Super-Achromatic Rapid Scanning Microendoscope for Ultrahigh-Resolution OCT Imaging

Danling Wang 1, Boyd V Hunter 2, Michael J Cobb 3, Xingde Li 4
PMCID: PMC11741674  NIHMSID: NIHMS2004893  PMID: 39830863

Abstract

Microendoscope is a critical technology to enable high-resolution imaging of internal luminal organs with optical coherence tomography. This paper reports the development of an achromatic compound microlens and a rapid scanning microendoscope based on the microlens that offers an ultrahigh transverse resolution of 4 μm (and an axial resolution of 2.2 μm when using a low-coherence light source with a broad spectrum bandwidth of 150 nm). The rapid scanning endoscope is capable of ultrahigh-resolution (UHR) optical coherence tomography (OCT) imaging in real time at an imaging speed of about 1220 lateral scans/s, with the image quality comparable to a slow bench-top UHR-OCT system (~100 scans/s). The superior performance of a scanning endoscope made of a compound microlens over an endoscope made of a conventional GRIN (gradient index) lens in imaging biological tissues has also been demonstrated.

Index Terms—: Biomedical optical imaging, endoscope, optical coherence tomography (OCT), ultrahigh resolution (UHR) imaging

I. Introduction

There is an increasing clinical need for a noninvasive high-resolution imaging technology, particularly in areas of early disease detection and guidance of biopsy and surgical interventions. Optical coherence tomography (OCT) has emerged as a high-resolution optical biopsy technique [1], capable of imaging tissue microanatomy in situ at a resolution approaching conventional histology, but without the need for tissue removal [2]. The working principle of OCT is similar to that of ultrasound B-mode imaging except that OCT utilizes near-infrared (NIR) low-coherence light rather than acoustic waves. The superb axial resolution of OCT is achieved by employing coherence gating, and a light source with a broader spectrum provides a narrower coherence gate, and thus, a higher axial resolution. Different from the axial resolution, the transverse resolution of OCT is determined by the focused beam spot size, which is governed by the focusing optics (similar to microscopy). Significant effort has been focused on developing advanced low coherence light sources for achieving ultrahigh axial resolution [3], [4]. In addition, advanced lenses are also critical for handling the broad spectrum of a low-coherence light source, and producing a high transverse resolution with minimal chromatic aberration. In a bench-top OCT system, some of the off-the-shelf NIR-corrected objective lenses can be a convenient choice.

The imaging penetration depth of OCT in many highly scattering tissues, determined by the tissue optical properties (i.e., absorption and scattering), is approximately 0.5–2.5 mm, depending upon the wavelength of the light source. This imaging depth is ideal for assessing skin and the epithelial or endothelial lining of internal organs, from which many diseases originate such as various epithelial cancers (e.g., in the gastrointestinal, respiratory, and female reproductive tract) and atherosclerotic plaques. The development of innovative miniature imaging probes has enabled a variety of in vivo OCT applications for imaging internal organs or solid tissues. These probes include an OCT endoscope/catheter for imaging hollow organs [5]–[9], a miniature needle for interstitial imaging of solid tissues [10], and a Doppler intravascular catheter for assessing blood flow [11]. Some of these miniature probes are making a significant impact on the clinical translation of the OCT technology. According to the imaging beam direction with respect to the longitudinal axis of the probe, these probes can be divided into two groups: transverse-viewing probes and forward-viewing probes. In a transverse-viewing OCT endoscope/catheter/needle, the imaging beam exits the probe at about 90° with respect to the probe’s longitudinal axis. A 2-D beam scanning can be achieved by rotating the probe through a speedometer cable for circumferential imaging [12], or by longitudinally translating the probe for rectangular imaging [13], [14]. In a forward-viewing endoscope, the imaging beam exits the probe in the forward direction (i.e., along the probes’ longitudinal axis), and lateral beam scanning can be performed by actuating the fiber with an electromagnetic or a piezoelectric actuator [8], [9]. In practice, the choice of one probe over the other depends upon the application. For example, a side-viewing endoscope is more suitable for a systematic 3-D imaging of a lumen in a spiral fashion when equipped with a pullback mechanism, while a forward-viewing probe can be more suitable for surveying a large area (of an arbitrary shape) and for guiding biopsy. It is noticed that when imaging a portion of a large luminal area, the transverse-viewing probe requires fine control of the probe orientation to keep the imaging beam nearly perpendicular to the target surface, whereas the forward-view probe is easy to position on the target. Furthermore, the forward-viewing probe does not require bending the imaging beam by 90°, as in a transverse-viewing probe; thus, an objective lens with a high numerical aperture and a short working distance can be used without compromising the overall diameter of the probe. As a result, an ultrahigh-resolution (UHR) OCT endoscope is more feasible with the forward-viewing configuration.

Recently, we developed a rapid scanning forward-viewing endoscope based on a fiber-optic resonant scanner [9]. The endoscope can perform rapid lateral beam scanning in the kilohertz range, which represents an improvement on beam scanning speed by more than two orders of magnitude over a previously published forward-viewing probe [8], and permits real-time OCT imaging. This fast scanning endoscope enables a new OCT image acquisition sequence, namely, fast lateral scanning followed by slow depth scanning. The new image acquisition sequence permits real-time focus tracking for the first time on an endoscope setting, which is essential for maintaining a high transverse resolution (or focused spot size) at various depths [15]. As previously discussed, a high-performance achromatic lens is critical for performing the UHR OCT imaging. Such an achromatic microlens was not available, and a conventional gradient-index (GRIN) lens had to be employed in the previous scanning endoscope, resulting in a suboptimal imaging quality due to the severe chromatic aberration in the GRIN lens. In this paper, we report on the design and characterization of a recently developed super-achromatic micro-objective lens suitable for a scanning endoscope. By using the achromatic microlens, a rapid scanning microendoscope has been built and tested by imaging biological tissues. Representative UHR OCT images acquired with the fast scanning microendoscope are presented.

II. Achromatic Microlens Design and Characterization

A. Microlens Design

The design of a miniature achromatic compound lens took into account a number of challenges, including: 1) small diameter (1.5 mm in optics and 2 mm with a housing tube); 2) superb secondary spectrum control over the broad spectrum of an OCT light source (i.e., 650–950 nm); 3) low magnification to image the beam at the fiber tip with a size of the fiber mode-field diameter to a small spot size with an ~1:0.8 reduction ratio or better; and 4) good telecentricity so that the beam after the compound lens will be nearly perpendicular to the surface of the target tissue during lateral scanning. With these considerations in view, a preliminary design of a multiple-element compound lens for the scanning microendoscope was generated. All elements had a 1.5 mm diameter (with a 1.4 mm clear aperture) so that they could be housed in a protective tube of a 2 mm outer diameter. A 0.5-mm-thick protective glass window was also considered. Beyond the window, body fluid of a refractive index of ~1.38 was assumed. The designed working distance between the last lens element and the focal plane was about 2.8 mm, which provides ample room for the lens assembly to translate with respect to the window for performing focus tracking over a depth of 1–2 mm, if needed. The target beam scanning range was 320 μm in any transverse direction. Glasses of various refractive indices and dispersion properties were used for achieving sufficient beam bending and chromatic correction. This preliminary design had excellent achromatic characteristic, and the on-axis focus position at 700, 800, and 900 nm differed by less than 2 μm, as shown in Fig. 1. A modulation transfer function (MTF) analysis indicated that an approximately 45% image contrast would be achieved even at a frequency of 200 line pairs/mm (corresponding to a 5-μm resolution) at a test wavelength of 632 nm (see Fig. 2). The design also took the field curvature traced out by the scanning fiber tip into consideration (see discussion in the endoscope design, Section III-A). The maximum deviation from perfect telecentricity was less than 0.3°, which occurred near the edge of the scan field and at the edge of the wavelength band. This deviation is small enough to keep the change in the collection of the backscattered light by the fiber within a few percent.

Fig. 1.

Fig. 1.

Low chromatic aberration of the compound microlens over a broad spectrum (600–1000 nm). The design curve indicates that the longitudinal aberration (or secondary spectrum) is about 1.64 μm over a 600–1000 nm wide spectrum range.

Fig. 2.

Fig. 2.

Ideal and measured on-axis MTF curves.

B. Microlens Performance Characterization

The achromatic microlens has been successfully fabricated. The end surfaces (except the outer surface of the window that was designed to directly contact body fluid) were antireflection (AR)-coated over a broad spectrum bandwidth (650–1050 nm). All the lenses were housed in a thin-wall hypodermic tube with an overall diameter of 2 mm. A photo of the finished miniature compound lens is shown in Fig. 3(A). The measured MTF is shown in Fig. 2, which is very close to the theoretical design. Even at 250 lines/mm spatial frequency, the measured MTF is about 0.25, suggesting that an image contrast of ~1.7 is achievable for resolving a structure with a 4 μm spatial dimension. Once implemented in a scanning fiber-optic endoscope (detailed in the next section), we were able to achieve a measured transverse resolution of about 4 μm with a fiber tip of a 4.6 μm mode-field diameter. This corresponds to a magnification of M = −0.87, which is within our design tolerance (considering an expected value of Mtheory = −0.8). Compared to a previous endoscope made of a GRIN lens [9], the use of the achromatic microlens dramatically improves the transverse resolution by more than a factor of 2.5.

Fig. 3.

Fig. 3.

(A) Photo of the compound microlens. The optics has a 1.5 mm diameter and a 1.4 mm clear aperture. The overall diameter including a protective hypodermic housing is 2 mm. The ruler tick is 1 mm. (B) Illustration of the configuration used to measure the spectra backreflected from a mirror at different planes. (C) Measured spectra at different mirror positions for an endoscope that uses the compound microlens. Notice that the spectra remain almost unchanged indicating a superb achromatic performance of the microlens. (D) Measured spectra at different mirror positions for an endoscope that uses a conventional GRIN lens. Notice the dramatic change in the backreflected spectra due to the chromatic aberration of the GRIN lens. b: confocal length (or confocal parameter).

In order to characterize the chromatic aberration, a customized Ti:Sapphire laser with an ~150 nm full-width at half-maximum (FWHM) spectrum bandwidth was used as a broadband light source. The laser light, sent through the fiber-optic microlens-equipped endoscope, was initially focused on a mirror placed at the focal plane. The spectra backreflected by the mirror were measured when the mirror was placed at the focal plane and at other planes with different distances away from the focal plane [see Fig. 3(B)]. As shown in Fig. 3(C), the backreflected spectra exhibit minimal change over a large distance from 0 to 3 in the units of the confocal parameter of the microlens bmicrolens (30.5 μm), indicating the superb achromaticity of the microlens. In comparison, the same experiment was performed with an endoscope made of a 1.8 mm GRIN lens. The backreflected spectra were again measured with the mirror placed at planes over a distance of 0–2 from the focal plane in the units of the GRIN lens confocal parameter bGRIN (350 μm), and the spectra are shown in Fig. 3(D). Obviously, the backreflected spectra experienced dramatic changes due to the strong chromatic aberration of the GRIN lens. The superb achromatic performance of the microlens is critical to maintain a high axial resolution at different depths for OCT imaging.

III. Scanning Fiber-Optic Microendoscope Using the Microlens

A. Basic Layout for the Scanning Microendoscope

The small size, superb achromatic performance, and the capability of producing a tight beam focus make the compound microlens very suitable for implementation in a scanning fiber-optical microendoscope. Compared to many other customized microlenses [16], the one reported here has a much smaller size (e. g. 1.5 mm versus 4 mm).

With the miniature compound objective lens, we have developed a scanning fiber-optic UHR endoscope with an overall diameter of 2.4 mm. Fig. 4(A) shows the schematic of the fiber-optic scanner head for the endoscope that adopts a layout similar to that in [9]. The use of the miniature compound lens dramatically improves the transverse resolution (1/e2 diameter) from the previous ~13–16 μm (GRIN lens) to ~4 μm in addition to the benefit of achromaticity. In essence, the endoscope consists of a tubular piezoelectric actuator, a single-mode fiber (SMF) cantilever going through the piezo tube and glued to the end of the tube, and the compound microlens. A photo of the entire flexible microendoscope is shown in Fig. 4(B), and the length of the rigid portion at the distal end (including the piezo actuator) is about2.5–3.0 cm, depending upon the piezo tube used. The small endoscope diameter was intended to facilitate its delivery to internal organs [e.g., gastrointestinal (GI) tract] through the working channel of a standard GI endoscope. Rapid lateral beam scanning is achieved by: 1) driving the piezoelectric actuator at the mechanical resonant frequency of the fiber-optic cantilever; 2) imaging the sweeping fiber tip to the target through the microlens. In our case, the resonant frequency was about 1.22 kHz for an 8.2-mm-long SMF cantilever, and the fast beam scanning permits OCT imaging in real time. In addition to rapid 1-D lateral beam scanning, the fiber-optic scanner is also capable of circular or 2-D spiral scanning [17]. Compared to a recently reported superb side-viewing UHR endoscope [18], the UHR microendoscope presented here offers better transverse resolution (i.e. 4 μm versus 7.5 μm 1/e2 diameter) and significantly higher imaging speed (i.e., 1220 scans/s versus 240 scans/s). The piezo drive voltage is generally not high, e.g., for about ±3.0° angular fiber cantilever scanning, the peak-to-peak drive voltage is only about 40 V. This corresponds to a lateral sweeping distance of approximately 0.86 mm by the tip of an 8.2-mm-long fiber cantilever, and an ~0.69 mm beam focus scanning when using an imaging lens of 0.8 magnification (as in our URH microendoscope).

Fig. 4.

Fig. 4.

(A) Schematic of the fiber-optic scanner head for a scanning microendoscope. Fast lateral beam scanning is achieved by driving a fiber-optic cantilever at it mechanical resonant frequency with a tubular piezoelectric actuator. (B) Photo of a rapid scanning all-fiber-optic microendoscope that has an overall diameter of 2.4 mm.

B. Real-Time Imaging with the Scanning Microendoscope and Performance Comparison

In order to generate a Doppler frequency shift and perform highly sensitive heterodyne detection in the time domain, a pair of acousto-optic modulators (AOMs) were used in the reference arm of our time-domain OCT system, with one AOM upshifting the light frequency and the other downshifting the frequency, producing a net frequency shift of about 1.0 MHz for a double-pass configuration [19], [20]. The dispersion of the AOM crystals in the reference arm and the microlens in the endoscope (i.e., in the sample arm) were compensated by using our previously developed method based on the combination of a single-mode fiber and a pulse stretcher [19], [20]. To fine-tune the dispersion compensation, a CaF2 glass window and a pair of BK7 wedges were placed in the reference arm of the OCT system. Fig. 5 illustrates the interference signal when using a light source (Ti:Sapphire laser) with a center wavelength of 825 nm and a spectrum bandwidth of ~150 nm. The measured axial resolution (i.e. the FWHM of the envelope of the interference signal in Fig. 5) was approximately 3 μm in air (corresponding to ~2.2 μm in tissue when the tissue refractive index of ~1.38 is taken into account). The interference signal is nearly symmetric, indicating the dispersion in the two arms of the OCT system was reasonably balanced. Owning to the broadband AR coating on the microlens, the loss of the entire endoscope including the uncoated fiber tip was low (i.e., with a throughput of approximately 90%).

Fig. 5.

Fig. 5.

Interference signal measured in the endoscope OCT system. The FWHM of the envelope of the interference signal represents the axial resolution of the OCT system, which is about 3 μm in air (corresponding to 2.2 μm in tissue).

The performance of the scanning microendoscope was tested by imaging different tissues ex vivo with a measured imaging resolution of 4.0 × 3.0 μ m (transverse × depth in air). The light power delivered to the sample was about 6.5–7 mW, and the OCT system sensitivity was about 96 dB. A representative image of rat tail acquired by the scanning endoscope in real-time is shown Fig. 6(A), where the tendons (i.e., collagen fiber bundles) can be clearly identified as indicated with the red arrows. The image size was 811 × 1230 pixels, corresponding to a physical dimension of 0.8 × 0.6 mm (transverse × depth). The 1.22 kHz lateral beam scanning rate (e.g., 1220 lateral scans per second) permitted real-time imaging at a speed of about 1.5 frames per second (with each frame consisting of 811 lateral scans). For comparison, the same rat tail was also imaged by a bench-top time-domain UHR OCT system at nearly the same position, with an imaging speed of ~100 depth scans per second, a resolution of approximately 5.0 × 2.8 μm (transverse × depth in air) and an image size of 800 × 1000 pixels (0.8 × 0.6 mm, transverse × depth) [Fig. 6(B)]. It should be mentioned that the image acquisition sequence with the scanning endoscope is fast lateral scanning followed by a slow depth scanning, which is opposite to the image acquisition sequence by the bench-top system (which performed fast depth scanning followed by slow lateral scanning). As evident in Figs. 6(A) and (B), the image quality achieved by the rapid scanning microendoscope is very comparable to the one achieved by the bench-top system even though the bench-top system affords the use of large optics. One significant difference, however, is that the UHR imaging speed enabled by the endoscope is at least 12 times faster than that of the bench-top system.

Fig. 6.

Fig. 6.

Representative OCT images of rat tail acquired. (A) With a rapid scanning microendoscope equipped with the compound microlens. (B) With a slow bench-top ultrahigh-resolution OCT system. The tendons (collagen fiber bundles) can be clearly identified on both images (indicated with arrows). The image quality in (A) and (B) is comparable, but the imaging speed with the scanning microendoscope is about 12 times faster than that of the bench-top system. The beam focus in both images was set about 50 μm below the tissue surface.

We have also carried out another comparison study by imaging the same tissue (e.g. rabbit trachea ex vivo) with a scanning endoscope made of the achromatic compound microlens and by a scanning endoscope made of a 1.8-mm-diameter GRIN lens. Representative images of rabbit trachea are shown in Fig. 7. The two images were acquired at two frames per second and have a physical dimension of about 0.8 × 0.35 mm. The image resolution was ~4.0 × 3.0 μm for the compound microlens endoscope and ~13.5 × 3.5 μm for the GRIN lens endoscope (transverse × depth in air). Fine tissue structures such as the epithelium (E), lamina propria (LP), submucosa (SM), and glands (G) are evident on both images. Although the depth of focus for the microlens endoscope is much shorter than for the GRIN lens endoscope (i.e., 30.5 μm versus 350 μm), the image quality achieved by the microlens endoscope [Fig. 7(A)] deteriorates much slower than the one acquired by the GRIN lens endoscope [Fig. 7(B)]. In particular, the cartilage can be clearly identified in Fig. 7(A), while it is barely visible in Fig. 7(B). These findings strongly suggest the superior performance of the compound microlens endoscope over a conventional GRIN lens endoscope, and the excellent achromaticity of the compound microlens could be the key factor contributing to the superb performance of the microendoscope made of the microlens. We also noticed that the imaging depth in Fig. 7(B) seems shallower than the one in [15], where rabbit esophagus was imaged with a similar GRIN lens endoscope, and this depth reduction most likely resulted from the difference in tissue types (esophagus versus trachea), tissue freshness, and the incident power [6.5–7 mW for Fig. 7(B) versus 9.3 mW in Reference [15]].

Fig. 7.

Fig. 7.

Representative OCT images of rabbit trachea (ex vivo). (A) Acquired with a rapid scanning microendoscope made of the compound microlens. (B) Acquired with a rapid scanning microendoscope made of a conventional 1.8 mm GRIN lens. E, epithelium; LP, lamina propria; SM, submucosa; G, gland (indicated by the short arrows 1); C, cartilage ring (indicated by the long arrows 2). The beam focus in both images was set about 50 μm below the tissue surface.

IV. Conclusion

In summary, a 1.5-mm-diameter achromatic microlens and a rapid scanning UHR microendoscope (of a 2.4 mm diameter) with the microlens were developed. A real-time UHR imaging has been demonstrated with the fast scanning microendoscope. Its performance is comparable to a bench-top UHR OCT system, but with a significant improvement on imaging speed by a factor of more than 12. One flexible parameter of the scanning endoscope is the scanning speed (or the resonant frequency of the fiber-optic scanner), which is controlled by the length of the freely standing fiber cantilever. A scanning speed between a few hundred to a few thousand scans per second can be conveniently achieved depending upon the specific need. In conjunction with recently developed advanced swept sources [21]–[23], the rapid scanning UHR microendoscope can potentially enable ultrahigh-speed and UHR 3-D imaging of internal organs in vivo.

Acknowledgment

The authors would like to thank Y. Leng, X. Li, A. Warsen, J. Magula, and D. Macdonald for their generous help with endoscope fabrication, tissue preparation, laser operation, and data acquisition.

This work was supported in part by the National Institutes of Health under Grant EB-003284 and Grant CA110239 and in part by the National Science Foundation (Career Award XDL) under Grant BES-0348720.

Biographies

graphic file with name nihms-2004893-b0001.gif

Danling Wang received the Ph.D. degree in ultrafast optical spectroscopy from the Department of Physics, Peking University, Beijing, China, in 2003.

From 2003 to 2005, she was a Research Associate at the University of Wisconsin-Madison, Wisconsin, where she was working on time-/frequency-domain 2-D vibrational spectroscopy. She is currently a Postdoctoral Senor Fellow in the Department of Bioengineering, University of Washington, Seattle. Her current research interests include ultrahigh-resolution optical imaging techniques, especially the development of miniature optical imaging probes and applications of optical coherence tomography (OCT).

Boyd V. Hunter received the B.S. degree in physics with a second major in German from Brigham Young University, Provo, UT, in 1987, the M.S. degree in industrial administration from Krannert Graduate School of Management, Purdue University, West Lafayette, IN, in 1998, and the Ph.D. degree in physics with a minor in optical sciences from the University of Arizona, Tucson, in 1993.

He has been a Senior Engineer at the TPL, Inc., and was engaged in engineering, product development, and sales for LightPath Technologies. He was a Postdoctoral Appointee at the Argonne National Laboratory in the Laser Applications Laboratory. He has also been an Assistant Professor of physics at Eastern New Mexico University, Portales, NM, and as a Materials Analyst for Morton Thiokol’s (now ATK) Space Operations Division. He is currently the Director of Engineering at the Archer OpTx, Inc., Rowlett, TX. He is the author or coauthor of more than 20 technical articles and conference papers. He is the holder of four issued patents. His current research interests include the design and manufacture of high precision aspheric optics.

Dr. Hunter is a member of the Optical Society of America.

graphic file with name nihms-2004893-b0002.gif

Michael J. Cobb received the B.S. and the M.S. degrees in biomedical engineering from Marquette University, Milwaukee, WI, in 1999 and 2001, respectively. He joined the Biophotonics Research Laboratory, Department of Bioengineering, University of Washington, Seattle, in 2003 for his Ph.D. degree.

His current research interests include using optical coherence tomography and signal processing techniques for the noninvasive assessment of biological tissues.

Mr. Cobb is a member of the Alpha Eta Mu Beta, the Alpha Sigma Nu, and the Tau Beta Pi honor societies.

graphic file with name nihms-2004893-b0003.gif

Xingde Li (M’06) received the B.S. degree in physics from the University of Science and Technology of China (USTC), Hefei, China, in 1990, and the Ph.D. degree in physics from the University of Pennsylvania, Philadelphia, in 1998.

He was a Postdoctoral Research Associate at the Ultrafast Optics Group, Research Laboratory of Electronics, Massachusetts Institute of Technology, Cambridge, where he was engaged in research on biomedical optics. In fall 2001, he joined the Department of Bioengineering, University of Washington, Seattle, as an Assistant Professor, and was promoted to an Associate Professor in 2007. His current research interests include biophotonics imaging technologies (including optical coherence tomography, light scattering, multiphoton fluorescence, diffuse light tomography, endomicroscopy, etc.) and their applications in biomedicine, nanobiophotonics such as development of nanocontrast agents for improving the imaging sensitivity and molecular specificity.

Dr. Li is a member of the Optical Society of America (OSA), the International Society for Optical Engineering (SPIE), and the Biomedical Engineering Society (BMES). He is currently the Chair of the Emerging Technologies Committee of the IEEE Engineering in Medicine and Biology Society. He was the recipient of the National Science Foundation (NSF) Career Award in 2004.

Contributor Information

Danling Wang, Department of Bioengineering, University of Washington, Seattle, WA 98195 USA.

Boyd V. Hunter, the Archer OpTx, Inc., Rowlett, TX 75088 USA

Michael J. Cobb, Department of Bioengineering, University of Washington, Seattle, WA 98195 USA

Xingde Li, Department of Bioengineering, University of Washington, Seattle, WA 98195 USA.

References

  • [1].Huang D, Swanson EA, Lin CP, Schuman JS, Stinson WG, Chang W, Hee MR, Flotte T, Gregory K, Puliafito CA, and Fujimoto JG, “Optical coherence tomography,” Science, vol. 254, pp. 1178–1181, 2007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [2].Fujimoto JG, “Optical coherence tomography for ultrahigh resolution in vivo imaging,” Nat. Biotechnol, vol. 21, pp. 1361–1367, 2003. [DOI] [PubMed] [Google Scholar]
  • [3].Drexler W, Morgner U, Kartner FX, Pitris C, Boppart SA, Li XD, Ippen EP, and Fujimoto JG, “In vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett, vol. 24, pp. 1221–1223, 1999. [DOI] [PubMed] [Google Scholar]
  • [4].Hartl I, Li XD, Chudoba C, Ghanta RK, Ko TH, Fujimoto JG, Ranka JK, and Windeler RS, “Ultrahigh-resolution optical coherence tomography using continuum generation in an air–silica microstructure optical fiber,” Opt. Lett, vol. 26, pp. 608–610, 2001. [DOI] [PubMed] [Google Scholar]
  • [5].Tearney GJ, Boppart SA, Bouma BE, Brezinski ME, Weissman NJ, Southern JF, and Fujimoto JG, “Scanning single-mode fiber optic catheter-endoscope for optical coherence tomography,” Opt. Lett, vol. 21, pp. 543–545, 1996. [DOI] [PubMed] [Google Scholar]
  • [6].Bouma BE and Tearney GJ, “Power-efficient nonreciprocal interferometer and linear-scanning fiber-optic catheter for optical coherence tomography,” Opt. Lett, vol. 24, pp. 531–533, 1999. [DOI] [PubMed] [Google Scholar]
  • [7].Rollins AM, Ung-arunyawee R, Chak A, Wong RCK, Kobayashi K, Sivak MV, and Izatt JA, “Real-time in vivo imaging of human gastrointestinal ultrastructure by use of endoscopic optical coherence tomography with a novel efficient interferometer design,” Opt. Lett, vol. 24, pp. 1358–1360, 1999. [DOI] [PubMed] [Google Scholar]
  • [8].Sergeev AM, Gelikonov VM, Gelikonov GV, Feldchtein FI, Kuranov RV, Gladkova ND, Shakhova NM, Suopova LB, Shakhov AV, Kuznetzova IA, Denisenko AN, Pochinko VV, Yu PC, and Streltzova OS, “In vivo endoscopic OCT imaging of precancer and cancer states of human mucosa,” Opt. Exp, vol. 1, no. 13, pp. 432–440, Dec. 1997. [DOI] [PubMed] [Google Scholar]
  • [9].Liu XM, Cobb MJ, Chen Y, Kimmey MB, and Li XD, “Rapid-scanning forward-imaging miniature endoscope for real-time optical coherence tomography,” Opt. Lett, vol. 29, pp. 1763–1765, 2004. [DOI] [PubMed] [Google Scholar]
  • [10].Li XD, Chudoba C, Ko T, Pitris C, and Fujimoto JG, “Imaging needle for optical coherence tomography,” Opt. Lett, vol. 25, pp. 1520–1522, 2000. [DOI] [PubMed] [Google Scholar]
  • [11].Li XD, Ko TH, and Fujimoto JG, “Intraluminal fiber-optic Doppler imaging catheter for structural and functional coherence tomography,” Opt. Lett, vol. 26, pp. 1906–1908, 2001. [DOI] [PubMed] [Google Scholar]
  • [12].Sivak MV Jr., Kobayashi K, Izatt JA, Rollins AM, Ung-Runyawee R, Chak A, Wong RC, Isenberg GA, and Willis J, “High-resolution endoscopic imaging of the GI tract using optical coherence tomography,” Gastrointest. Endosc, vol. 51, pp. 474–479, 2007. [DOI] [PubMed] [Google Scholar]
  • [13].Bouma BE, Tearney GJ, Compton CC, and Nishioka NS, “High-resolution imaging of the human esophagus and stomach in vivo using optical coherence tomography,” Gastrointest. Endosc, vol. 51, pp. 467–74, 2000. [DOI] [PubMed] [Google Scholar]
  • [14].Li XD, Boppart SA, Van Dam J, Mashimo H, Mutinga M, Drexler W, Klein M, Pitris C, Krinsky ML, Brezinski ME, and Fujimoto JG, “Optical coherence tomography: Advanced technology for the endoscopic imaging of Barrett’s esophagus,” Endoscopy, vol. 32, pp. 921–30, 2000. [DOI] [PubMed] [Google Scholar]
  • [15].Cobb MJ, Liu XM, and Li XD, “Continuous focus tracking for real-time optical coherence tomography,” Opt. Lett, vol. 30, pp. 1680–1682, 2005. [DOI] [PubMed] [Google Scholar]
  • [16].Liang C, Sung KB, Richards-Kortum RR, and Descour MR, “Design of a high-numerical-aperture miniature microscope objective for an endoscopic fiber confocal reflectance microscope,” Appl. Opt, vol. 41, pp. 4603–4610, 2002. [DOI] [PubMed] [Google Scholar]
  • [17].Myaing MT, MacDonald DJ, and Li XD, “Fiber-optic scanning two-photon fluorescence endoscope,” Opt. Lett, vol. 31, pp. 1076–1078, 2006. [DOI] [PubMed] [Google Scholar]
  • [18].Tumlinson AR, Povazay B, Hariri LP, McNally J, Unterhuber A, Hermann B, Sattmann H, Drexler W, and Barton JK, “In vivo ultrahigh-resolution optical coherence tomography of mouse colon with an achromatized endoscope,” J. Biomed. Opt, vol. 11, pp. 064003–1–064003–8, 2006. [DOI] [PubMed] [Google Scholar]
  • [19].Chen YC and Li XD, “Dispersion management up to the third order for real-time optical coherence tomography involving a phase or frequency modulator,” Opt. Exp, vol. 12, pp. 5968–5978, 2004. [DOI] [PubMed] [Google Scholar]
  • [20].Chen YC, Liu XM, Cobb MJ, Myaing MT, Sun T, and Li XD, “Optimization of optical spectral throughput of acousto-optic modulators for high-speed optical coherence tomography,” Opt. Exp, vol. 13, pp. 7816–7822, 2005. [DOI] [PubMed] [Google Scholar]
  • [21].Yun SH, Boudoux C, Tearney GJ, and Bouma BE, “High-speed wavelength-swept semiconductor laser with a polygon-scanner-based wavelength filter,” Opt. Lett, vol. 28, pp. 1981–1983, 2003. [DOI] [PubMed] [Google Scholar]
  • [22].Lim H, de Boer JF, Park BH, Lee ECW, Yelin R, and Yun SH, “Optical frequency domain imaging with a rapidly swept laser in the 815–870 nm range,” Opt. Exp, vol. 14, pp. 5937–5944, 2006. [DOI] [PubMed] [Google Scholar]
  • [23].Huber R, Wojtkowski M, and Fujimoto JG, “Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography,” Opt. Exp, vol. 14, pp. 3225–3237, 2006. [DOI] [PubMed] [Google Scholar]

RESOURCES