for dosimetry audits in 2018, which included the use of deformable and anthropomorphic phantoms, as well as novel approaches to non-phantoms such as log files and EPID measurements, but stated that the harmonization of dosimetry audit techniques are required first. Brooks et al104 focused on auditing outlining in clinical trials, while not machine QC, outlining is still an important part to audit if considering the full radiotherapy pathway.
Lehmann et al91 proposed a new concept for validating PSQC for clinical trial participation. Centers sent treatment plans to the audit team. Delivery errors were introduced to the plans and returned. Participating centers performed PSQC on these plans and informed the audit team of their results. Centers were not always able to detect the clinically significant delivery errors and often passed plans that increased the risk of spinal cord injury. Also, sometimes plans were failed that had no clinically significant impact. No consistency was found between centers, leading Lehmann et al91 to state that the sensitivity of PSQC should be audited, optimized, and justified as clinical trial data is pooled and therefore consistency is critical. On a similar theme, Brooks et al105 developed reference plans with introduced errors to compare various audit methodologies within the Global Harmonization Group (GHG),106 as part of the GHG work to test the comparability of audit systems and align international audits across the globe. Jomy et al107 performed a systematic review between 2021 and 2024, which showed that deviations from trial protocols were linked to worse clinical outcomes. Retrospective QC was not shown to correct existing issues. Deviations were found in contouring, planning, and dose delivery, proving the importance of pretrial participation E2E audits.
Looking to the future, Misson et al95 published a position paper on audits in the UK. IPEM has interdepartmental audit (IDA) groups setup across the UK in which regions work among themselves to carry out annual dosimetry audits. NPL and RTTQA may perform voluntary or obligatory audits prior to centers going clinical with new techniques and equipment and if they request to participate in clinical trials. Misson et al95 proposes to centralize these audits, to allow IDA groups to benchmark against each other, and reduce demand on NPL and RTTQA.
Nielsen et al100 gives recommendations for a comprehensive QC program and states a robust framework is essential for safety. They suggested that FMEA could be integrated into the clinical trial QC design which would allow QC protocols to be customized to the study, resources to be directed to QC failure modes that have the highest potential impact on trial outcomes, and identify processes that are most variable among centers to provide targeted support.
Kerns et al108 extended the value of dosimetry audits using E2E tests by producing an additional test for IROC thorax phantom audits. They created a beam model to facilitate independent recalculation of centers plans. A comparison of the center versus IROC plan calculation indicated if failing measurements were a result of calculation errors in the treatment planning system.
Discussion
A conventional technique to QC using published guidance documents maintains a minimum safe standard of equipment performance and aims to keep consistency across departments.34 However, this technique can have drawbacks such as time-consuming manual methods,15, 16 evaluating components in isolation,5, 14 performing outdated tests, and employing test tolerances and frequencies with no quantitative background, consideration of risk or age/workload of machine.36
These shortfalls can be addressed by employing the newer QC techniques discussed in this review. Automated and MIQC methods have been shown to require less time than conventional QC methods.44, 45 Researchers recommend E2E tests should be employed more routinely to consider the failure modes that exist when adding components together.11, 89 However, E2E tests often require additional testing to be developed108 or for all conventional QC to be performed11 to help identify the cause of error, which can increase a departments’ workload. A novel approach could be to design an E2E test that incorporates intermediate metrics to assess the stage at which any deviation could appear. Schreiner11 suggests implementation of FMEA to determine a combination of E2E and conventional testing that balances efficiency and error cause identification. FMEA has also been shown to be valuable in determining which QC tests should be carried out to provide optimal clinical benefit and their optimal tolerance and frequency.56, 59 AAPM task groups have produced reports which can be used by centers to help move toward FMEA and process-based radiotherapy. AAPM TG-10054 can be used to identify high-risk failure modes within the radiotherapy workflow and the draft TG-315109 produces a standardized framework for implementing the checks of these risks. Researchers have produced automated methods of implementing these checklists to further improve their practicality and efficiency.110
FMEA can also be used to determine optimal combinations of other QC techniques such as conventional methods and MIQC.61 As well as be used in conjunction with QC techniques such as SPC to determine a more appropriate output frequency for earlier detection of machine failure.36 SPC offers promise of being able to predict failures prior to occurrence, allow for preventative maintenance, and therefore reduced clinical disruption.63, 65 To improve QC efficiency is a common aim of research. For example, Hoisak et al111 developed a cloud-based system that queried the linacs machine log to record all instances and causes of machine breakdown with the aim of using the knowledge of what, when, and why components breakdown could produce a new QC and preventative maintenance schedule that improves on linac uptime, efficiency, and patient safety.
The use of AI increasingly looks beneficial to improving QC and PSQC efficiency. By using machine log files to predict machine performance,80 and previous PSQC results compared with complexity metrics70, 71 or MLC positions51–53, 76 to predict failing rates, improve plan optimization,52, 70 and reduce the need for measured results.51, 71, 72 However, there are concerns regarding the use of AI in radiotherapy, such as a lack of generalizability and lack of QC tools for the AI itself.112 Some studies address this, Yang et al113 improved generalizability by taking into account different delivery equipment, QC devices, and treatment planning systems. They also performed daily and monthly QC on the model with measurement consistency results within 2% and created a novel systematic evaluation of clinical deployment using an E2E test. The ethics of the use of AI regarding patient data, potential biases, reduced human oversight, and deskilling have also been raised. Yirmibesoglu Erkal et al114 used the 4 topics approach to evaluate the ethics of AI in radiotherapy. Overall, stating AI should only be used for decision support and not a replacement for clinical judgment. In the UK, the Medicines and Healthcare products Regulatory Agency launched a National Commission into the Regulations of AI in Healthcare in September 2025 with the goal to publish a new regulatory framework in 2026.115 This will provide centers with guidance when trying to implement AI into their practice.
Embracing a combination of techniques to create a new approach to QC practice could improve efficiency. Many studies have qualitatively quoted efficiency savings as a benefit of MIQC, FMEA, and AI,43–45, 56, 59, 111 with some detailing specific time savings seen for daily QC44, 45 and detailed added value by testing more modes in a similar time frame.45 Reduced time for QC would help alleviate clinical pressures and may improve staff well-being, for example, by providing the potential to move away from unsocial hours (only 12% of UK centers complete all their QC within a normal working day).34 This ideal approach could be a combination of techniques; potentially an E2E assessment that has individual optimized measurements of critical factors as indicated by FMEA analysis. The E2E system would have intermediate stages to identify potential failure points while also ensuring the whole system works together. All results would be analyzed and trended by AI. The system would be tested against ability to detect degrading performance, value, and efficiency.
Despite there being many recent publications covering QC techniques, it is worth noting that many of the ideas and justifications for the research are not novel. A critique of the conventional quality assurance paradigm was published by Thomadsen116 in 2008. This stated conventional QC procedures were time-consuming, a burden on departments and do not necessarily prevent patient injury, as human error is the more likely cause. Thomadsen116 was hopeful that new techniques such as the use of FMEA would address this problem. Thomadsen116 also queried whether measured PSQC was necessary, much like Barnes.13 Why have these techniques not yet become widespread? This could be due to many factors such as clinical pressures preventing time spent on a paradigm shift, a cautious approach resulting in a lack of desire to be the first to deviate from standard practice, a lack of technological maturity, or inherently they will never replace all conventional QC. Whatever the reason, a reevaluation of optimal QC methods may be overdue to ensure QC practices can keep pace with ever evolving delivery practices, and the need for improved efficiencies.
Conclusion
Different QC techniques have been discussed. Conventional QC methods are still widely used and useful, but often neglects system level QC, impact of human involvement, and is reactive not proactive or preventative. This increases risk and inefficiencies within the radiotherapy pathway. It is possible that the use of risk-based methods, SPC and E2E may help solve these issues, but these are themselves time-consuming. Introduction of automated QC and AI methods may help to reduce clinical downtime for QC which is attractive but may require further proof of concept and more rigorous testing before widespread adoption. Dosimetry audits remain a valuable addition to a QC program, fundamentally confirming delivered dose, to ensure standardized safe practice, and could benefit itself from the implementation of FMEA, automation, and AI.
Contributor Information
Elisha Tassano-Smith, Department of Medical Physics and Biomedical Engineering, University College London, London WC1E 6BT, United Kingdom; Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire PO6 3LY, United Kingdom.
Antony L Palmer, Department of Medical Physics and Biomedical Engineering, University College London, London WC1E 6BT, United Kingdom; Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire PO6 3LY, United Kingdom.
Wojciech Polak, Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire PO6 3LY, United Kingdom.
Andrew Nisbet, Department of Medical Physics and Biomedical Engineering, University College London, London WC1E 6BT, United Kingdom.
Conflicts of interest
No authors have any conflicts of interest.
Funding
No funding to declare.
References
- 1. Hunte SO, Clark CH, Zyuzikov N, Nisbet A. Volumetric modulated arc therapy (VMAT): a review of clinical outcomes—what is the clinical evidence for the most effective implementation? Br J Radiol. 2022;95:20201289. 10.1259/bjr.20201289 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Diez P, Hanna GG, Aitken KL, et al. UK 2022 consensus on normal tissue dose-volume constraints for oligometastatic, primary lung and hepatocellular carcinoma stereotactic ablative radiotherapy. Clin Oncol (R Coll Radiol). 2022;34:288-300. 10.1016/j.clon.2022.02.010 [DOI] [PubMed] [Google Scholar]
- 3. Yen A, Shen C, Albuquerque K. The new kid on the block: online adaptive radiotherapy in the treatment of gynecologic cancers. Curr Oncol. 2023;30:865-874. 10.3390/curroncol30010066 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Mast M, Perryck S. Introduction to: surface guided radiotherapy (SGRT). Tech Innov Patient Support Radiat Oncol. 2022;22:37-38. 10.1016/j.tipsro.2022.04.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Xiao Q, Li G. Application and challenges of statistical process control in radiation therapy quality assurance. Int J Radiat Oncol Biol Phys. 2024;118:295-305. 10.1016/j.ijrobp.2023.08.020 [DOI] [PubMed] [Google Scholar]
- 6. Mills JA, Colligan SJ. Megavoltage quality control, resources and demand: a pragmatic review. Br J Radiol. 2016;89:20150709. 10.1259/bjr.20150709 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Institute of Physics and Engineering in Medicine (IPEM). The Impact of Extended Clinical Hours on Radiotherapy Physics Service. IPEM; 2020. https://www.ipem.ac.uk/media/vpjfvrgu/extended-hours-rt.pdf [Google Scholar]
- 8. American Society for Radiation Oncology (ASTRO). Safety is No Accident: A Framework for Quality Radiation Oncology Care. 2nd ed. American Society for Radiation Oncology; 2019. [Google Scholar]
- 9. British Institute of Radiology, Institute of Physics and Engineering in Medicine, National Patient Safety Agency, Society and College of Radiographers, The Royal College of Radiologists. Towards Safer Radiotherapy. British Institute of Radiology; 2008. [Google Scholar]
- 10. Health and Safety Executive. The Ionising Radiation (Medical Exposure) Regulations 2017. Health and Safety Executive; 2017. [Google Scholar]
- 11. Schreiner LJ. End-to-end QA in radiation therapy quality management. J Phys Conf Ser. 2023;2630:012007. 10.1088/1742-6596/2630/1/012007 [DOI] [Google Scholar]
- 12. Ono T, Iramina H, Hirashima H, Adachi T, Nakamura M, Mizowaki T. Applications of artificial intelligence for machine- and patient-specific quality assurance in radiation therapy: current status and future directions. J Radiat Res. 2024;65:421-432. 10.1093/jrr/rrae033 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Barnes M. A pathway away from measurement-based patient-specific QA: a review of contemporary quality assurance for external-beam radiotherapy Sub-systems. J Appl Clin Med Phys. 2025;26:e70407. 10.1002/acm2.70407 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Kron T, Fox C, Ebert MA, Thwaites D. Quality management in radiotherapy treatment delivery. J Med Imaging Radiat Oncol. 2022;66:279-290. 10.1111/1754-9485.13348 [DOI] [PubMed] [Google Scholar]
- 15. Zhou S, Li J, Du Y, et al. Development and longitudinal analysis of plan-based streamlined quality assurance on multiple positioning guidance systems with single phantom setup. Front Oncol. 2021;11:683733. 10.3389/fonc.2021.683733 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Palta JR, Liu C, Li JG. Current external beam radiation therapy quality assurance guidance: does it meet the challenges of emerging image-guided technologies? Int J Radiat Oncol Biol Phys. 2008;71:S13-S17. 10.1016/j.ijrobp.2007.06.084 [DOI] [PubMed] [Google Scholar]
- 17. Pearson M, Barnes MP, Brown KF, et al. IPEM topical report: results of a 2022 UK survey on the use of linac manufacturer integrated quality control (MIQC). Phys Med Biol. 2023;68:245018. 10.1088/1361-6560/ad0eb3 [DOI] [PubMed] [Google Scholar]
- 18. Pearson M, Barnes MP, Brown KF, et al. IPEM topical report: guidance for the use of linac manufacturer integrated quality control. Phys Med Biol. 2025;70:085016. 10.1088/1361-6560/adc8f7 [DOI] [PubMed] [Google Scholar]
- 19. AAPM Task Group. Report 332: Verification of Vendor‐Provided Data, Tools, and Test Procedures in Radiotherapy. 2016. https://aapm.onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17879 10.1002/mp.17879 [DOI]
- 20. Pourel N, Meyrieux C, Lisbona A. Quality/risk management system in radiotherapy: changes afoot. Cancer Radiother J Soc Francaise Radiother Oncol. 2022;26:14-19. 10.1016/j.canrad.2021.10.004 [DOI] [PubMed] [Google Scholar]
- 21. Ford EC, Gaudette R, Myers L, et al. Evaluation of safety in a radiation oncology setting using failure mode and effects analysis. Int J Radiat Oncol Biol Phys. 2009;74:852-858. 10.1016/j.ijrobp.2008.10.038 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Wong LM, Pawlicki T. A systems theory-based safety assessment of pre-treatment patient-specific quality assurance for intensity-modulated treatments in a single-vendor environment. Radiother Oncol. 2024;201:110569. 10.1016/j.radonc.2024.110569 [DOI] [PubMed] [Google Scholar]
- 23. O’Daniel JC, Giles W, Cui Y, Adamson J. A structured FMEA approach to optimizing combinations of plan-specific quality assurance techniques for IMRT and VMAT QA. Med Phys. 2023;50:5387-5397. 10.1002/mp.16630 [DOI] [PubMed] [Google Scholar]
- 24. Palmer A, Mzenda B, Kearton J, Wills R. Analysis of regional radiotherapy dosimetry audit data and recommendations for future audits. Br J Radiol. 2011;84:733-742. 10.1259/bjr/18691638 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Clark CH, Aird EGA, Bolton S, et al. Radiotherapy dosimetry audit: three decades of improving standards and accuracy in UK clinical practice and trials. Br J Radiol. 2015;88:20150251. 10.1259/bjr.20150251 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Patel I, Weston S, Palmer AL. Physics Aspects of Quality Control in Radiotherapy - IPEM Report 81. 2nd ed. Institute of Physics and Engineering in Medicine; 2018. [Google Scholar]
- 27. Klein EE, Hanley J, Bayouth J, et al. ; Task Group 142, American Association of Physicists in Medicine. Task group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197-4212. 10.1118/1.3190392 [DOI] [PubMed] [Google Scholar]
- 28. Smith K, Balter P, Duhon J, et al. AAPM medical physics practice guideline 8.a.: linear accelerator performance tests. J Appl Clin Med Phys. 2017;18:23-39. 10.1002/acm2.12080 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Hanley J, Dresser S, Simon W, et al. AAPM task group 198 report: an implementation guide for TG 142 quality assurance of medical accelerators. Med Phys. 2021;48:e830-e885. 10.1002/mp.14992 [DOI] [PubMed] [Google Scholar]
- 30. Krauss RF, Balik S, Cirino ET, et al. AAPM medical physics practice guideline 8.b: linear accelerator performance tests. J Appl Clin Med Phys. 2023;24:e14160. 10.1002/acm2.14160 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Van Der Wal E, Wiersma J, Ausma AH, NCS Report 22: Code of Practice for the Quality Assurance and Control for Intensity Modulated Radiotherapy. 1st ed. Netherlands Commission for Radiation Dosimetry; 2013. 10.25030/ncs-022. [DOI] [Google Scholar]
- 32. Mans A, Schuring D, Arends M, NCS Report 24: Code of Practice for the Quality Assurance and Control for Volumetric Modulated Arc Therapy. 1st ed. Netherlands Commission for Radiation Dosimetry; 2015. 10.25030/ncs-024. [DOI] [Google Scholar]
- 33. Fog LS, Webb LK, Barber J, et al. ; ACPSEM Radiation Oncology Specialty Group (ROSG). ACPSEM position paper: pre-treatment patient specific plan checks and quality assurance in radiation oncology. Phys Eng Sci Med. 2024;47:7-15. 10.1007/s13246-023-01367-9 [DOI] [PubMed] [Google Scholar]
- 34. Tassano-Smith E, Palmer AL, Polak W, Nisbet A. A survey and analysis of radiotherapy linear accelerator quality control in the United Kingdom, for c-arm and ring gantry systems and associated patient specific quality control. Br J Radiol. 2025;98:1463-1482. 10.1093/bjr/tqaf140 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Raveendran V, R GR, P T A, Bhasi S, C P R, Kinhikar RA. Moving towards process-based radiotherapy quality assurance using statistical process control. Phys Med. 2023;112:102651. 10.1016/j.ejmp.2023.102651 [DOI] [PubMed] [Google Scholar]
- 36. Ma M, Men K, Dai J. A patient risk model to determine the optimal output constancy check frequency for a radiotherapy machine. Phys Med. 2021;84:192-197. 10.1016/j.ejmp.2021.04.013 [DOI] [PubMed] [Google Scholar]
- 37. Clivio A, Vanetti E, Rose S, et al. Evaluation of the machine performance check application for TrueBeam linac. Radiat Oncol. 2015;10:97. 10.1186/s13014-015-0381-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Barnes MP, Greer PB. Evaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter‐free (FFF) photon beams. J Appl Clin Med Phys. 2017;18:139-150. 10.1002/acm2.12016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Barnes MP, Greer PB. Evaluation of the truebeam machine performance check (MPC): mechanical and collimation checks. J Appl Clin Med Phys. 2017;18:56-66. 10.1002/acm2.12072 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Barnes MP, Greer PB. Evaluation of the truebeam machine performance check (MPC) geometric checks for daily IGRT geometric accuracy quality assurance. J Appl Clin Med Phys. 2017;18:200-206. 10.1002/acm2.12064 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Pearson M, Eaton D, Greener T. Long‐term experience of MPC across multiple TrueBeam linacs: MPC concordance with conventional QC and sensitivity to real‐world faults. J Appl Clin Med Phys. 2020;21:224-235. 10.1002/acm2.12950 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Binny D, Aland T, Archibald‐Heeren BR, Trapp JV, Kairn T, Crowe SB. A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control. J Appl Clin Med Phys. 2019;20:71-80. 10.1002/acm2.12547 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Stambaugh C, Yancey J, Shukla U, Melhus C, Stambaugh N. Daily quality assurance efficiency evaluation using SunCHECK machine and machine performance check. Cureus. 2023;15:e35695. 10.7759/cureus.35695 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Dhoundiyal M, Rasal S, Gupte A, Dandekar PR, Jadhav A, Awate O. Validation and efficiency evaluation of automated quality assurance software SunCHECKTM machine for mechanical and dosimetric quality assurance. J Med Phys. 2024;49:311-315. 10.4103/jmp.jmp_158_23 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Bonanno E, Borzì GR, Cavalli N, et al. Use of an automated software module for monthly routine machine QA tests. J Instrum. 2023;18:T07010. 10.1088/1748-0221/18/07/T07010 [DOI] [Google Scholar]
- 46. Qian J, Lee L, Liu W, et al. Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files. Phys Med Biol. 2010;55:3597-3610. 10.1088/0031-9155/55/13/002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Defoor DL, Stathakis S, Roring JE, et al. Investigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methods. J Appl Clin Med Phys. 2017;18:172-179. 10.1002/acm2.12114 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48. Rangaraj D, Zhu M, Yang D, et al. Catching errors with patient-specific pretreatment machine log file analysis. Pract Radiat Oncol. 2013;3:80-90. 10.1016/j.prro.2012.05.002 [DOI] [PubMed] [Google Scholar]
- 49. Son NK, Phat LT, Tu HA, et al. Evaluation of multi-leaf collimator accuracy in halcyon system using logfile analysis. Radiat Phys Chem. 2025;236:112741. 10.1016/j.radphyschem.2025.112741 [DOI] [Google Scholar]
- 50. Wojtasik AM, Bolt M, Clark CH, Nisbet A, Chen T. Multivariate log file analysis for multi-leaf collimator failure prediction in radiotherapy delivery. Phys Imaging Radiat Oncol. 2020;15:72-76. 10.1016/j.phro.2020.07.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Lv J, Chen L, Zhu Z, et al. ; FDS Consortium. Advanced prediction of multi‐leaf collimator leaf position using artificial neural network. Med Phys. 2025;52:3439-3449. 10.1002/mp.17690 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52. Carlson JNK, Park JM, Park SY, Park JI, Choi Y, Ye SJ. A machine learning approach to the accurate prediction of multi-leaf collimator positional errors. Phys Med Biol. 2016;61:2514-2531. 10.1088/0031-9155/61/6/2514 [DOI] [PubMed] [Google Scholar]
- 53. Sivabhaskar S, Li R, Roy A, Kirby N, Fakhreddine M, Papanikolaou N. Machine learning models to predict the delivered positions of elekta multileaf collimator leaves for volumetric modulated arc therapy. J Appl Clin Med Phys. 2022;23:e13667. 10.1002/acm2.13667 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54. Huq MS, Fraass BA, Dunscombe PB, et al. The report of task group 100 of the AAPM: application of risk analysis methods to radiation therapy quality management. Med Phys. 2016;43:4209. 10.1118/1.4947547 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Liu HC, Zhang LJ, Ping YJ, Wang L. Failure mode and effects analysis for proactive healthcare risk evaluation: a systematic literature review. J Eval Clin Pract. 2020;26:1320-1337. 10.1111/jep.13317 [DOI] [PubMed] [Google Scholar]
- 56. Ochi Y, Saito A, Kawahara D, et al. A novel risk analysis of clinical reference dosimetry based on failure modes and effects analysis. Phys Med. 2019;58:59-65. 10.1016/j.ejmp.2019.01.014 [DOI] [PubMed] [Google Scholar]
- 57. Rah JE, Manger RP, Yock AD, Kim GY. A comparison of two prospective risk analysis methods: traditional FMEA and a modified healthcare FMEA. Med Phys. 2016;43:6347. 10.1118/1.4966129 [DOI] [PubMed] [Google Scholar]
- 58. Kawahara D, Nakano H, Saito A, Ochi Y, Nagata Y. Formulation of objective indices to quantify machine failure risk analysis for interruptions in radiotherapy. J Appl Clin Med Phys. 2021;22:165-173. 10.1002/acm2.13126 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59. O’Daniel JC, Yin FF. Quantitative approach to failure mode and effect analysis for linear accelerator quality assurance. Int J Radiat Oncol. 2017;98:56-62. 10.1016/j.ijrobp.2017.01.035 [DOI] [PubMed] [Google Scholar]
- 60. Bonfantini F, Giandini T, Meroni S, et al. Application of failure mode and effects analysis to optimization of linac quality controls protocol. Med Phys. 2019;46:2541-2555. 10.1002/mp.13538 [DOI] [PubMed] [Google Scholar]
- 61. Pearson M, Butterworth V, Misson-Yates S, et al. Application of failure mode and effects analysis to validate a novel hybrid linac QC program that integrates automated and conventional QC testing. J Appl Clin Med Phys. 2022;23:e13798. 10.1002/acm2.13798 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. De Kerf G, Barragán-Montero A, Brouwer CL, et al. Multicentre prospective risk analysis of a fully automated radiotherapy workflow. Phys Imaging Radiat Oncol. 2025;34:100765. 10.1016/j.phro.2025.100765 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Ma M, Li M, Zhang K, et al. Applying the six-sigma methodology to determine the limits of quality control (QC) tests for a specific linear accelerator. J Appl Clin Med Phys. 2024;25:e14460. 10.1002/acm2.14460 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64. NHS Improvement. An Overview of Six Sigma. NHS England; 2011. https://www.england.nhs.uk/improvement-hub/wp-content/uploads/sites/44/2018/06/An-overview-of-Six-Sigma.pdf [Google Scholar]
- 65. Able CM, Baydush AH, Nguyen C, et al. A model for preemptive maintenance of medical linear accelerators—predictive maintenance. Radiat Oncol. 2016;11:36. 10.1186/s13014-016-0602-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66. Shen C, Nguyen D, Zhou Z, Jiang SB, Dong B, Jia X. An introduction to deep learning in medical physics: advantages, potential, and challenges. Phys Med Biol. 2020;65:05TR01. 10.1088/1361-6560/ab6f51 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67. Santoro M, Strolin S, Paolani G, et al. Recent applications of artificial intelligence in radiotherapy: where we are and beyond. Appl Sci. 2022;12:3223. 10.3390/app12073223 [DOI] [Google Scholar]
- 68. Avanzo M, Trianni A, Botta F, Talamonti C, Stasi M, Iori M. Artificial intelligence and the medical physicist: welcome to the machine. Appl Sci. 2021;11:1691. 10.3390/app11041691 [DOI] [Google Scholar]
- 69. Osman AFI, Maalej NM. Applications of machine and deep learning to patient-specific IMRT/VMAT quality assurance. J Appl Clin Med Phys. 2021;22:20-36. 10.1002/acm2.13375 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Wall PDH, Fontenot JD. Application and comparison of machine learning models for predicting quality assurance outcomes in radiation therapy treatment planning. Inform Med Unlocked. 2020;18:100292. 10.1016/j.imu.2020.100292 [DOI] [Google Scholar]
- 71. Wall PDH, Hirata E, Morin O, Valdes G, Witztum A. Prospective clinical validation of virtual patient-specific quality assurance of volumetric modulated arc therapy radiation therapy plans. Int J Radiat Oncol Biol Phys. 2022;113:1091-1102. 10.1016/j.ijrobp.2022.04.040 [DOI] [PubMed] [Google Scholar]
- 72. Yang X, Li S, Shao Q, Cao Y, Yang Z, Zhao YQ. Uncertainty-guided man-machine integrated patient-specific quality assurance. Radiother Oncol J Eur Soc Ther Radiol Oncol. 2022;173:1-9. 10.1016/j.radonc.2022.05.016 [DOI] [PubMed] [Google Scholar]
- 73. Zeng X, Zhu Q, Ahmed A, et al. Multi-granularity prior networks for uncertainty-informed patient-specific quality assurance. Comput Biol Med. 2024;179:108925. 10.1016/j.compbiomed.2024.108925 [DOI] [PubMed] [Google Scholar]
- 74. Wang Y, Pang X, Wang H, Bai Y. Research on the error pattern recognition of dosimetric quality assurance by Bayesian optimization. Transl Cancer Res. 2025;14:2029-2042. 10.21037/tcr-2025-337 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Luk SMH, Meyer J, Young LA, et al. Characterization of a Bayesian network-based radiotherapy plan verification model. Med Phys. 2019;46:2006-2014. 10.1002/mp.13515 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Osman AFI, Maalej NM, Jayesh K. Prediction of the individual multileaf collimator positional deviations during dynamic IMRT delivery priori with artificial neural network. Med Phys. 2020;47:1421-1430. 10.1002/mp.14014 [DOI] [PubMed] [Google Scholar]
- 77. McNutt TR, Moore KL, Wu B, Wright JL. Use of big data for quality assurance in radiation therapy. Semin Radiat Oncol. 2019;29:326-332. 10.1016/j.semradonc.2019.05.006 [DOI] [PubMed] [Google Scholar]
- 78. Wu B, Zhang P, Tsirakis B, Kanchaveli D, LoSasso T. Utilizing historical MLC performance data from trajectory logs and service reports to establish a proactive maintenance model for minimizing treatment disruptions. Med Phys. 2019;46:475-483. 10.1002/mp.13363 [DOI] [PubMed] [Google Scholar]
- 79. Kalet AM, Luk SMH, Phillips MH. Radiation therapy quality assurance tasks and tools: the many roles of machine learning. Med Phys. 2020;47:e168-e177. 10.1002/mp.13445 [DOI] [PubMed] [Google Scholar]
- 80. Ma M, Liu C, Wei R, Liang B, Dai J. Predicting machine’s performance record using the stacked long short‐term memory (LSTM) neural networks. J Appl Clin Med Phys. 2022;23:e13558. 10.1002/acm2.13558 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81. Mehrens H, Kry S. 3817 Predicting IROC’s thoracic dosimetry audit performance using random forest modeling and feature interpretability. Radiother Oncol. 2025;206:S3677-S3678. 10.1016/S0167-8140(25)01978-4 [DOI] [Google Scholar]
- 82. Mertens L, Fleckenstein J, Steil V, Schneider F. A novel end-to-end test for combined dosimetric and geometric treatment verification using a 3D-printed phantom. Med Dosim Off J Am Assoc Med Dosim. 2022;47:177-183. 10.1016/j.meddos.2022.02.002 [DOI] [PubMed] [Google Scholar]
- 83. Tajik M, Akhlaqi MM, Gholami S. Advances in anthropomorphic thorax phantoms for radiotherapy: a review. Biomed Phys Eng Express. 2022;8:ac369c. 10.1088/2057-1976/ac369c [DOI] [PubMed] [Google Scholar]
- 84. Shiinoki T, Fujii F, Fujimoto K, Yuasa Y, Sera T. A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy. J Appl Clin Med Phys. 2020;21:16-28. 10.1002/acm2.12876 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85. Burton A, Beveridge S, Hardcastle N, Malfitana S, Madamperuma J, Franich R. Design and validation of a novel dosimetry phantom for motion management audits. J Appl Clin Med Phys. 2025;26:e70091. 10.1002/acm2.70091 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86. Shariff M, Grigo J, Masitho S, et al. End-to-end testing for stereotactic radiotherapy including the development of a multi-modality phantom. Z Med Phys. 2024;34:477-484. 10.1016/j.zemedi.2022.11.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87. Hayashi N, Saito T, Ouchi K, et al. Development of dedicated phantom for end-to-end testing of multiple image-guidance configurations incorporating light-section-based surface guidance system. J Appl Clin Med Phys. 2025;26:e14517. 10.1002/acm2.14517 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88. Zhang F, Zhang H, Zhao H, et al. Design and fabrication of a personalized anthropomorphic phantom using 3D printing and tissue equivalent materials. Quant Imaging Med Surg. 2019;9:94-100. 10.21037/qims.2018.08.01 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Loughery B, Knill C, Silverstein E, et al. Multi-institutional evaluation of end-to-end protocol for IMRT/VMAT treatment chains utilizing conventional linacs. Med Dosim Off J Am Assoc Med Dosim. 2019;44:61-66. 10.1016/j.meddos.2018.02.002 [DOI] [PubMed] [Google Scholar]
- 90. Kry SF, Molineu A, Kerns JR, et al. Institutional patient-specific IMRT QA does not predict unacceptable plan delivery. Int J Radiat Oncol Biol Phys. 2014;90:1195-1201. 10.1016/j.ijrobp.2014.08.334 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91. Lehmann J, Hussein M, Barry MA, et al. SEAFARER - a new concept for validating radiotherapy patient specific QA for clinical trials and clinical practice. Radiother Oncol J Eur Soc Ther Radiol Oncol. 2022;171:121-128. 10.1016/j.radonc.2022.04.019 [DOI] [PubMed] [Google Scholar]
- 92. Barber J, Vial P, White P, et al. A survey of modulated radiotherapy use in Australia and New Zealand in 2015. Australas Phys Eng Sci Med. 2017;40:811-822. 10.1007/s13246-017-0590-y [DOI] [PubMed] [Google Scholar]
- 93. Abolaban F, Zaman S, Cashmore J, Nisbet A, Clark CH. Changes in patterns of intensity-modulated radiotherapy verification and quality assurance in the UK. Clin Oncol (R Coll Radiol). 2016;28:e28-e34. 10.1016/j.clon.2016.01.013 [DOI] [PubMed] [Google Scholar]
- 94. O’Daniel J, Hernandez V, Clark C, et al. Which failures do patient-specific quality assurance systems need to catch? Med Phys. 2025;52:88-98. 10.1002/mp.17468 [DOI] [PubMed] [Google Scholar]
- 95. Misson S, Palmer AL, Diez P, Hussein M, Clark CH. National radiotherapy dosimetry audit in the UK - a vision and roadmap. IPEM-Transl. 2024;12:100032. 10.1016/j.ipemt.2024.100032 [DOI] [Google Scholar]
- 96. Lye J, Kry S, Shaw M, et al. A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry. Med Phys. 2019;46:5878-5887. 10.1002/mp.13800 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97. Vaandering A, Roels S, Yalvaç B, et al. Favouring quality improvement initiatives: the experience of the Belgian college of radiation oncology. Precis Cancer Med. 2023;6:4-4. 10.21037/pcm-22-15 [DOI] [Google Scholar]
- 98. Izewska J, Bokulic T, Kazantsev P, Wesolowska P, van der Merwe D. 50 Years of the IAEA/WHO postal dose audit programme for radiotherapy: what can we learn from 13756 results? Acta Oncol. 2020;59:495-502. 10.1080/0284186X.2020.1723162 [DOI] [PubMed] [Google Scholar]
- 99. Nishio T, Nakamura M, Okamoto H, et al. An overview of the medical-physics-related verification system for radiotherapy multicenter clinical trials by the medical physics working group in the Japan clinical oncology Group-Radiation therapy study group. J Radiat Res. 2020;61:999-1008. 10.1093/jrr/rraa089 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100. Nielsen CP, Samsøe E, Offersen BV, et al. Recommendations for radiotherapy quality assurance in clinical trials. Radiother Oncol. 2025;209:110950. 10.1016/j.radonc.2025.110950 [DOI] [PubMed] [Google Scholar]
- 101. Ibbott GS, Thwaites DI. Audits for advanced treatment dosimetry. J Phys Conf Ser. 2015;573:012002. 10.1088/1742-6596/573/1/012002 [DOI] [Google Scholar]
- 102. Muhammad W, Ullah A, Mahmood K, Null M. Assessment of national dosimetry quality audits results for teletherapy machines from 1989 to 2015. J Appl Clin Med Phys. 2016;17:145-152. 10.1120/jacmp.v17i2.5984 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103. Pasler M, Hernandez V, Jornet N, Clark CH. Novel methodologies for dosimetry audits: adapting to advanced radiotherapy techniques. Phys Imaging Radiat Oncol. 2018;5:76-84. 10.1016/j.phro.2018.03.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104. Brooks C, Miles E, Hoskin PJ. Radiotherapy trial quality assurance processes: a systematic review. Lancet Oncol. 2024;25:e104-e113. 10.1016/S1470-2045(23)00625-3 [DOI] [PubMed] [Google Scholar]
- 105. Brooks FMD, Hussein M, Lye J, et al. Developing reference plans for evaluating global clinical trials credentialing and PSQA systems. J Appl Clin Med Phys. 2025;26:e70113. 10.1002/acm2.70113 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106. Global Quality Assurance of Radiation Therapy Clinical Trials Harmonization Group. Global Quality Assurance of Radiation Therapy Clinical Trials Harmonization Group. 2026. https://rtqaharmonization.org/
- 107. Jomy J, Sharma R, Lu R, et al. Clinical impact of radiotherapy quality assurance results in contemporary cancer trials: a systematic review and meta-analysis. Radiother Oncol. 2025;207:110875. 10.1016/j.radonc.2025.110875 [DOI] [PubMed] [Google Scholar]
- 108. Kerns JR, Stingo F, Followill D, Howell R, Melancon A, Kry SF. Treatment planning system calculation errors are present in the majority of IROC-Houston phantom failures. Int J Radiat Oncol Biol Phys. 2017;98:1197-1203. 10.1016/j.ijrobp.2017.03.049 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109. Xia P, Chuang C, Colussi VC, et al. Task Group 315: Medical Physics Practice Guideline (MPPG) For Plan and Chart Review in External Beam Radiotherapy and Brachytherapy. 2026. https://www.aapm.org/announcements/MPPGPlanChartChecksDraft.pdf [DOI] [PMC free article] [PubMed]
- 110. Xu H, Zhang B, Guerrero M, et al. Toward automation of initial chart check for photon/electron EBRT: the clinical implementation of new AAPM task group reports and automation techniques. J Appl Clin Med Phys. 2021;22:234-245. 10.1002/acm2.13200 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111. Hoisak JD, Kim GYG, Atwood TF, Pawlicki T. Operational insights from the longitudinal analysis of a linear accelerator machine log. Cureus. 2021;13:e16038. 10.7759/cureus.16038 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112. Vandewinckele L, Claessens M, Dinkla A, et al. Overview of artificial intelligence-based applications in radiotherapy: recommendations for implementation and quality assurance. Radiother Oncol. 2020;153:55-66. 10.1016/j.radonc.2020.09.008 [DOI] [PubMed] [Google Scholar]
- 113. Yang R, Yang X, Wang L, et al. Commissioning and clinical implementation of an autoencoder based classification-regression model for VMAT patient-specific QA in a multi-institution scenario. Radiother Oncol J Eur Soc Ther Radiol Oncol. 2021;161:230-240. 10.1016/j.radonc.2021.06.024 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114. Yirmibesoglu Erkal E, Akpınar A, Erkal HŞ. Ethical evaluation of artificial intelligence applications in radiotherapy using the four topics approach. Artif Intell Med. 2021;115:102055. 10.1016/j.artmed.2021.102055 [DOI] [PubMed] [Google Scholar]
- 115. GOV.UK. National Commission into the Regulation of AI in Healthcare. 2026. https://www.gov.uk/government/groups/national-commission-into-the-regulation-of-ai-in-healthcare
- 116. Thomadsen B. Critique of traditional quality assurance paradigm. Int J Radiat Oncol Biol Phys. 2008;71:S166-S169. 10.1016/j.ijrobp.2007.07.2391 [DOI] [PubMed] [Google Scholar]