Abstract
Reliable processing of biospecimens such as plasma in liquid biopsy is crucial to achieving robust analytical results; however, manual procedures involved in sample preparation remain a major source of variability. Here, we present MECHA (Magnetic-Enrichment and Centrifugal-Homogenization Assembly), a fully automated system that standardizes magnetic bead (MB)-based sample handling. MECHA integrates coordinated magnetic actuation, controlled vibration, and orbital mixing to perform target capture, washing, and labeling steps with minimal operator input. The prototype was designed to process eight samples in parallel for electrochemiluminescence (ECL) detection. The signal variation was <10% across replicates. In a pilot analytical study, the MECHA-ECL pipeline enabled quantitative profiling of extracellular vesicles (EVs) from clinical plasma samples of colorectal cancer patients. The automated EV assay achieved a detection limit of 4 × 104 EVs/mL, a dynamic range spanning 4 orders of magnitude, and identified EGFR and CD133 as predictive biomarkers for chemotherapy response.
INTRODUCTION
Processing biospecimens, such as biofluids in liquid biopsy, is a crucial preanalytical factor that influences the reliability of clinical assays1,2. Inconsistencies or errors introduced at this stage can compromise downstream analytical results and affect diagnostic accuracy. Manual sample preparation is particularly susceptible to inter- and intra-operator variability, limiting assay throughput. Automating sample processing can help improve the consistency and efficiency of diagnostic workflows3-5.
Electrochemiluminescence (ECL) is a potent biosensing modality6-8. Its signaling mechanism, chemiluminescence triggered by electrical potential, offers advantages including high signal-to-background contrast, electrical control over reaction kinetics, and straightforward instrumentation compared to other assay platforms. Such features make ECL assays well-suited for sensitive, high-throughput analysis. However, most ECL assays rely on manual sample preparation or require sophisticated instrumentation for automation, which limits their broader implementation in research and clinical laboratories. Conveniently, ECL assays are well-suited for adaptation to magnetic bead (MB)-based formats, in which MBs serve as a solid support for target capture and labeling9-12. This configuration is suitable for automation, as MBs can be manipulated by magnetic forces for capture, washing, and elution, and uniformly dispersed by vortexing.
Here, we introduce MECHA (Magnetic-Enrichment and Centrifugal-Homogenization Assembly), an integrated device designed to automate MB–based sample processing. MECHA features three core capabilities: magnetic manipulation for bead capture, transport, and elution; sonic vibration for bead release and dispersion; and spill-free vortexing for sample mixing. A working prototype was implemented to process up to eight samples in parallel, completing target capture, washing, and ECL labeling within 15 minutes, with signal variation of less than 10% across all samples. To demonstrate its clinical applicability, we applied MECHA to process plasma samples from patients with colorectal cancer (CRC). Extracellular vesicles (EVs), biological nanoparticles secreted by cells, were selected as the analytical target due to their established potential as cancer biomarkers13-17. We utilized MECHA to immunocapture EVs directly from plasma and subsequently label them for a panel of CRC-candidate markers (EGFR, EpCAM, GPA33, MDR, CEA, and CD133). A custom-designed ECL detection chip was then used to concurrently detect eight protein markers. The results revealed a marked reduction in CRC EV burden following curative surgery and identified candidate markers predictive of response to neoadjuvant chemotherapy.
RESULTS
Overall design architecture of MECHA
Figure 1a shows a standard workflow of ECL assay. The process begins by mixing a sample with target-specific magnetic beads (MBs) to capture the analytes of interest. The resulting MB-target complexes are then labeled with probe antibodies conjugated to the ECL luminophore [Ru(bpy)3]2+. These labeled MBs are deposited onto an electrode together with the coreactant tripropylamine (TPrA) for signal generation (Figure. 1b). Upon applying a potential between the working and reference electrodes, both [Ru(bpy)3]2+ and TPrA are oxidized, generating the highly oxidizing species [Ru(bpy)3]3+ and the strongly reducing radical TPrA*. Their subsequent redox reaction forms the excited-state [Ru(bpy)3]2+*, which emits light10,18,19. This MB-based assay can be directly applied to native samples such as plasma or urine without purification, achieving a rapid turnaround time of ~1 hour. Furthermore, MBs can be magnetically concentrated on the electrode surface to enhance signal intensity20,21.
Figure 1. Automated ECL assay workflow using MECHA.

(a) Overview of the ECL assay. Samples are incubated with target-specific magnetic beads (MBs) for analyte capture, followed by labeling with [Ru(bpy)3]2+-conjugated antibodies. (b) Labeled MBs are deposited on an electrode together with the coreactant tripropylamine (TPrA). Upon application of an electrical potential, light is emitted through the ECL reaction. (c) Sequential steps in MECHA sample processing. A linear array of wells is preloaded with reagents and buffers. Magnetic rods with removable sheaths capture and transport MBs; bead release is aided by sheath vibration, while vortexing ensures thorough mixing of MBs and reagents. PBS, phosphate-buffered saline; PBST, PBS with Tween 20. (d) Schematic (left) and photograph of the MECHA prototype (right). The system is compatible with 96-well plates, enabling the concurrent processing of 8 samples. It features eight magnetic rods, an orbital shaker, and a touchscreen interface within a compact benchtop device. (e) Custom-designed ECL chip for parallel eight-sample measurement. The chip consists of eight working electrodes with shared counter and reference electrodes. (f) ECL measurement setup. The chip is mounted on a socket containing pogo pins for electrical contacts and an embedded magnet array for MB collection. An EM-CCD camera captures images of working electrodes, enabling parallel quantitative readout.
We designed the MECHA system to automate sample preparation, including target capture, labeling, mixing, and washing. The system design mimicked the manual workflow by adopting a linear-well format (Figure. 1c): necessary reagents (e.g., MBs, antibodies) and wash buffers were preloaded in separate wells, and the MBs were sequentially transferred between these wells. The bead transfer was executed using a magnet rod encased in a removable plastic sheath. This rod captured MBs in one well and transported them to the next. MBs were released by retracting the magnet rod from the sheath, assisted by sonic agitation. Sample mixing within the wells was achieved by vortexing the entire well structure.
We constructed a prototype (Figure. 1d) compatible with a conventional 96-well plate (12 columns × 8 rows). Each row on the plate constituted the linear well path for a separate sample, enabling the system to run eight concurrent assays. To support this operation, an array of eight magnetic rods was installed on a mechanical actuator, and the plate holder was designed to perform an orbital rotation for sample mixing. The system featured a small form factor (w x d x h (in cm) = 33 × 44 × 26 ] as a benchtop machine. It included a touchscreen user interface and could run user-defined command scripts for customized workflows.
Matching the MECHA's throughput, we built an ECL setup for eight parallel measurements. An ECL chip was fabricated, comprising eight working electrodes, a shared counter electrode, and a shared reference electrode (Figure. 1e). This chip was mounted onto a custom socket (Figure. 1f) that incorporated pogo pins for electrical contacts and an embedded magnet array to concentrate MBs onto the working electrodes. During the ECL reaction, the chip was imaged using an electron-multiplying charge-coupled device (EMCCD) camera, and the luminescence intensity from each working electrode was extracted for quantitative analysis.
Design of the magnetic manipulators
Four types of MB manipulations are needed for automated sample preparation, i.e. bead capture, transport, release, and dispersion. To execute these functions, we designed and integrated (Figure. 2a), a magnetic manipulator, a plate translator, and an orbital shaker into MECHA (see Figure. S1 and Data S1-S4 for details).
Figure 2. Design and operation of the MECHA system.

(a) Three core modules in the MECHA architecture: a magnetic manipulator, a plate translator, and an orbital shaker. The magnetic manipulator consists of vertically aligned magnet pairs and a removable plastic sheath, with each component controlled by a separate motorized stage for z-axis movement. The plate stage moves samples in the x-direction. Positioned beneath the plate stage, the orbital shaker generates controlled circular motion to ensure uniform sample mixing. (b) A vibration module was integrated into the sheath stage to facilitate MB detachment. (c) Stepwise sequence of MB transfer as described in the main text. (d) Design schematic of the orbital shaker. The motor and the camshafts were arranged to drive a controlled circular motion. (e) Computational fluid dynamics simulations of orbital shaking. Three hydrodynamic regimes were identified: Regime I (low speed) with weak bottom-confined circulation; Regime II (intermediate speed) characterized by a toroidal vortex spanning most of the liquid volume; and Regime III (high speed) showing pronounced free-surface deformation with spillover risk. (f) Experimental validation of MB dispersion under orbital shaking. Representative images show bead distribution before and after 1 min of orbital shaking at increasing rotational speeds. Uniform dispersion throughout the well volume was achieved at 800 rpm.
The magnetic manipulator consisted of eight magnet assemblies, each comprising a vertically aligned pair of magnets. The upper magnet was rod-shaped, encased in a detachable plastic sheath (Figure. S2 and Plastic sheath.STL for plastic sheath); this assembly could be immersed in a sample to collect MBs. The lower magnet was positioned beneath each well to attract MBs downward (Bottom magnet holder.STL). Both magnets and the sheath were mounted on separate motorized stages, which independently operated along the z-axis (as defined in Figure. 2a, and PS holder part1&2.STL, Top magnet holder.STL for assembly of top magnet control module). The sheath stage also included a vibration motor that generated oscillatory motion (Figure. 2b). For lateral movement, the 96-well plate was mounted on another motorized stage that moved along the x-axis; this stage also served as an orbital shaker (Plate bed.STL, Shaking plate.STL, Cam 1 & 2.STL). All parts were assembled onto the MECHA base (Base plate.STL and Main bracket.STL) and then covered in the cases. (Base case.STL, Back case.STL, and Front case.STL).
Figure 2c illustrates the MB transfer sequence. (i) The initial MECH actuation involves stage movements to enclose the upper magnets with the sheaths (Video 1). (ii) The upper magnet assembly is immersed in the sample wells to collect MBs (Video 2). The lower magnet is located away from the well bottom. (iii) The upper magnet assembly is raised, and the plate is translated in the x direction, positioning MBs above the next well. (iv) The upper assembly descends into the well. The sheath is detached (Video 3), and the rod magnet is withdrawn. (v) The vibration module is activated to release MBs into the solution (Video 4). Concurrently, the lower magnet is raised to the bottom of the plate to accelerate MB recovery. (vi) The plastic sheath is raised to cover the upper magnet, and the lower magnet is retracted. The entire plate then undergoes orbital rotation to disperse the MBs.
Video 1.
Operation of inserting and releasing the plastic sheath
Video 2.
Operation of magnetic bead collection.
Video 3.
Operation of detaching the plastic sheath
Video 4.
Operation of magnetic bead release.
The system control was implemented using a microcontroller integrated with position sensors and stepper motors. An embedded software, developed in Python, coordinated all actuator movements and timing sequences. Key operational parameters, including manipulator movements, assay duration, vibration period, vortex speed, and mixing time, were programmable through a touchscreen interface (Figure. S3) or by uploading command scripts.
Design of the orbital shaker
We chose an orbital shaker to generate uniform flow profiles across all 96 wells. The shaking module consisted of four eccentric cams, with a main cam (orbiting radius, 1.5 mm) driven by a direct-current (DC) motor (Figure. 2d). To determine the optimal operating conditions, we performed three-dimensional transient computational fluid dynamics (CFD) simulations. We used the actual well geometry (height: 10.67 mm; bottom diameter: 6.35 mm; top diameter: 6.86 mm) and the input sample volume (100 μL). The simulation model incorporated inertial forcing, gravity, surface tension, and wetting boundary conditions (see Methods) to evaluate free-surface deformation and internal flow structures over a range of rotational speeds. The design objective was to maximize MB dispersion while preventing liquid spillovers.
The simulation identified three hydrodynamic regimes (Figure. 2e and Figure. S4). At low rotational speeds (~600 rpm), circulation was weak, and the fluid was largely confined near the bottom of the well (Regime I; Video 5). At intermediate speeds (~600–1000 rpm), the flow displayed a toroidal vortex structure spanning most of the liquid volume (Regime II; Video 6): the closed circulation loop indicated continuous exchange of fluid between the well bottom and the free surface, thereby enhancing convective transport and bead dispersion. At higher speeds (≳1000 rpm), pronounced free-surface deformation increased the risk of spillovers (Regime III; Video 7).
Video 5.
CFD simulation of fluidic motion within a micropate well (rotation, 200 rpm; duration, 1.2 sec).
Video 6.
CFD simulation of fluidic motion within a micropate well (rotation, 800 rpm; duration, 0.3 sec).
Video 7.
CFD simulation of fluidic motion within a micropate well (rotation, 1200 rpm; duration, 0.2 sec).
Experimental observations were consistent with these predictions (Figure. 2f). At ≤400 rpm, MBs remained aggregated after 1 min of shaking. At 600 rpm, beads were mobilized but remained concentrated near the well center, indicating incomplete volumetric circulation. At 800 rpm, aggregated beads were uniformly dispersed throughout the well volume. Based on the combined CFD and experimental results, 800 rpm was selected as the operating condition for rapid MB dispersion.
Performance evaluation of MECHA in sample preparation
We next evaluated MECHA’s performance in automated sample preparation. A key metric was MB transport efficiency, defined as the ratio of transferred beads to input beads. This ratio was estimated by counting residual beads remaining in a well after the transfer step (Figure. S5). To reduce non-specific bead binding, we treated both the well plates and plastic sheath with 1% BSA (see Methods).
We tested magnetic beads of two sizes (2.8 μm and 1 μm) to evaluate the effect of bead size on transport efficiency. MECHA demonstrated strong performance with 2.8-μm magnetic beads, whereas manipulating 1-μm beads was suboptimal (Figure. S6). Magnetic capture force scales with particle volume, while viscous drag scales linearly with particle radius. Therefore, smaller beads experience reduced capture efficiency and slower transport velocities under identical field conditions. This limitation can be addressed by using a larger magnet, which will increase both the magnetic force and the bead-capture area.
In the current design, we opted to use 2.8 μm beads. The average bead transport efficiency between adjacent wells reached approximately 91% (Figure. 3a). After nine sequential transfers across the well array, the final bead retention was 47% (Figure. 3b), consistent with the inter-well transport efficiency. The retention values were highly reproducible, with a coefficient of variation (CV) below 5%.
Figure 3. Performance evaluation of the MECHA system.

(a) MBs were counted between two adjacent wells before and after magnetic transport. The estimated transport efficiency was approximately 91%. Data are presented as mean ± s.d. from technical replicates (n = 20). (b) MB retention after nine consecutive transfers through the full well array. The residual bead counts were <10% of the initial MB input in intermediate wells (well numbers 1–8), and the final bead recovery was 47% of the initial input. Data are presented as mean ± s.d. from technical quadruples. (c) MECHA concurrently processed eight samples of the same EV concentrations for ECL labeling. The measured ECL intensities were consistent among all samples. Bars indicate the mean from duplicate measurements. (d) The experiment in (c) was repeated using samples with different input EV concentrations. Coefficients of variation (CVs) were <10% across all input levels, confirming the assay's reproducibility. Each bar shows the CV calculated from eight samples. (e) Comparison of ECL signal intensities obtained from MECHA-prepared and manually processed EV samples. The strong correlation (Pearson’s r = 0.97) indicated analytical equivalence between automated and manual sample preparation.
We assessed the uniformity of ECL signals by processing multiple samples in parallel. A parent EV sample was aliquoted into eight input wells, and MECHA prepared eight ECL-ready MB complexes by completing automated steps: EV capture on MBs, washing, labeling with ECL probes, and final buffer rinses. ECL measurements exhibited high signal consistency (Figure. 3c). The CV values were <10% across varying input EV concentrations (Figure. 3d), highlighting MECHA’s capability for reproducible sample handling.
We compared ECL signal intensities between MECHA-prepared and manually processed samples. Two types of analytical targets, EVs (Figure. 3e) and soluble protein (IL-6; Figure. S7), were tested at varying concentrations. The immunoassay workflow involved five MB-transfer steps, corresponding to an expected final bead recovery of ~68%. Manual sample processing yielded a comparable recovery of ~75% (Figure. S8). Consistent with these results, ECL signals obtained from MECHA-processed samples showed strong agreement with those from manual preparation (Pearson’s r = 0.90), supporting the analytical equivalence of the automated workflow to conventional manual methods. Bland-Altman analysis supported this agreement, showing a mean bias of 0.025 ± 0.11 (95% limits of agreement: −0.19 to 0.24), indicating negligible systematic deviation between the two preparation methods (Figure. S9).
EV assays using MECHA-prepared samples
We optimized the MECHA protocol for EV protein detection through ECL assays (see Table S1 for operation parameters). The sample wells were preloaded with reagents for sequential automated processing (Figure. 4a). For EV capture, MBs (diameter, 2.8 μm) were functionalized with antibodies against CD63, a tetraspanin expressed on EV surfaces. To generate probe antibodies conjugated to ECL reporters, we employed a click-chemistry approach using half-antibodies (Figure. 4a, bottom). Full antibodies were first reduced to cleave interchain disulfide bonds, exposing free sulfhydryl groups on the heavy chains. These thiols were then reacted with a maleimide-functionalized [Ru(bpy)3]2+ complex. This conjugation approach consistently produced a labeling ratio of 0.6 [Ru(bpy)3]2+ per half antibody (Figure. S8).
Figure 4. EV assay workflow for MECHA–ECL detection.

(a) Automated MECHA workflow for EV protein detection. Reagents were preloaded into a linear well array, and MECHA performed sequential EV capture on magnetic beads (MBs), followed by washing and labeling. (Bottom) Preparation of half antibodies labeled with ECL probes. Full antibodies were reduced to expose sulfhydryl groups, which were subsequently reacted with maleimide-functionalized [Ru(bpy)3]2+ reporters. (b) Labeled MBs exhibited strong luminescence under ECL microscopy, confirming successful sample preparation. (c) Samples with varying EV concentrations were processed by MECHA, and ECL measurements were performed using the custom-designed eight-plex ECL chip. EVs were captured by CD63-specific MBs and labeled for CD63. The signal level was EV dose-dependent. (d) EV titration curve obtained using a CD63-specific ECL assay. The assay achieved a limit of detection of 4 × 104 EVs/mL (mean blank + 3 × SD; indicated by a black arrow) with a dynamic range spanning four orders of magnitude. The ECL assay outperformed conventional ELISA in both sensitivity and linear range. Data are presented as mean ± s.d. from technical replicates (n = 3 for ECL and n = 2 for ELISA).
After MECHA processing, labeled MBs were mixed along with the coreactant TPrA and deposited onto an electrode for ECL measurements. An electrical potential of 1.2 V was applied between the working and counter electrodes to trigger light emission. ECL microscopy (see Methods) confirmed successful sample preparation, revealing individual MBs with strong luminescence signals (Figure. 4b). We subsequently used MECHA to process samples with varying EV concentrations. ECL measurements were performed on the custom-designed multiplex chip. The signal level was dose-dependent (Figure. 4c), indicating that quantitative EV detection was achieved using MECHA-prepared samples.
Using a CD63-specific ECL probe, we constructed an EV titration curve based on the mean ECL intensity per electrode (Figure. 4d). The assay achieved a limit of detection (LOD) of 4 × 104 EVs/mL (technical replicates, n = 3) and a dynamic range spanning four orders of magnitude. The ECL assay outperformed conventional ELISA in both sensitivity and dynamic range, with an LOD approximately 1,800 times lower and a dynamic range more than 100 times wider.
Application demonstration using EV proteins in clinical samples
As a pilot utility study, we applied the MECHA-ECL pipeline to profile EV proteins in clinical plasma samples, with the demographic information presented in Table S2. Figure 5a shows the study design. Plasma was collected from 37 patients with advanced colorectal cancer (CRC) undergoing preoperative chemotherapy with the FOLFOX regimen (folinic acid, fluorouracil, and oxaliplatin). After completing a 2-month treatment course, all patients underwent curative surgery. Surgical tissue specimens were evaluated by histopathology to assess therapeutic response to chemotherapy, classifying 20 patients as good responders and 17 as poor responders. Longitudinal plasma samples were obtained at three defined time points: before chemotherapy, after chemotherapy, and after surgery. MECHA-ECL analysis was performed to quantify treatment-associated changes in EV marker expression.
Figure 5. Application of MECHA–ECL analysis to clinical plasma samples.

(a) Study design for longitudinal EV analysis in colorectal cancer (CRC) patients undergoing preoperative chemotherapy (FOLFOX) and curative surgery. Plasma samples were collected at three defined time points: before chemotherapy, after chemotherapy, and after surgery. Tissue specimens were obtained during surgery for histopathological examination, and therapeutic responses were classified. (b) Normalized expression levels of EV surface markers before and after chemotherapy, stratified by the treatment response to preoperative chemotherapy. Marker expression was generally higher in poor responders than in good responders. (c) Logistic regression analysis using pretreatment EV data identified EGFR and CD133 as the most predictive marker pair for chemotherapy response. The risk score from the regression model was used to construct a receiver operating characteristic (ROC) curve. The area under the curve (AUC) was 0.74, indicating fair predictive performance. (d) Risk scores derived from the EGFR + CD133 regression model were calculated for pre- and post-surgery plasma samples. Both poor and good responder groups showed decreased scores after surgery, consistent with tumor removal and a reduction in tumor-derived EV burden.
For EV capture, we prepared MBs functionalized with antibodies against CD63 or CD9. A mixture of CD63- and CD9-conjugated MBs was used to capture EVs directly from plasma. The captured EVs were profiled for six CRC-associated surface markers: epithelial cell adhesion molecule (EpCAM), epidermal growth factor receptor (EGFR), glycoprotein 33 (GPA33), multidrug resistance protein (MDR), carcinoembryonic antigen (CEA), and prominin-1 (CD133). These markers are previously validated as EV biomarkers for CRC diagnostics14. Two additional targets were included: CD63 to estimate EV loading and control IgG to account for nonspecific background signals. CD63 was included to estimate total EV loading, and IgG served as a nonspecific control. For each marker (M), the net signal was defined as ΔIM = IM − IIgG. Marker expression was normalized to EV abundance as ΔIM/ΔICD63, with CD63 expression serving as an estimate of EV amount14.
Figure 5b shows the normalized expression levels of EV markers before and after chemotherapy, stratified by treatment response determined from histopathology. Although marker expression was generally elevated in poor responders relative to good responders, no single marker provided sufficient discriminative power. To identify an EV signature prognostic of therapeutic response, we evaluated all candidate marker combinations using logistic regression applied to pretreatment samples. We then selected the optimal model that simultaneously minimized the Akaike information criterion (AIC) and maximized the area under the receiver operating characteristic (ROC) curve. This combined criterion identified the EGFR-CD133 pair as the most informative and parsimonious marker combination (Figure. S9). The resulting model yielded an AUC of 0.74 (Figure. 5c), indicating fair predictive performance.
We applied the same regression model (EGFR + CD133) to calculate risk scores for pre- and post-surgery plasma samples (Figure. 5d). All CRC patients exhibited decreased risk scores after surgery, likely reflecting a reduction in tumor burden. This observation was consistent with previous studies that reported the decline of CRC EV markers following tumor resection14,22.
DISCUSSION AND OUTLOOK
Standardizing sample preparation is crucial for reliable, reproducible assays1,5,23. We developed the MECHA system to automate MB-based processing, a widely used format in bioanalytical assays. MECHA integrates programmable magnetic manipulation, controlled vibration, and orbital mixing, and performs all liquid-handling operations within a closed system. The prototype device was engineered to process eight samples concurrently, performing target capture, washing, and labeling steps without operator intervention. A signal variation of less than 10% across the samples was achieved, demonstrating the feasibility of precise and reproducible sample preparation. In a pilot study, we coupled MECHA with highly sensitive ECL detection to profile EV proteins in clinical plasma samples of CRC patients. These results demonstrated the translational potential of the MECHA-ECL pipeline and highlighted the utility of EV analysis for assessing therapeutic efficacy and monitoring disease progression.
MBs can be aggregated and separated under an external magnetic field, simplifying their handling11. However, even superparamagnetic beads tend to remain clustered after the field is removed. Inter-particle attractions, arising from hydrophobic and magnetic dipole–dipole interactions, can overcome the randomizing force from Brownian motion, hindering MB dispersion in a buffer24,25. MECHA resolved this issue by incorporating mechanical agitation of both tip vibration and orbital shaking. CFD simulations guided the shaker design to prevent sample spillage and ensure efficient mixing within the 96-well format. This combined approach enabled consistent bead dispersion, supporting reliable, reproducible ECL assay performance. Although several automated platforms have been developed for EV isolation and immunomagnetic workflows (see Table S3), MECHA distinguishes itself through its fully reprogrammable architecture, enabling seamless adaptation to diverse assay protocols for MB-based chemistries and detection modalities (e.g., electrochemical assays and ELISAs). Such versatility is expected to facilitate seamless translation of new assays into standardized, automated workflows.
We applied MECHA to process serial plasma samples collected from CRC patients who received preoperative chemotherapy. EVs were chosen as an analytical target in an MB-based assay, considering their growing recognition as noninvasive diagnostic and prognostic biomarkers. Among six surface markers (EpCAM, EGFR, GPA33, MDR, CEA, and CD133) profiled, CD133 and EGFR were the most prognostic of chemotherapy response, with higher pretreatment levels associated with poorer therapeutic outcomes. This finding aligns with previous studies identifying CD133 as a cancer stem cell marker linked to chemotherapy resistance and poor prognosis in multiple cancer types26-28.
The current study has several limitations. While MECHA demonstrated reproducible performance within a single prototype, validation across multiple fabricated units will be necessary to confirm scalability. Further improvements in magnet strength and reduced magnet–sheath spacing may enhance bead-handling efficiency, particularly for smaller beads, thereby expanding the platform's versatility. However, highly viscous samples (e.g., sputum) may still present challenges for efficient MECHA processing. It should also be noted that the clinical analysis presented here is based on a relatively small cohort and lacks an independent validation set, which may introduce a risk of model overfitting. Future studies utilizing larger, independent cohorts are needed to validate robustness and generalizability. There are also opportunities to reduce the cost and size of the current setup. For example, the current system uses an EMCCD camera to detect the ECL signal, which may be replaced by lower-cost imaging hardware, such as CMOS-based detectors with higher-efficiency ECL coreactants9,10,29 to maintain sufficient signal intensity. Alternative readout modalities, such as electrochemical sensing, may offer compact, cost-effective detection formats for portable applications.
METHODS
MATERIAL
4-(dimethylamino)pyridine (4-DMAP, 98%), Amicon centrifugal filter (3 kDa, 0.5 mL), Tripropylamine (TPrA), tris(2-carboxyethyl)phosphine hydrochloride (TECP), Aminomaleimide, Dimethyl sulfoxide (DMSO), and Ru(bpy)2(mcbpy-O-Su-ester)(PF6)2 were purchased from Milipore Sigma. Bovine serum albumin, magnetic beads (1μm and 2.8 μm), 1xphosphate buffer saline (PBS), and 96-well plate were purchased from Thermo Fisher Scientific. Size-exclusion chromatography (SEC) column was purchased from iZon. ECL chip was printed from UXN. The submitted drawing was manufactured through the following machining: 3D Printing parts (ABS): Back case.STL, Base case.STL, bottom magnet holder.STL, Cam 1.STL, and Cam2.STL (Data S1); CNC Machining parts (Aluminum): Base plate.STL, Main bracket.STL, plate bed.STL, PS holder part1.STL, PS holder part2.STL, shaking plate.STL, and Top magnet holder.STL (Data S2); Injection molding (polystyrene): Plastic sheath (PS).STL (Data S3); Laser cutting and acrylic bonding (acrylic plate): Front case.STL (Data S4).
MECHA design and fabrication
The system was designed to process eight samples simultaneously in a 96-well plate format (see Figure. S1 for details). The assembly included one motorized x-axis stage for the sample plate and three independently driven z-axis stages supporting two magnet modules and removable plastic sheaths. Two vibration motors were mounted on the sheath stage, and an additional motor was attached to the plate stage to generate orbital vortex motion. Metal components were fabricated via computer numerical control (CNC) machining. The device’s plastic parts were produced by 3D printing or laser cutting. The plastic sheath was fabricated by injection molding. Embedded control software was developed in Arduino C++ and deployed on a Teensy 4.0 microcontroller. A touchscreen interface (Nextion HMI LCD touchscreen) was integrated to provide real-time user interaction and system monitoring. Through this interface, users can adjust key operational parameters, including assay time, vibration duration, vortex speed (revolutions per minute, RPM), and vortex time. Each assay sequence was executed along the column direction of the 96-well plate.
ECL chip fabrication and detector setup.
The electrode pattern for the ECL chip was designed using EAGLE software (Autodesk) and patterned onto a gold film by UXN Research Center (UXN, South Korea). To facilitate handling, the thin gold film was supported by a 1 mm-thick acrylic plate, which was laser-cut to the desired size and bonded to the back of the gold film using double-sided adhesive (468MP, 3M). A second, donut-shaped acrylic plate (3–4 mm thick) was also laser-cut and attached to the front side of the gold film, precisely aligned with the reaction electrode area, to form a reservoir for the ECL reagent. The ECL connector was fabricated using a 3D printer and assembled with pogo pins to interface with the potentiostat (Digi-Ivy Inc, DY2216). The detection module consisted of an EM-CCD camera (Falcon) equipped with a 2–6× zoom lens to capture the ECL signal emitted from the electrode surface.
CFD modeling of orbital shaking
Three-dimensional transient CFD simulations were conducted to optimize orbital shaking conditions in a 96-well plate. The modeled geometry replicated a standard well (height 10.67 mm; bottom diameter 6.35 mm; top diameter 6.86 mm) with a working liquid volume of 100 μL, corresponding to an initial liquid height of approximately 3.09 mm. The incompressible, unsteady Navier–Stokes equations were solved using a transient finite-element framework implemented in COMSOL Multiphysics (COMSOL Inc., USA). Orbital motion was introduced as a time-dependent body acceleration representing circular translation of the plate30,31. The liquid-air interface was captured using a phase-field method based on the Cahn–Hilliard formulation, with surface tension incorporated through a free-energy functional32,33. Wetting at the polystyrene wall was represented by a prescribed static contact angle of 87.4°. No-slip boundary conditions were applied to solid surfaces, and the top boundary of the air domain was set to atmospheric pressure. Simulations were advanced until a periodic steady state was reached, after which velocity fields and free-surface morphologies were extracted for analysis.
Magnetic bead preparation
Two milligrams of streptavidin-coated magnetic beads (Dynabeads M-280 Streptavidin, Invitrogen) were suspended in 1× phosphate-buffered saline (PBS) solution (500 μL) at 20 °C and washed three times (1× PBS). The beads were then resuspended in 200 μL of 1× PBS. This suspension was divided into two 100 μL aliquots. To one aliquot, 20 μg of biotinylated anti-CD63 antibodies (Ancell) was added, and to the other, 20 μg of biotinylated anti-CD9 antibodies (Abcam) was added. The resulting mixtures were incubated for 2 hours (20 °C) with continuous rotation. After incubation, the beads were washed three times with 1× PBS, and each antibody-conjugated bead solution was resuspended in 50 μL SuperBlock (Thermo Fisher Scientific) in PBS.
Antibody cleavage and preparation of the ruthenium complex
Antibody disulfide bonds were reduced to generate free thiol groups. Briefly, 10 μL of antibody solution (1 mg/mL) was diluted with 180 μL PBS (pH 7.4), followed by the addition of 10 μL of tris(2-carboxyethyl)phosphine hydrochloride (TCEP; 100 mM in deionized water, MilliporeSigma). The mixture was incubated at 25 °C for 30 min with gentle mixing using a HulaMixer (Thermo Fisher Scientific). For preparation of the maleimide-functionalized ruthenium complex, 0.5 μL of amino-maleimide (50 mM in dimethyl sulfoxide, DMSO) was mixed with 5 μL of Ru(bpy)2(mcbpy-O-Su-ester)(PF6)2 (4.927 mM in deionized water), followed by dilution with 194.5 μL of 1× PBS to yield a maleimide-activated [Ru(bpy)3]2+ solution.
Ruthenium conjugation with cleaved antibody
The reduced antibody was mixed with the maleimide-functionalized [Ru(bpy)3]2+ and incubated at 25 °C for 1 h in the dark with gentle mixing using a HulaMixer (Thermo Fisher Scientific) to allow thiol–maleimide coupling. Following the reaction, the mixture was purified twice using a 3 kDa molecular weight cutoff centrifugal filter (Amicon Ultra, MilliporeSigma) to remove unreacted ruthenium complex. The purified antibody–[Ru(bpy)3]2+ conjugate was diluted to 10 μg/mL in 0.1% bovine serum albumin (BSA) and stored at −20 °C until use. Antibodies used in this study are listed in Table S4.
Sample preparation sequence
The automated sample processing was performed using the MECHA system (see Table S1 for the operation settings). To reduce non-specific bead binding, we treated both the well plates and plastic sheath with 1% BSA for 30 min at 20-24 °C before use. A 96-well plate was preloaded with reagents according to the programmed sequence: mixed MB solution (2 μL), anti-CD63 and anti-CD9 antibodies (1 μL each), wash buffers, and labeling reagents. EV capture. EV samples (100 μL of EV-spiked PBS or 100 μL of plasma) were loaded into the MB well, and the mixture was incubated for 30 min with vortexing (200 rpm). Washing. MBs were magnetically collected and transferred through a well containing 0.05% PBS-Tween 20 (PBS-T). EV labeling. MBs were transferred to a well containing 100 μL of [Ru(bpy)3]2+ labeled antibodies (2 μg/mL in 0.1% BSA) and incubated for 15 min. Final wash and collection. Labeled MBs were washed three times with 1× PBS and resuspended in 30 μL of PBS for ECL measurement.
ECL measurement
For ECL measurements, 7 μL of each prepared MB suspension was deposited onto the working electrode (8 electrodes in total), followed by the addition of 1 mL of 180 mM tripropylamine (TPrA) in PBS (pH 7.4). ECL signals were recorded using the custom-built detection system described above. Chronoamperometric measurements were initiated, during which light emission from the electrode surface was acquired and quantified as the ECL intensity.
Cell culture and EV isolation
SW480 and SW620 colorectal cancer cell lines were obtained from the American Type Culture Collection (ATCC) and maintained in Dulbecco’s Modified Eagle Medium (DMEM; Cellgro) supplemented with 10% fetal bovine serum (FBS) and 1% penicillin–streptomycin (Cellgro). Cells were routinely tested and confirmed to be free of mycoplasma contamination using a MycoAlert Mycoplasma Detection Kit (Lonza, LT07-418). Cells at passages 1–20 were cultured in complete medium until they reached ~80% confluency, then incubated for 48 hours in medium containing 5% EV-depleted FBS. Conditioned medium was collected, sequentially filtered through a 0.2 μm membrane (Millipore), and centrifuged at 300 × g for 10 min to remove cell debris. The supernatant was concentrated to 0.5 mL using a Centricon Plus-70 centrifugal filter unit (10 kDa cutoff; Millipore) and loaded onto a qEV 70 nm size-exclusion chromatography (SEC) column (iZon). During elution, 4 mL of 1× PBS was added in increments of 100–200 μL. The first 3 mL of eluate was discarded, and the subsequent 1.5 mL fraction, containing EVs, was collected. The collected EVs were further concentrated using an Amicon Ultra-2 centrifugal filter unit (Millipore) and stored at -80 °C until use.
Clinical samples
Clinical plasma samples were collected under protocols approved by the Institutional Review Board (IRB) of Massachusetts General Hospital (2019P003441 and 2019P003472; principal investigator, H.L.) and Chonnam National University Hwasun Hospital (CNUHH 2019-228; principal investigator, C.H.K.). All procedures were conducted in accordance with institutional guidelines. Written informed consent was obtained from all participants. Peripheral blood samples (~10 mL) were collected from patients with advanced colorectal cancer at three defined time points: before chemotherapy, after chemotherapy, and within 5 days of surgery. Samples were centrifuged at 2,000 × g for 3 min to separate plasma from the red blood cell and buffy coat layers. The resulting plasma fractions were aliquoted and stored at -80 °C until use.
Tumor tissue analysis
Tumor regression was evaluated on surgical specimens obtained after preoperative chemotherapy. Histopathological analysis was performed at Chonnam National University Hwasun Hospital. Tissue sections were stained with hematoxylin and eosin (H&E) and graded according to the American Joint Committee on Cancer / College of American Pathologists (AJCC/CAP) tumor regression grading (TRG) system. Tumor response was classified as follows: TRG 1, near complete response (<10% residual tumor); TRG 2, moderate response (10–50%); TRG 3, minimal response (>50%); and TRG 4, poor or no response (extensive residual tumor without regression). For analysis, TRG 1–2 were grouped as good responders and TRG 3 and 4 as poor responders.
Data processing and statistical analysis
ECL intensity values were obtained for each target marker (IM) and the IgG control (IIgG). The net signal for each marker was calculated as ΔIM = IM − IIgG. Marker expression levels were normalized to CD63 as ΔIM/IΔCD63. For statistical evaluation, data were analyzed using GraphPad Prism (version 10.4) and R (version 4.2.2). Comparisons between groups were performed using unpaired two-tailed t-tests unless otherwise specified. To identify predictive markers of chemotherapy response, logistic regression was performed using marker data collected prior to chemotherapy. Risk scores were computed from regression coefficients, and receiver operating characteristic (ROC) curves were generated to evaluate model performance. The area under the ROC curve (AUC) was used as a measure of predictive accuracy.
Supplementary Material
Data S3: Injection molding.zip, Plastic injection molded part (Plastic sheath)
Data S1: 3D Printing.zip, 3D printed parts
Data S4: Laser cutting.zip, Laser cutting and acrylic bonding part (Front case)
Data S2: Aluminum CNC.zip, Aluminum CNC-machined parts
Acknowledgements
This work was supported in part by National Institutes of Health (NIH) 1U01CA279858 (H.L.), U01CA284982 (H.L.), R01CA304204 (H.L.), R01HL163513 (H.L.), R21CA267222 (H.L.), R61CA297878 (H.L.); Massachusetts General Hospital (MGH) Scholar Fund (H.L.); the National Research Foundation (NRF) of Korea grant RS-2024-00354654 (I.-S.S.) funded by the Ministry of Science and ICT; and the Basic Science Research Program (RS-2025-02633604) through the NRF funded by the Ministry of Education (Y.K.C.)
Footnotes
Resource availability
Lead Contact
Requests for further information and resources should be directed to and will be fulfilled by the lead contact, Hakho Lee (hlee@mgh.harvard.edu)
Material availability
This study did not generate new unique materials.
- All data supporting the results in this study are available within the paper and its supplementary information. Additional data are available from the corresponding authors upon reasonable request.
- This paper does not report original code.
- Any additional information required to reanalyze the data reported in this paper, including the raw patient datasets generated and analyzed dur-ing the study are available from the corresponding authors on reason-able request, subject to approval from the Institutional Review Board of Massachusetts General Hospital.
REFERENCE
- 1.Fu Q, Murray CI, Karpov OA, and Van Eyk JE (2023). Automated proteomic sample preparation: The key component for high throughput and quantitative mass spectrometry analysis. Mass Spectrom. Rev, 42, 873–886. 10.1002/mas.21750. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Radnik J, Hodoroaba V-D, Jungnickel H, Tentschert J, Luch A, Sogne V, Meier F, Burr L, Schmid D, Schlager C, et al. (2022). Automation and Standardization-A Coupled Approach towards Reproducible Sample Preparation Protocols for Nanomaterial Analysis. Molecules, 27, 985. 10.3390/molecules27030985. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Guan XL, Chang DPS, Mok ZX, and Lee B (2023). Assessing variations in manual pipetting: An under-investigated requirement of good laboratory practice. J. Mass Spectrom. Adv. Clin. Lab, 30, 25–29. 10.1016/j.jmsacl.2023.09.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Li Y, Liu C, Shen W, and Tang S (2026). Recent advances in high-throughput automated sample preparation technologies in bioanalysis. Talanta, 300, 129171. 10.1016/j.talanta.2025.129171. [DOI] [PubMed] [Google Scholar]
- 5.More D, Khan N, Tekade RK, and Sengupta P (2025). An Update on Current Trend in Sample Preparation Automation in Bioanalysis: strategies, Challenges and Future Direction. Crit. Rev. Anal. Chem, 55, 1461–1485. 10.1080/10408347.2024.2362707. [DOI] [PubMed] [Google Scholar]
- 6.Calabria D, Lazzarini E, Pace A, Trozzi I, Zangheri M, Cinti S, Difonzo M, Valenti G, Guardigli M, Paolucci F, et al. (2023). Smartphone-based 3D-printed electrochemiluminescence enzyme biosensor for reagentless glucose quantification in real matrices. Biosens. Bioelectron, 227, 115146. 10.1016/j.bios.2023.115146. [DOI] [PubMed] [Google Scholar]
- 7.Cho YK, Kim H, Bénard A, Woo H-K, Czubayko F, David P, Hansen FJ, Lee JI, Park JH, Schneck E, et al. (2022). Electrochemiluminescence in paired signal electrode (ECLipse) enables modular and scalable biosensing. Sci. Adv, 8, eabq4022. 10.1126/sciadv.abq4022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Chinnadayyala SR, Park J, Le HTN, Santhosh M, Kadam AN, and Cho S (2019). Recent advances in microfluidic paper-based electrochemiluminescence analytical devices for point-of-care testing applications. Biosens. Bioelectron, 126, 68–81. 10.1016/j.bios.2018.10.038. [DOI] [PubMed] [Google Scholar]
- 9.Cho YK, Jeon S-H, Park JY, Kim H, Shim B-J, Nam K, Köcher SS, Lee H, Woo H-K, Park JH, et al. (2025). Discovery of a new coreactant for highly efficient and reliable electrochemiluminescence. Cell Rep. Phys. Sci, 6, 102864. 10.1016/j.xcrp.2025.102864. [DOI] [Google Scholar]
- 10.Zanut A, Fiorani A, Canola S, Saito T, Ziebart N, Rapino S, Rebeccani S, Barbon A, Irie T, Josel H-P, et al. (2020). Insights into the mechanism of coreactant electrochemiluminescence facilitating enhanced bioanalytical performance. Nat. Commun, 11, 2668. 10.1038/s41467-020-16476-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Fabiani L, Mazzaracchio V, Moscone D, Fillo S, De Santis R, Monte A, Amatore D, Lista F, and Arduini F (2022). Paper-based immunoassay based on 96-well wax-printed paper plate combined with magnetic beads and colorimetric smartphone-assisted measure for reliable detection of SARS-CoV-2 in saliva. Biosens. Bioelectron, 200, 113909. 10.1016/j.bios.2021.113909. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Feng Y, Wang C, Zhou W, Yang X, Paolucci F, Valenti G, and Qi H (2025). Tomography Electrogenerated Chemiluminescence Imaging from Magnetic Microbeads. Small, 21, e2500804. 10.1002/smll.202500804. [DOI] [PubMed] [Google Scholar]
- 13.Cha BS, Park KS, and Park JS (2020). Signature mRNA markers in extracellular vesicles for the accurate diagnosis of colorectal cancer. J. Biol. Eng, 14, 4. 10.1186/s13036-020-0225-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Park J, Park JS, Huang C-H, Jo A, Cook K, Wang R, Lin H-Y, Van Deun J, Li H, Min J, et al. (2021). An integrated magneto-electrochemical device for the rapid profiling of tumour extracellular vesicles from blood plasma. Nat. Biomed. Eng, 5, 678–689. 10.1038/s41551-021-00752-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Shao H, Im H, Castro CM, Breakefield X, Weissleder R, and Lee H (2018). New Technologies for Analysis of Extracellular Vesicles. Chem. Rev, 118, 1917–1950. 10.1021/acs.chemrev.7b00534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Zhai C, Xie F, Xu J, Yang Y, Zheng W, Hu H, Ding X, and Yu H (2023). Correlation between membrane proteins and sizes of extracellular vesicles and particles: A potential signature for cancer diagnosis. J. Extracell. Vesicles, 12, e12391. 10.1002/jev2.12391. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Zhang L, Shen H, and Zhang N (2025). Advances in extracellular vesicle (EV) biomarkers for precision diagnosis and therapeutic in colorectal cancer. Front. Oncol, 15, 7. 10.3389/fonc.2025.1581015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Miao W, Choi J-P, and Bard AJ (2002). Electrogenerated chemiluminescence 69: the tris(2,2’-bipyridine)ruthenium(II), (Ru(bpy)3(2+))/tri-n-propylamine (TPrA) system revisited-a new route involving TPrA*+ cation radicals. J. Am. Chem. Soc, 124, 14478–14485. 10.1021/ja027532v. [DOI] [PubMed] [Google Scholar]
- 19.Sornambigai M, Bouffier L, Sojic N, and Kumar SS (2023). Tris(2,2’-bipyridyl)ruthenium (II) complex as a universal reagent for the fabrication of heterogeneous electrochemiluminescence platforms and its recent analytical applications. Anal. Bioanal. Chem, 415, 5875–5898. 10.1007/s00216-023-04876-4. [DOI] [PubMed] [Google Scholar]
- 20.Cho YK, Choi Y, Kim S, Kim H, Chow K-F, Shin I-S, Park JH, and Lee H (2025). Scalable electrochemical system for rapid on-site detection of food allergens. Biosens. Bioelectron, 273, 117142. 10.1016/j.bios.2025.117142. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Ecke A, Westphalen T, Hornung J, Voetz M, and Schneider RJ (2022). A rapid magnetic bead-based immunoassay for sensitive determination of diclofenac. Anal. Bioanal. Chem, 414, 1563–1573. 10.1007/s00216-021-03778-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Song J, Cho MH, Cho H, Song Y, Lee SW, Nam HC, Yoon TH, Shin JC, Hong J-S, Kim Y, et al. (2025). Amplifying mutational profiling of extracellular vesicle mRNA with SCOPE. Nat. Biotechnol, 43, 1485–1495. 10.1038/s41587-024-02426-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Sens A, Rischke S, Hahnefeld L, Dorochow E, Schäfer SMG, Thomas D, Köhm M, Geisslinger G, Behrens F, and Gurke R (2023). Pre-analytical sample handling standardization for reliable measurement of metabolites and lipids in LC-MS-based clinical research. J. Mass Spectrom. Adv. Clin. Lab, 28, 35–46. 10.1016/j.jmsacl.2023.02.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Mørup S, Hansen MF, and Frandsen C (2010). Magnetic interactions between nanoparticles. Beilstein J. Nanotechnol, 1, 182–190. 10.3762/bjnano.1.22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Rampini S, Kilinc D, Li P, Monteil C, Gandhi D, and Lee GU (2015). Micromagnet arrays for on-chip focusing, switching, and separation of superparamagnetic beads and single cells. Lab Chip, 15, 3370–3379. 10.1039/c5lc00581g. [DOI] [PubMed] [Google Scholar]
- 26.Barzegar Behrooz A, Syahir A, and Ahmad S (2019). CD133: beyond a cancer stem cell biomarker. J. Drug Target, 27, 257–269. 10.1080/1061186x.2018.1479756. [DOI] [PubMed] [Google Scholar]
- 27.Ong CW, Kim LG, Kong HH, Low LY, Iacopetta B, Soong R, and Salto-Tellez M (2010). CD133 expression predicts for non-response to chemotherapy in colorectal cancer. Mod. Pathol, 23, 450–457. 10.1038/modpathol.2009.181. [DOI] [PubMed] [Google Scholar]
- 28.Ren F. (2013). CD133: A cancer stem cells marker, is used in colorectal cancers. World J. Gastroenterol, 19, 2603. 10.3748/wjg.v19.i17.2603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hong G, Su C, Lai M, Huang Z, Weng Z, Chen Y, Deng H, Chen W, and Peng H (2022). Co-Reactant-Mediated Low-Potential Anodic Electrochemiluminescence Platform and Its Immunosensing Application. Anal. Chem, 94, 12500–12506. 10.1021/acs.analchem.2c02726. [DOI] [PubMed] [Google Scholar]
- 30.Bouvard J, Herreman W, and Moisy F (2017). Mean mass transport in an orbitally shaken cylindrical container. Phys. Rev. Fluids, 2, 28. 10.1103/physrevfluids.2.084801. [DOI] [Google Scholar]
- 31.Reclari M, Dreyer M, Tissot S, Obreschkow D, Wurm FM, and Farhat M (2014). Surface wave dynamics in orbital shaken cylindrical containers. Phys. Fluids, 26, 68. 10.1063/1.4874612. [DOI] [Google Scholar]
- 32.Jacqmin D. (1999). Calculation of Two-Phase Navier–Stokes Flows Using Phase-Field Modeling. J. Comput. Phys, 155, 96–127. 10.1006/jcph.1999.6332. [DOI] [Google Scholar]
- 33.Yue P, Feng J, Llu C, and Shen J (2004). A diffuse-interface method for simulating two-phase flows of complex fluids. J. Fluid Mech, 515, 293–317. 10.1017/s0022112004000370. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data S3: Injection molding.zip, Plastic injection molded part (Plastic sheath)
Data S1: 3D Printing.zip, 3D printed parts
Data S4: Laser cutting.zip, Laser cutting and acrylic bonding part (Front case)
Data S2: Aluminum CNC.zip, Aluminum CNC-machined parts