Abstract
The field of Brillouin microscopy and imaging was established approximately 20 years ago, thanks to the development of non-scanning high-resolution optical spectrometers. Since then, the field has experienced rapid expansion, incorporating technologies from telecommunications, astrophotonics, multiplexed microscopy, quantum optics and machine learning. Consequently, these advancements have led to much-needed improvements in imaging speed, spectral resolution and sensitivity. The progress in Brillouin microscopy is driven by a strong demand for label-free and contact-free methods to characterize the mechanical properties of biomaterials at the cellular and subcellular scales. Understanding the local biomechanics of cells and tissues has become crucial in predicting cellular fate and tissue pathogenesis. This Primer aims to provide a comprehensive overview of the methods and applications of Brillouin microscopy. It includes key demonstrations of Brillouin microscopy and imaging that can serve as a reference for the existing research community and new adopters of this technology. The article concludes with an outlook, presenting the authors’ vision for future developments in this vibrant field. The Primer also highlights specific examples where Brillouin microscopy can have a transformative impact on biology and biomedicine.
Introduction
The effect of mechanical forces on tissue development and pathology was first recorded more than 100 years ago in relation to the structural formation of trabeculae bone1. At the smaller scale of a single cell, mechanical forces and changes in viscoelastic properties of intracellular and extracellular environments are as important to the cell function as biochemistry2. It took many decades, however, to acknowledge the role of mechanosensing and mechanotransduction in cell biology — the molecular mechanisms by which cells sense and respond to mechanical stimuli3. Largely, probing cells’ mechanical properties with sufficient resolution and sensitivity, and without causing damage or changing their behaviour, is a very challenging task. The past 30 years have brought tremendous progress in experimental techniques for biomechanical measurements, and it is now agreed that mechanical forces play a crucial role in the development of many pathologies, including cardiovascular diseases and cancers4,5.
Modern techniques for mechanical probing of cells and tissues include deformability cytometry6, optical tweezers7,8, atomic force microscopy (AFM) indentation9,10, optical elastography11, Brillouin microscopy12–15 and other methods based on optical, mechanical and acoustic principles and their combinations (for example, picosecond ultrasonics16 and time-domain opto-acoustic techniques17,18). Brillouin microscopy, the focus for this Primer, has the advantage of being a non-contact method capable of characterizing local micromechanical properties in biomaterials, compared with compression and rheology tests, traditionally used to measure volume-averaged elastic moduli (Fig. 1a). Brillouin microscopy relies on the principle of Brillouin light scattering (BLS), a process of inelastic scattering in which light interacts with gigahertz acoustic waves at the frequency . Because of this interaction, the frequency of the scattered light undergoes a shift by exactly the acoustic frequency (Fig. 1a). The detection of such frequency shift, called the Brillouin frequency shift (BFS), and its spectral line shape can lead to valuable insights into the local viscoelastic properties of biological samples19. As a label-free and contact-free method, Brillouin microscopy is suitable for in vivo and in situ 3D imaging of a broad range of biological systems19–22.
Fig. 1 |. Principles of Brillouin light scattering microscopy.

a, A conventional compression test (bulk properties) and Brillouin microscopy (local properties) for measuring mechanics in biomaterials. In Brillouin microscopy, a focused light beam is scattered by the gigahertz pressure waves producing () Brillouin frequency shift (BFS) of the scattered light. b, Brillouin spectrum showing a doublet of Stokes and anti-Stokes peaks, centred symmetrically around the unshifted Rayleigh scattered light. c, Brillouin map for a respiratory cell with colour coding indicating the BFS magnitude. d, Anti-Stokes peak for a location outside (1) and inside (2) the cell demonstrating the relative shift in Brillouin frequency across the sample. BLS, Brillouin light scattering.
In the spectral domain, a spontaneous BLS process in a homogeneous and isotropic medium produces Stokes and anti-Stokes peaks, located symmetrically around the elastically scattered Rayleigh band at frequency as shown in Fig. 1b. Energy transfer from optical to acoustic waves produces downshifted Stokes light , whereas energy transfer from acoustic to optical waves produces upshifted anti-Stokes light . In most hydrated biological materials, including tissues and cells, BFS tends to be close to that of water, within the 5–8 GHz range. The absolute value of the BFS depends on the specific tissue, probing laser wavelength and scattering geometry. The linewidth of Brillouin peaks, , falls between 100 MHz and a few gigahertz, and is also specific to the sample and the scattering geometry13. For a material with a known density and refractive index23, can be linked to the elastic response whereas is linked to the viscous response of the material in the gigahertz frequency range, namely the longitudinal storage and loss moduli. Their ratio, the loss tangent, where tan , is a useful measure that neither depends on the refractive index nor density, and can be used to determine whether the mechanical properties alone influence the changes in and over time14,24,25.
Importantly, the longitudinal modulus probes the ratio of uniaxial stress to strain under a confined condition as opposed to the unconfined deformation used to define Young’s modulus and shear modulus. As such, the longitudinal modulus corresponds to one of many components within the material elasticity tensor (the full tensor of an arbitrary material has a total of 81 components). The differences between the longitudinal, Young’s and shear moduli are discussed in detail elsewhere13,14. BLS experiments are not restricted by measurements of the longitudinal modulus only but are capable of reconstruction of the full elastic tensor by varying the scattering angle as routinely performed for glassy and solid-state materials26–28 and curing epoxy liquids29. Owing to a high level of hydration in biological matter, the frequencies of transverse and quasi-transverse phonons required for tensor reconstruction can fall below a few gigahertz and become particularly difficult to distinguish from the central Rayleigh peak28. Scanning the scattering angle at each measurement point adds additional time, making the full tensor reconstruction impractical for the current state-of-the-art Brillouin microscopy. Nonetheless, a few successful demonstrations of elastic tensor measurements in biological materials using BLS have been reported for collagen and silk fibres25,30,31.
To resolve the shape of the Brillouin peak, an instrument resolution well below is required. Pump–probe heterodyne techniques based on stimulated Brillouin scattering (SBS) achieve high spectral resolution <100 MHz, in theory only limited by the laser bandwidth32,33. Interferometric techniques using multi-pass Fabry–Perot spectrometers34 or highly dispersive virtually imaged phase arrays, or virtually imaged phased arrays (VIPAs)35, reach spectral resolution of 0.1–0.5 GHz.
Brillouin microscopy is sensitive to the local mechanical properties of materials. In heterogeneous samples, Brillouin spectra acquired at different points within the sample allow to build 2D and 3D images where and are used as contrast methods (Fig. 1c). The Brillouin images can depict the sample structures, help characterize tissue morphology32,36,37, and visualize cell aggregates38–41 and subcellular compartments22,42–46. These achievements make Brillouin micro scopy a promising technique in many applications including biomedical diagnostics47,48, mechanobiology49–52, tissue engineering53,54 and biomaterial characterization55–57.
A peculiarity of Brillouin microscopy lies in the effective spatial resolution of the mechanical images. Unlike far-field optical microscopy, where the Abbe diffraction limit defines the reachable longitudinal and transversal spatial resolutions when the optical configuration is known, in Brillouin spectroscopy the light interacts with acoustic phonons that are extended collective excitations. Thus, both optical and acoustic fields need to be considered to define the spatial resolution of Brillouin images58 (see Fig. 2 for an explanation of different spatial scales in Brillouin microscopy). The coherence length, , estimates the characteristic size over which the density fluctuations probed by the BLS process are correlated. is a sample-dependent parameter that can be estimated directly from the Brillouin spectra through the peak width, , and position, , knowing the phonon wavelength , so that (ref. 59). ranges from tens of microns in low dissipative solids to a few microns in high dissipative soft matter. When the coherence length is larger than the voxel of the focused optical beam, it represents the lowest reachable limit in the spatial resolution60. This limit influences the longitudinal as well as the transversal spatial resolution. In particular, the spatial extension of the phonons will be oriented with respect to the optical axis depending on the exchanged wave vector . Hence, the angular distribution of the exchanged vector influences the longitudinal and transversal spatial resolution: the use of high numerical aperture objectives can increase the longitudinal spatial resolution but can occur to the detriment of the transversal one60. This intrinsic limit can be overcome in the specific case of phonon confinement, which occurs when the sample morphology contains structures characterized by significant acoustic mismatch with their environment and whose characteristic size is smaller than . In this instance, BLS offers insights into the mechanical characteristics of nanostructures, for example in studies with nanoparticles61,62 or thin films59,63 where dimensions are comparable with the phonon wavelength ().
Fig. 2 |. Length scales of the Brillouin light scattering process.

The principal length scales that drive the spatial resolution of Brillouin spectroscopy. The scattering volume, , which depends on the optical components (microscope objective and laser wavelength), the coherence length, , which is a sample-dependent parameter that defines the so-called acoustic volume and, finally, the wavelength of the probed phonons, . Adapted with permission from ref. 60 © The Optical Society.
The efficiency of the spontaneous BLS process depends on the material but is generally very low. For example, at room temperature the ratio between the number of inelastically scattered photons and the number of incoming photons is approximately one in a billion64. Additional losses within the optical system and collection optics mean that approximately 105 photons per second of 1 mW of incident visible laser power will be scattered and reach the detector64. The overall detected power typically is no larger than a nanowatt as each visible photon carries only 10−19 J of energy. Therefore, high quantum efficiency scientific-grade cameras and detectors should be used for Brillouin microscopy based on the spontaneous scattering process, preferably with electronic amplification such as electron-multiplying charge-coupled device cameras. Brillouin microscopy data typically have a low signal-to-noise ratio (SNR). One way to increase the SNR of the measurement is to collect the signal over a longer acquisition time. For confocal microscopy applications, however, this is too slow as many thousands of points within the sample need to be measured to complete the full sample scan. SBS methods can be one possible solution, as the efficiency of the stimulated scattering process is largely enhanced as compared with spontaneous scattering32,33,65–67.
This Primer introduces the main methods of Brillouin microscopy and discusses the advantages, limitations and future directions in this vibrant field of research. Experimentation describes existing approaches to Brillouin microscopy, including spontaneous and SBS techniques, as well as practical aspects of the experimental set-ups including the requirements for the laser source, spectral filtering and biological sample preparation. The Results section showcases cell mechanobiology demonstrations and discusses Brillouin microscopy data processing approaches. Applications highlight the main areas of biomedicine and biology where Brillouin microscopy has already made significant impact or is expected to bring more advances in the future. The remaining sections address issues with reproducibility, data storage and deposition, challenges of Brillouin microscopy methods and the outlook for this field of research.
Experimentation
Depending on the nature of the biological sample and the intended application, various approaches to Brillouin microscopy can be adopted. The set-up for Brillouin microscopy can rely on either a single laser source, such as in spontaneous BLS, or multiple lasers, as seen in stimulated and impulsive Brillouin microscopy. In the following sections, the key considerations for assembling each type of Brillouin microscopy system are described, with a summary of the parameters presented in Table 1 and Fig. 3. Additionally, techniques for spectral filtering to eliminate unwanted light are described with comprehensive guidance for preparing biological samples, ranging from hydrogels to multicellular tissues and organoids.
Table 1 |.
Requirements for the laser source
| Technique | Laser type | Output power/energy | Spectral characterization |
|---|---|---|---|
|
| |||
| Spontaneous Brillouin microscopy12,20–23,28,38,40,41 | Single longitudinal mode laser | >50 mW | Linewidth: <10 MHz Spectral purity: >60 dB Frequency stability: <0.2 GHz h−1 |
| Continuous-wave SBS32,33,65 | Tunable continuous-wave lasers | Pump: >180 mW Probe: >40 mW |
Tuning range: >2 GHz Linewidth: <10 MHz Preferably locked to a frequency standard |
| Pulsed SBS67,101 | Tunable continuous-wave lasers with an additional electro-optic or acousto-optic modulator to form pulses | Pulse duration: 10–100 ns Repetition rate: 0.5~1 MHz Peak power: 0.1–1 W (pump) 10–100 mW (probe) |
Tuning range: > 2 GHz Linewidth: <10 MHz Preferably locked to a frequency standard |
| Impulsive SBS102,104,105 | Pulsed laser (pump) | Pulse energy: >10 μJ Pulse duration: ~1–10 ps |
Energy stability: <2% rms |
| Continuous-wave laser (probe) | Power: >40 mW | Single frequency operation Linewidth: <10 MHz |
|
SBS, stimulated Brillouin scattering.
Fig. 3 |. Characteristic parameters of Brillouin microscopy experiments.

The acquisition time per pixel, (seconds) and the lateral imaging resolution, (microns). Parameters are based on the following experiment type: tandem Fabry–Perot interferometer (TFPI)36,38,53,54,57,76,78,139,155,187, virtually imaged phase array (VIPA) spectrometer12,19,20,22,24,41,42,52,150,165, line-scanning VIPA spectrometer (Line)37,39, stimulated Brillouin scattering (SBS) microscopy32,67,100 and impulsive stimulated Brillouin scattering (ISBS) microscopy102,103. Further details on the sample type are presented in Supplementary Table 1.
Spontaneous Brillouin scattering techniques
Confocal microscopy.
Brillouin spectroscopy has been traditionally performed with tandem Fabry–Perot (TFP) etalon scanning interferometry68. TFP-based Brillouin spectroscopy has represented the BLS experiment’s workhorse for decades and still remains in widespread use for applied physics, material science69 and environmental sensing70. Furthermore, in biomedical applications, TFP-based Brillouin spectroscopy has been reported for living cells38, collagen fibres30,71,72, hydrogels53,54,57,73, the cornea and crystalline lens of the eye74–76, biofilms77,78 and bones36,79,80. Owing to the scanning operation modality, TFP spectrometers require long acquisition times. The fastest acquisition time reported using a TFP system to date requires 512 ms per single spectrum78, with typical measurement speeds being tens of seconds depending on the transparency of the sample and imaging volume36,57,73,81. This characteristic makes it unsuitable for extensive characterization of a large area and/or large number of samples. Non-scanning Fabry–Perot spectrometers can shorten the acquisition time by removing the frequency scanning operation79,82, but are limited in throughput as a significant portion of the light is lost in reflection. Introducing a VIPA, a different non-scanning etalon, fundamentally improves the Brillouin spectrometer speed (Fig. 4a). The back surface of a VIPA etalon is fully reflective, other than for a small input anti-reflection window, so that no light is lost in reflection resulting in increased throughput by approximately the etalon’s finesse35,82. VIPA-based Brillouin confocal microscopy was first introduced in 2008 (ref. 12) and has been further optimized in the past 10 years to improve power throughput and noise rejection ability83–88. At present, the best Brillouin confocal microscopy systems exhibit an acquisition time ranging from 20 to 50 ms per pixel/spectrum within the shot-noise limited regime20,42,52,89,90.
Fig. 4 |. Brillouin imaging technology.

a, Confocal Brillouin microscopy. b, Line-scanning Brillouin microscopy (LSBM). c, Continuous-wave stimulated Brillouin scattering (SBS) microscopy. d, Impulsive stimulated Brillouin scattering (ISBS) microscopy with heterodyne configuration. , quarter-wave plate; AOM, acousto-optic modulator; ND filter, neutral density filter; PBS, polarized beam splitter; PC, personal computer; PD, photodetector; TG, transmission grating; VIPA, virtually imaged phased array.
Line-scanning microscopy.
Line-scanning Brillouin microscopy (LSBM) uses signal multiplexing, similar to light-sheet fluorescence microscopy91,92. In LSBM93, one objective lens generates an illumination beam line, and the other collects the scattered light at 90° geometry for signal analysis of a multiplexed spectrometer (Fig. 4b), which enables simultaneous imaging and single-shot spectral analysis of hundreds of points with much lower irradiance dosage compared with confocal Brillouin microscopy. Therefore, the acquisition speed of the LSBM in biological samples can be more than one order of magnitude faster compared with confocal Brillouin microscopy, or up to 1 ms per pixel37,39. The multiplexing of the spectrometer is accomplished by aligning the illumination beam line perpendicular to the dispersion axis of a VIPA etalon. The lateral optical resolution of the LSBM is dictated by the numerical aperture of the collecting objective lens. The axial resolution is determined either by the size of the illumination beam or the collection objective’s depth of field, whichever is the smallest. In practice, LSBM can be implemented in different forms. For instance, the illumination line can be generated by axially scanning the focus position with a tunable lens37; and a dual-line configuration has been developed to mitigate artefacts induced by the heterogeneous refractive index of a thick sample39. In addition, an epi-line geometry has been implemented for embryo samples37.
Stimulated Brillouin scattering techniques
Stimulated Brillouin scattering microscopy.
Stimulated Brillouin scattering (SBS) microscopy operates on the principle of stimulated light scattering, which occurs owing to induced variations in material density caused by optical waves. In SBS microscopy, two counter-propagating, continuous-wave, pump and probe laser beams (Fig. 4c), slightly detuned in frequency to match the BFS of the material , are focused into a joint point in the sample to stimulate a longitudinal acoustic wave with frequency . The acoustic wave, in turn, leads to a travelling grating of dielectric variations. Diffraction from this moving grating is detected as an increase (decrease), or gain (loss), in the probe intensity light at lower (higher) frequency. The gain magnitude is proportional to the pump power and influenced by material param eters such as mass density66. The SBS spectrum is acquired by scanning the pump–probe frequency detuning over (refs. 32,33,65,94–96). The notch filter is a rubidium gas cell used to suppress reflected stray light from the pump beam (Fig. 4c). An alternative approach to SBS microscopy that uses only a single pump laser and applies a heterodyne detection scheme has also been demonstrated97.
To improve temporal resolution significantly, the SBS gain/loss can be recorded only at . This is possible when preliminary knowledge of the sample properties is available, for example the range of BFSs66. Spectrum acquisition times down to 2 ms over 2 GHz in water and 20 ms over 4 GHz in living specimens are practical, resulting in adequate spectrum SNRs (~27–30 dB with overall incident power of ~230–265 mW). A larger frequency scanning range in biological samples may be required for a reliable multi-component fit analysis32,98. Spectral resolution similar to that in TFP-based Brillouin microscopy (~100 MHz) can be achieved in SBS microscopy and is determined by the laser linewidth, the frequency scanning-induced dispersion and the aperture-induced spectral broadening99. SBS images are obtained by raster scanning the sample through the focus and measuring the SBS spectrum or the SBS gain/loss at at each image pixel. The 3D spatial resolution of the SBS image is determined by diffraction and the non-linear excitation of SBS. For reduced incident power, SBS microscopy with squeezed light100 or quasi-continuous-wave light has also been demonstrated67,101. Quasi-continuous-wave SBS, also known as pulsed SBS, is based on external modulation of the continuous-wave pump and probe by an acousto-optic modulator to form optical pulses of 10–100 ns. Depending on the pulse train duty cycle and the pump laser peak power, the overall radiation dose on the sample can be reduced by at least two orders of magnitude67,101.
Impulsive stimulated Brillouin scattering microscopy.
In impulsive stimulated Brillouin scattering (ISBS) microscopy, instead of using two continuous-wave laser beams, an optical pulse from a laser is divided into two pump beams focused into a joint point in the sample to produce a transient grating through the interaction between the pulsed light and the material (Fig. 4d). The generated acoustic wave is measured by the diffraction of a probe light that is incident to the focal point at the Bragg angle102–108. The probe light can be either continuous-wave or a delayed pulse depending on the time constants of the generated acoustic waves and the detection electronics107,108. Improved detection sensitivity is obtained by first mixing the diffracted light and the reference probe beam, and then measuring the intensity of the beat signal106.
Requirements for the laser source
To achieve high-quality Brillouin microscopy measurements, the laser wavelength, power, spectral bandwidth and noise characteristics should be considered. The important laser characteristics for each type of Brillouin microscopy measurements are summarized in Table 1. Wavelengths within the biological transparency region (650–800 nm) are sought to be ideal to avoid sample photodamage88. Noteworthy, the efficiency of the scattering process declines rapidly with the light wavelength (), presenting a trade-off between the laser power and acquisition time. As with other optical microscopy methods, the penetration depth of Brillouin microscopy is limited by the sample absorption and multiple scattering and typically falls within the range of a few hundred microns.
Rayleigh band filtering
Brillouin signals are intrinsically weak and require additional filtering to be detected. Various strategies, passive or active, have been proposed for filtering unwanted light in the Brillouin system, which are generalized to the Rayleigh band as they are difficult to distinguish spectrally. Passive filters, commonly used in other spectroscopy methods for their simplicity and stability, are challenging to implement for Brillouin microscopy, owing to the requirement of a narrow notch/edge width (<10 GHz). Thin film technology is limited to around 100 GHz in width and cannot be achieved passively109. To overcome this, atomic/molecular filters based on gas absorption bands can be used, providing an additional 50 dB of extinction. However, the availability of these filters depends on the materials used, although popular bands such as 780 nm can be targeted110. It is important to note that some absorption agents may cause residual absorption of Brillouin signals, leading to spectral asymmetry and negatively affecting data interpretation. An alternative passive filtering design, based on spatial frequency filtering by using a Lyot stop87, is simple and readily available, but has a limited 20 dB extinction.
Active optical filtering strategies have also been successful in providing extra spectral contrast for Brillouin microscopy. A common-path interferometric notch filter is another approach that can help reach 50 dB extinction111. The device’s long-term stability is achieved by using feedback closed-loop control and adjusting the glass positioning adaptively. To suppress amplified spontaneous emission, Bragg gratings and Fabry–Perot etalons can be combined, achieving ~90 dB of amplified spontaneous emission suppression37,39. Spatial separation of unwanted light using multiple dispersive elements and filter masks is another well-established approach, such as introducing another VIPA in the spectrometer placed orthogonal to the first, resulting in an extra 30 dB of extinction with a well-designed mask112. Similar blocking effects can be achieved in the Fourier domain44.
The active strategy used in TFP set-ups employs the multi-pass configuration: the light passes three times through each of the two Fabry–Perot interferometers, which synchronously perform a scan. The movement of the mirrors and their alignment are actively controlled in real time by piezoelectric transducers. Such a spectrometer is characterized by high resolution and high contrast: the newly developed configuration, TFP-2 HC, reports the record-high contrast of 150 dB (ref. 78). With the emergence of high-throughput LSBM configurations37,39, it may become necessary to combine various passive and active filters. Therefore, a balanced trade-off between optical throughput and overall extinction ratio should be considered during experimental design.
Sample preparation and considerations
Hydrogels.
Hydrogels are commonly used for designing 3D cell and tissue models as they mimic the natural extracellular matrix and provide cells with mechanical support, hydration and delivery of nutrients. Owing to relative simplicity and ability to control the gel fabrication process, this biological material presents a viable model for Brillouin microscopy benchmarking and standardization15. For example, efficient tuning of mechanical properties (Fig. 5a) was demonstrated in gelatin hydrogels53,113. Additionally, hydrogel beads can be used for characterization of the sample heterogeneity114–116. Important aspects to consider in the preparation of hydrogel models for Brillouin microscopy imaging are time-dependent swelling and non-uniform gel hydration117. The incubation of hydrogels in media can cause the gel’s polymer network to swell and a dynamic increase of the local hydration in the 48 h post incubation. Furthermore, cross-linking of gels and tissues results in material stiffening and reduction in local hydration. The effect of hydration on the BFS and linewidth has been investigated in various fibrous materials30, gelatin gels118, hydrogels117,119,120 and human corneas121,122. The impact of hydration changes on the longitudinal modulus and BFS has been elucidated using a two-phase model117,121. It was found that the dehydration process occurring in collagen gels displays the same behaviour as in the glass transition process, for example the phase transition process in amorphous materials from a liquid-like to a solid-like glassy state. In this case, the polymer concentration is the control parameter and plays the same role as temperature or pressure in the analytical description of the transition53 (Fig. 5a). It is crucial to exercise caution while devising Brillouin imaging experiments to maintain a consistent hydration level. One feasible approach is to permit samples to gradually swell over time until they attain a state of equilibrium hydration.
Fig. 5 |. Representative data obtained using Brillouin technology.

a, Brillouin frequency shift (BFS) and linewidth of gelatine versus polymer volume fraction, acquired by a spontaneous Brillouin microscope with 532 nm laser source. Red triangles represent theoretical values derived from the viscoelastic fit. b, Brillouin image of a live 3T3 cell, acquired by a spontaneous Brillouin microscope with 660 nm laser source. Blue shadow and dashed line indicate the cellular nucleus. Scale bar: 5 μm. c, Brillouin images of a single cell and three spheroids at different stages of growth, acquired by a spontaneous Brillouin microscope with 660 nm laser source. The human breast cancer line MCF10CA1h was used. Scale bar: 25 μm. d, Brillouin images of cranial neural tube tissue of four embryos at different culturing stages, acquired by a spontaneous Brillouin microscope with 660 nm laser source. Red dashed line outlines the neural plate region. Scale bar: 50 μm. e, Brillouin images of live nematodes at three larval stages (L2–L4) and two adult stages, acquired by a stimulated Brillouin microscope with 780 nm laser sources. Scale bar: 50 μm. Insert is a magnified image of the gonad. Scale bar: 25 μm. f, BFS values for various tissues, acquired by a spontaneous Brillouin microscope with 532 nm laser source: brain tissue (cortex (C), putamen (P), corpus callosum (CC), hippocampus (H)), cartilage tissue (superficial (S), transitional (T), deep (D)) and retina tissue (basal (B) — nerve fibre layer, retinal ganglion cell layer, inner plexiform layer; apical (A) — inner nuclear layer, outer plexiform layer; and photoreceptor cell layer (P) — outer nuclear layer, photoreceptor inner segment, photoreceptor outer segment). HH, Hamburger Hamilton staging; YA, young adult. Part a reprinted from ref. 53, CC BY 4.0. Part b reprinted from ref. 133, Springer Nature Limited. Part c reprinted with permission from ref. 137, Elsevier. Part d adapted from ref. 201, Springer Nature Limited. Part e adapted from ref. 32, Springer Nature Limited. Part f adapted with permission from ref. 130 © The Optical Society.
Single cell.
A single cell typically has a size ranging from a few micrometres to several tens of micrometres, and its mechanical properties are sensitive to changes in the surrounding microenvironment (Fig. 5b). When probing cellular biomechanics using Brillouin microscopy, it is essential to maintain cells in their physiological or culture conditions. A good practice is to keep cells within an on-stage incubation chamber that ensures a suitable environment such as the temperature, humidity, and CO2 and O2 levels. For Brillouin microscopy with an inverted configuration, whenever feasible, the substrate for cell attachment should be thin and optically clear to minimize beam distortion. If appropriate, cells can be seeded on a Petri dish with a glass coverslip bottom (thickness <200 μm).
Owing to the strong back-reflection from the interface of the glass bottom and cell culture medium, noise from non-scattered light can become significant when the focal plane is close to this interface. In such cases, additional background filtering and noise reduction techniques may be necessary to extract the Brillouin signal44,86. Alternatively, the bottom of the dish can be pre-coated by a thin hydrogel layer (<500 μm) to match the refractive index of the medium, thus reducing back-reflection42. However, as cells may adjust their mechanical properties in response to a substrate’s stiffness22, the impact of the hydrogel layer should be carefully evaluated based on the purpose of the experiment. In addition, many standard culture media contain phenol red, which is a pH indicator but can photosensitize cells with strong absorption of visible light. To reduce the risk of phototoxicity, it is recommended to replace the standard medium with a phenol red-free alternative before conducting Brillouin microscopy experiments88.
Tissues, spheroids and whole organisms.
Compared with single cells, multicellular samples such as biological tissues, spheroids (Fig. 5c) and whole organisms have larger size, more heterogeneous distribution of biomechanics and more complicated compositions. Therefore, careful consideration should be given to the trade-off between spatial resolution and acquisition time, sample preparation and orientation and data interpretation. For time-extended experiments, the potential phototoxicity should be carefully considered. Brillouin microscopy has been used for quantifying a broad range of biological tissues ranging in size from ~100 μm to a few millimetres, including ocular tissues (for example, cornea, retina and crystalline lens)76,123–129, central nervous system tissues (such as cortex, putamen, corpus callosum, hippocampus and spinal cord)52,130, muscle, skin, cartilage and early-stage embryonic tissue20,89,131 (Fig. 5d–f). The acquisition speed of the Brillouin microscope can be altered, for example 20–500 ms per pixel, to preserve the biomechanics of the sample throughout the acquisition time and to allow sufficient mechanical details to be captured. However, the size of the mapping area and spatial resolution may become compromised.
Ocular tissue, such as the cornea, is among the outmost layer of an eyeball and possesses excellent optical transparency, allowing for in vivo testing121,132. Notably, zebrafish, especially in the larval stages, are relatively transparent and can be investigated completely in vivo20,52,127–129. Otherwise, biological tissues generally need to be prepared to fit a standard Petri dish or glass slide for ex vivo testing. To avoid alteration of tissue biomechanics, live tissue without fixation or freezing is required, even though the sacrificing of animals and the cutting required can lead to changes in material properties52. Considering the penetration depth of Brillouin microscopy of a few hundred microns133, the region of interest should be as close to the surface as possible. For anisotropic tissue such as the cornea, the orientation of the tissue layer should also be specified134,135. Additionally, Brillouin data can be carefully calibrated against conventional methods such as AFM22.
3D cell culture such as spheroids represents a physiologically more relevant culture system compared with conventional 2D cell culture136. With its ability to provide subcellular resolution in three dimensions without physical contact, Brillouin microscopy is well suited for mapping biomechanics of spheroids (Fig. 5c). In experiments, spheroids can be prepared in different ways: embedding cells into hydrogel, placing cells into a low attachment well or seeding cells on top of soft Matrigel19,38,40,137,138. In confocal Brillouin microscopy, experiments can be conducted in the spheroids’ native culture. Alternatively, LSBM can be used for rapid mapping with a lower irradiation dose39. Here, spheroids must be harvested and embedded in an agarose gel, although this can be overcome when using an inverted microscope37. Brillouin technology can also be used for whole organisms, such as Caenorhabditis elegans32 (Fig. 4e). To acquire biomechanical maps of live C. elegans using stimulated Brillouin microscopy, the sample along with growth medium is transferred to a homemade mount that sandwiches the sample between two coverslips. Two counter-propagating laser beams, namely the pump and the probe, pass through the coverslips and intersect at the common focal plane within the sample, allowing for the acquisition of Brillouin images through the raster scanning of the sample.
Results
Data acquisition and processing
Brillouin line shape analysis in homogeneous and heterogeneous materials.
The mechanical and thermodynamic properties of materials determine the spectral features of Brillouin signals. In a bio-Brillouin framework, the analysis considers the position of and the full width at half maximum of the Brillouin peak associated with the longitudinal acoustic wave. Close to , the spectrum can be described by a damped harmonic oscillator convoluted with the experimental response function :
| (1) |
where is the peak intensity and considers the peak enlargement, owing to the finite spectrometer resolution, and the spread in the exchanged phonon wave vectors, owing to the aperture of the focusing and collection optics36,139. In low dissipation conditions, such as , the damped harmonic oscillator function can be approximated by a Lorentzian peak.
For elastically heterogeneous samples, the final spatial resolution and the Brillouin peak line shape are derived from the spatial extension of the probed acoustic waves60. Depending on the characteristic size of the heterogeneities, , and on the acoustic mismatch among them, two scenarios are possible59. If the scattering volume is filled by heterogeneities characterized by (where is the phonon coherence length) with a small acoustic mismatch, a single phonon can extend across different structures and a single broadened Brillouin peak is expected in the spectrum. In this case, accounts for all of the different phonon attenuation processes and provides an average value of the sample mechanics. On the contrary, when or the acoustic mismatch is very high, the phonons are confined within the hetero geneities allowing to determine their mechanical properties. This latter condition was recently found in a study investigating human bones36,80.
Effect of the variation of refractive index and density on frequency shift.
The BLS measurement output is the frequency shift, , which — in addition to the longitudinal storage modulus — also depends on the refractive index, , and the mass density, , at the measurement location within the sample:
| (2) |
where is the scattering angle and is the laser wavelength. Both and typically scale with local electron density so that the ratio between them is at least constant, or even trivially cancels. For this reason, these two parameters are often not considered, and the BFS is interpreted as a direct proxy of the longitudinal modulus. Further, in some cases the variability of the refractive index across a tissue being measured, such as for zebrafish spinal cord, has directly been quantified using quantitative phase imaging and found to be much smaller than the variability of the BFS52.
However, there are instances where the ratio between and changes within a biological sample, which could lead to misinterpretation of the BFS data. The contribution of some of the most abundant biomolecules found in biological cells, such as proteins, nucleic acids and phospholipids, to the refractive index and density can be considered using a two-substance mixture model140. For each material, there is a so-called refraction increment that can vary by almost a factor of two between proteins and phospholipids, for example. As certain areas within cells have different compositions of these biomolecules, it is necessary to measure both the local refractive index and the mass density explicitly for accurate conversion from BFS to the longitudinal modulus. This is particularly obvious for lipid droplets within adipocytes, which have higher refractive index but lower mass density than water and where the two-substance mixture model cannot be applied, as it can translate to an error of up to 20%46. The 3D refractive index distributions within a cell can be determined by optical diffraction tomography with spatial resolutions of about 100 nm (ref. 141). Mass density can be extracted from the gain in stimulated Brillouin spectroscopy66 or computed from the known refractive index — if the assumptions of the two-substance mixture model are fulfilled and if the local composition of biomolecules is known. The local chemical composition can be either estimated from fluorescence microscopy46 or, even better, Raman spectroscopy78, ideally all in one instrument.
Line broadening and asymmetry of the Brillouin peak.
The asymmetric line shape of the Brillouin peak, accounted for by the damped harmonic oscillator fitting function, derives from the intrinsic attenuation processes present in the material and is particularly evident when investigating highly dissipative samples142. There are additional effects that distort and broaden Brillouin peaks, which require further discussion (Fig. 6).
Fig. 6 |. Ballistic and multiple Brillouin scattering processes.

a, A Brillouin light scattering (BLS) process occurring in a transparent medium, where all scattered photons are ballistic and the scattering geometry defines the exchanged vector . b, A multiple scattering process introduces uncertainty in the value. c, Brillouin spectrum (markers) acquired in the depth of a milk sample: ballistic (solid blue line) and multiple scattering (dot–dash line) contributions are provided. , scattering angle; , wave vector of incoming light; , wave vector of scattered light. Adapted from ref. 144, CC BY 4.0.
The linear dispersion relation links the BFS, , to the exchanged wave vector , with being the sound velocity of the acoustic modes (see Eq. (2)). This implies that each corresponds to a specific . The use of high numerical aperture (NA) focusing and collecting optics introduces a large distribution of possible scattering angles , extending the range of the exchanged wave vectors60,143. The resulting spectral distortions depend on the chosen microscope objective and the scattering geometry used. In back-scattering, where the exchanged wave vector reaches its maximum, high NA optics implies a reduction in the effective , corresponding to the introduction of a low-frequency tail in the peak, which modifies its frequency position and width99,139,143. Here, effective NA is important rather than nominal NA of the microscope objective lens, as the objective aperture can become underfilled, resulting in NA reduction.
The linewidth distortions are even more important when investigating turbid media. In fact, the propagation direction of light inside turbid systems continuously changes so that the external scattering geometry no longer defines the effective scattering angle . Recently, polarization gated Brillouin spectroscopy has been introduced to selectively determine the contributions in the Brillouin spectra, owing to the ballistic and the diffusive photons144 (see Fig. 6a,b). Both the use of high NA objectives and the influence of multiple scattering effects can be individually quantified or simulated (Fig. 6c), allowing for the spectra to be effectively analysed in a suitable manner.
Data pre-processing.
Brillouin microscopy data analysis is non-trivial owing to its high dimensionality. Prior to analysis, pre-processing steps are necessary to enhance data quality and interpretability. Although widely available for biomedical spectroscopy145, there is currently no standardized approach for Brillouin microscopy. In data sets acquired using low-contrast spectrometers, subtracting the Rayleigh band prior to analysis is essential, as it can directly affect the accuracy of the extracted information. Software filtering, such as direct subtraction using a pseudo-Voigt function146, is possible in addition to hardware filtering. Data normalization and scaling are additional pre-processing steps that remove systematic variabilities, allowing for the identification of subtle biological patterns. For example, methods such as standard normal variate transformation have been effective in stabilizing variance for subsequent analyses146. Brillouin microscopy data typically have low SNRs; denoising techniques such as wavelet and maximum entropy reconstruction have thus been proposed147, providing useful spectral signals even from challenging photon statistics (SNR ≥ 1). More recently, the matrix pencil method has shown promise for denoising non-linear Brillouin signals in the temporal domain148. Enhancing the SNR post acquisition may allow for faster imaging speed by sacrificing some photon budget experimentally and restoring it computationally.
Univariate analysis and peak detection.
Although Brillouin microscopy data are intrinsically high-dimensional, they are often sparse in the spectral domain. To analyse the data, peak fitting methods are commonly used to compute the peak position and width, which relate to bulk compressibility and viscosity, respectively. Different profiles such as the Lorentzian function, pseudo-Voigt function and damped harmonic oscillator have been employed to handle various assumptions and experimental set-ups15,146. Directly analogous to the same concept in localization microscopy, successful peak fitting allows super-resolution in spectral localization but is still fundamentally limited by detection and noise models, bound by information theoretic limits. This precision limit has been derived for different experimental parameters149 and should facilitate the performance evaluation of various hardware and software aspects of Brillouin microscopy, such as denoising tools. Interestingly, determining the spectral linewidth is more challenging and less accurate. Once average peak locations and widths are obtained for each spatial location in the Brillouin microscopy data, mechanical contrast images are formed. Univariate statistics can then be applied for hypothesis testing between different regions of interest or samples. For instance, t tests and analysis of variance with Bonferroni false discovery correction have been used to evaluate differences in Brillouin peak pro perties from various samples such as 3D bioprinted hydrogels, dentin and mice arterial sections56,57,150. Fitting, usually by means of least-square algorithms, can however become time consuming, especially for larger Brillouin microscopy data sets consisting of >104 pixels. As such, other approaches such as use of the spectral phasor and moment have also been presented151,152. The spectral phasor is a fitting-free method, widely used in other complex imaging data owing to excellent noise resilience. On the other hand, the moment approach is routinely applied for Raman data and can separate overlapped peaks assuming spectral symmetry. Both methods utilize faster computational processes, providing quicker options for data visualization.
Unsupervised and supervised multivariate analysis.
Considering the subtle spectral variations, it is often advantageous in Brillouin microscopy to analyse the entire spectral domain using multivariate analysis. Multivariate analysis explores the correlations between different spectral variables across the data set (Fig. 7). This approach, widely used in hyperspectral imaging, for example in remote sensing, leverages machine learning capabilities. Multivariate algorithms can be supervised or unsupervised depending on the level of prior knowledge required (Fig. 7). Although various algorithms are applicable to Brillouin microscopy with minimal modification, representative algorithms for high-level image analysis applications are discussed, without delving into their inner workings (more details can be found elsewhere146). Unsupervised analysis finds patterns in input data without any prior knowledge. Principal component analysis (PCA) performs a linear transformation of the input data into a set of basis vectors that are defined by the ranked intra-data variances, in turn projecting Brillouin microscopy data into distinct spectral components. This form of dimension reduction is one of the enabling powers of multivariate analysis, as it provides easy 2D/3D visualization and exploration of data trends, even in large, complex high-dimensional data. One important confounding factor in analysis of Brillouin microscopy data is that a spectral mixture can be acquired from a single spatial location. Spectral unmixing using methods such as minimum volume simplex analysis and non-negative matrix factorization have been proposed to separate overlapped components, also known as endmembers153,154. The number of endmembers dictates the algorithm performance, which can be estimated statistically or manually. The former can be realized by observing the natural groupings in the data and is directly linked to the idea of cluster analysis. -means cluster analysis and other fast density-based clustering methods are efficient for stratifying high-resolution Brillouin microscopy data146,155. Similarly, clustering can also be performed more optimally in a dimensionally reduced latent space by combining with, for example, PCA.
Fig. 7 |. A typical workflow depicting analysis of Brillouin microscopy data with modern machine learning approaches.

The high-dimensional raw data can generally be more easily interpreted through dimension reduction, for example by principal component analysis (PCA) as illustrated. The lower-dimensional representation can be used with unsupervised learning techniques, where natural trends and patterns in the data can be extracted statistically to divide the data into clusters and visualized spatially (segmentation); alternatively, supervised learning approaches may also be used with additional prior information, for instance labels of distinct cellular regions from another reference image. A model can then be trained using these labels with, for example, linear discriminant analysis (LDA), where the best-performing model can then be used to predict the class of unseen pixels or images (classification). With the increase in data quality and quantity, a centralized repository hosting a wide range of Brillouin microscopy data from different detectors, sources and domains can also be envisioned to enable deep learning approaches in the near future. Adapted with permission from ref. 146, Wiley.
Supervised analysis, on the other hand, requires prior knowledge for optimal functioning. It is useful for spectral classification and hypothesis testing, for example distinguishing diseased versus healthy samples. Although still rarely practised, the supervised classification of Brillouin microscopy data was first demonstrated on fat cells using linear discriminant analysis (LDA)146 and more recently adapted to discover spectral markers for the diagnosis of osteoarthritis36. On the back of successful implementations in related fields, the use of supervised deep learning is also anticipated to impact future Brillouin microscopy data analysis, especially in terms of multimodal integration and developing machined learned tools for in vivo applications. In this vein, routine use of shallow learning approaches such as the support vector machine156 should first be adopted to assess their suitability. The corresponding hyperparameters, including the optimal classification features, can then be optimized, which will also shed light on the underlying mechanobiology and discovery of Brillouin-based biomarkers for various applications.
Applications
Cell mechanobiology
The importance of the mechanical interaction between biological cells and the surrounding microenvironment has now been widely recognized157,158. Such interaction is regulated by the mechanotransduction through which cells sense mechanical stimuli from external forces or surrounding stiffness and convert them into biochemical signalling159. Thus, mapping the extracellular and intracellular distribution of the elastic modulus is critical to determine the location of environmental mechanical cues as well as how forces are transmitted within a cell160–162.
Brillouin microscopy holds great promise in mechanobiology to meet the resolution requirements for mapping mechanics at the subcellular scale as well as the experimental settings to investigate cell mechanics in 2D and 3D microenvironments, in tissues or in microfluidics platforms. To date, Brillouin microscopy has already been used to demonstrate actin polymerization and branching22, actin remodelling155, the correlation between intracellular modulus and cytoskeletal properties163 as well as a cell’s adhesion properties164 and intracellular stiffness changes owing to environment perturbations19. At the subcellular level, Brillouin microscopy has demonstrated the ability to characterize nuclear mechanical properties46,50,138 and the viscoelasticity of plant cells’ extracellular matrix165, and, in some cases, has even been validated by gold-standard tests and chemo-mechanical modelling42,166. At even higher resolution, measurements of biomolecular condensates, lipid droplets in adipocytes, and other aggregates within cells and nuclei have been recently reported44–46,167.
Biomedical diagnostics
Ocular disease.
The most mature of the applications of Brillouin microscopy is in ophthalmology, and specifically in the characterization of corneal mechanics for keratoconus and related ectatic disorders (Fig. 8). In the United States168, keratoconus is the leading cause of corneal transplantation169 and the most feared complication of refractive surgery, leading to post-refractive surgery corneal ectasia when undetected170,171. Currently, keratoconus presents clinically as a morphological disorder characterized by the thinning and increase in curvature of the cornea. However, it is widely believed that this clinical presentation is a late manifestation of the progression of the disorder, which probably starts with a mechanical degeneration of the corneal tissue172. Healthy corneas are mechanically strong thanks to tightly packed stromal collagen fibres which maintain shape under the outward intraocular pressure (IOP) (Fig. 8a). By contrast, keratoconic corneas have altered collagen organization173,174 and a reduced number of cross-links175, presenting a focal weakening176 (Fig. 8b). Such weakening is believed to be the main driver of shape change (Fig. 8c). Brillouin microscopy is uniquely capable of measuring the loss of mechanical strength in corneas in vivo21,132 (Fig. 8d). Recent advances in clinical instruments promise great utility in this space for keratoconus diagnostics (Fig. 8e,f), refractive surgery screening as well as therapy monitoring177. Within ophthalmology, another promising application of Brillouin microscopy is related to presbyopia, the loss of the dynamic vision accommodation response, which affects individuals as they age. Currently, presbyopia is treated with spectacles, which do not restore the accommodative amplitude. The issue is widely known to be caused by the stiffening of the crystalline lens leading to a strong recent effort to develop drugs that target lens biomechanics; in this respect, Brillouin microscopy has shown the ability to measure the relevant changes occurring in the lens178, and also in clinical settings179, and thus may represent a critical tool for drug development in this field.
Fig. 8 |. Application of Brillouin microscopy to ocular disease diagnostics and treatment.

a, Healthy corneas have nearly spherical shape to provide needed refraction for proper vision. b, In keratoconus, a focal weakening of the cornea disrupts the mechanical balance with intraocular pressure (IOP), c, Mechanical imbalance leads to corneal thinning and bulging. d, Clinical instrument user interface featuring Brillouin light scattering (BLS) and 3D tracking via pupil imaging and optical coherence tomography (OCT). e,f, Brillouin maps for healthy (panel e) and keratoconus (panel f) corneas reveal focal weakening at an early stage. Part d adapted with permission from ref. 177 © The Optical Society.
Parts e and f adapted with permission from ref. 132, Elsevier.
Cardiovascular disease.
The healthy heart functioning relies not only on the ability of its muscle fibres to contract and relax but also on maintaining the shape and structure of the organ. The way the heart muscle shortens and generates force, regardless of the existing conditions, is determined by the mechanical properties of the heart tissue180. When these properties are disturbed, as often happens in heart diseases, this leads to changes in the thickness of the heart’s walls, stiffness of the tissue and, eventually, malfunction. However, directly assessing these properties at a local level is challenging. Brillouin microscopy was introduced to assess heart tissue’s viscosity and elasticity181: it identified stiffness changes in healthy mouse myocardium ventricles with notably reduced stiffness in the compliant right ventricle. These results correlated strongly with speckle-tracking echocardiography, emphasizing the link between displacement and stiffness in the heart’s regions181. In a different study, the micromechanical properties of a thin cap fibroatheroma were assessed using Brillouin microscopy150. A fibroatheroma is a particular stage in the progression of an atherosclerotic plaque, characterized by a thickened fibrous cap, covering a lipid-rich core182. Fibroatheromas are considered high-risk plaques because they have a greater likelihood of developing complications such as rupture or erosion183. Thus, a method for precise mechanical characterization of the fibroatheroma’s stiffness and structure can prove to be invaluable for prediction of the plaque’s fate. A strong inverse correlation between BFS and the local accumulation of lipid was demonstrated together with a direct relationship between frequency shift and the distribution of collagen fibres150. Collectively, these results support Brillouin microscopy as a method to optically measure stiffness and structural composition in atherosclerotic vessel sections.
Bone, dentin and cartilage disease.
Specialized forms of connective tissue, bone and cartilage, are essential for providing structural support to muscles and organs within the human body. Their structure and mechanical properties are critical for their respective functions, whereas disruptions in these properties are associated with injuries and degenerative conditions such as arthritis and osteoporosis. Understanding of bone remodelling processes, especially in the context of implants, is limited owing to the absence of non-invasive methods to assess implant stability and osseointegration during the healing phase after implantation. The potential of Brillouin microscopy in distinguishing between mature and newly formed bone following implantation was recently suggested184. A comparative investigation between BLS and scanning acoustic microscopy was conducted to determine the sensitivity of BLS in detecting anisotropy and decalcification, hallmarks of bone diseases185. In a rabbit model, bone healing was quantified in terms of BFS and linewidth changes, both parameters reflecting the pace of bone generation and healing at the defect site186. The contribution of collagen to cartilage tissue mechanics in the regenerating axolotl model131 and the pathogenesis of osteoarthritis and disease-related early changes in articular cartilage36,53,81,187 have also been studied using Brillouin microscopy. Degradation of cartilage extracellular matrix typically goes hand in hand with changes in tissues’ hydration, which can be traced with Brillouin microscopy with high precision187.
Cancers and drug discovery
The mechanical properties of individual cancer cells and tumour tissue are some of the best studied topics in biomechanics. Cancer cells have almost consistently been found to be more compliant than their healthy counterparts, and increasingly so with metastatic competence188. This has led to the notion that the increased deformability of cancer cells is required for their migration and circulation through the body. However, to date there is no convincing evidence that this is the case. The problem is that most studies so far have been on either cell lines or cells isolated from tissues189,190 and not on cells behaving in situ. Also, tissue mechanical measurements on the cellular scale have generally been done with excised cancer tissue biopsies191. This is a technological short-coming that is starting to be addressed by Brillouin microscopy38,40. The ability to measure single cells in complex microenvironments has enabled unique studies about the adaptability/plasticity of cell mechanical properties in different circumstances to be performed. This is particularly relevant for investigations into the metastatic cascade where the mechanical microenvironments and mechanical challenges vary significantly192,193. During extravasation (cells leaving the bloodstream to get into tissues), the Brillouin longitudinal modulus of cells and their nuclei was markedly lower in the transmigration phase194, as predicted by chemo-mechanical models195. Ascitic fluid currents, a hallmark of poor prognosis in ovarian cancer, were shown to reduce the tumour cell modulus196, indicating higher tumour aggressiveness. In confined migration, the nuclear modulus was inversely correlated with the confined migration speed, indicating nuclear stiffness as a critical mechanical barrier for cell motility197.
As changes in tumour cell mechanics are required for successful metastatic capability, a promising anti-tumour strategy is to look for ways to stiffen cancer cells artificially. For this, screens of US Food and Drug Administration (FDA)-approved drug libraries or chemical compounds with a mechanical read-out of cell or nuclear stiffness could be performed. Even though Brillouin spectroscopy is relatively slow, Brillouin flow cytometry of cell nuclei at rates of 200 cells h−1 has already been demonstrated166.
Tissue morphogenesis
Tissue biomechanics is deeply involved in morphogenesis198–200. To date, Brillouin microscopy has been used for understanding the role of mechanics in various aspects of tissue morphogenesis, including spinal cord growth and repair after injury52, eye development127,128, neural tube closure (NTC)90,201, the development and regeneration of limbs131, and early development of the kidney202. The model organisms analysed by Brillouin microscopy include zebrafish52, chick201, axolotl131, mouse90,202, C. elegans32, Drosophila37 and Nematostella51.
Zebrafish are among the most common model organisms used for imaging, partly because of their optical accessibility. Using Brillouin microscopy, the tissue modulus of living zebrafish larvae in all anatomical planes has been mapped during development (3–5 days post fertilization) and after spinal cord injury52. The results demonstrate that Brillouin microscopy has sufficient sensitivity to distinguish the mechanical difference between tissues from distinct sites. Importantly, the results reveal that the spinal cord tissue shows a significant decrease in BFS right after injury. Moreover, the longitudinal measurement shows that the BFS of the lesion site gradually increases during the 2 days post injury but remains lower than that of adjacent uninjured tissue, which coincides with the decreased cell-body density in the lesion site.
Brillouin microscopy has also been used for quantifying tissue mechanics during NTC — a developmental process responsible for vertebrate neurulation. If failed, it is known to be a primary cause of birth defects. Preliminary studies using co-registered Brillouin microscopy and optical coherence tomography (OCT) suggested that the BFS of the neural tube tissue of mouse embryos is higher following closure compared with when it is open90,203. Recently, time-lapse Brillouin imaging of neural tube tissue during NTC was conducted by combining confocal Brillouin microscope with a modified ex ovo culture of chick embryos201. The longitudinal experiment observes the increasing BFS and thickness of the neural tube during NTC, which is concurrent with the bending of the neural plate. This study provides a platform for further understanding the underlying cellular and molecular mechanisms that regulate the procedure of NTC as well as neural tube defects.
In addition, Brillouin microscopy has been employed to understand the role of biomechanics in the packing of tip domains for early development of the kidney using a mouse model202. Results show a strong correlation between cap mesenchyme cell BFS and geometric patterning, suggesting mechanical changes during branching and expansion. Early nephrons appear to have higher BFS than cap mesenchymal cells.
3D bioprinting and tissue engineering
The advent of 3D bioprinting has transformed biofabrication, providing an accurate, cost-efficient and relatively straightforward approach for creating in vitro living systems on a large scale204,205. Quality control is vital, owing to complex designs of bioprinted constructs. Brillouin microscopy presents a promising avenue for assessing the micromechanical properties of bioprinted models, either during or post fabrication73. Complex hydrogel structures created through drop-on-demand bioprinting were imaged using Brillouin microscopy showing high sensitivity to small (5%) variations in the gel’s cross-linking and solid component57. This study also identified non-uniform distribution of mechanical properties across 3D prints despite constant printing conditions as the result of a non-homogeneous cross-linking process and non-uniform hydrogel swelling57. 3D mapping of micromechanical properties is also attractive for the development of artificial tissue models aimed at tissue regeneration applications in spinal cord injury and heart disease54,73. Both pathologies lead to fundamental alteration of mechanical properties in native tissues through formation of fibrous scarring, leading to the patient’s disability or organ failure. Brillouin microscopy was successfully utilized for characterization of 3D bioprinted neural cell models73 and cardiac spheroids54, helping assess the models’ long-term mechanical stability under physiological conditions.
Reproducibility and data deposition
The reproducibility and data quality of Brillouin spectra can be influenced by various factors. The principal factors together with the best practices for mitigating these effects are discussed below.
Factors affecting reproducibility
Modifications and variability in sample viscoelastic properties.
The mechanical properties of viscoelastic materials depend on the physical conditions of the samples, such as temperature, hydration and pressure. Moreover, the mechanical properties are frequency dependent; thus, to compare the measurements obtained in different laboratories, the chosen experimental conditions, together with the power and laser frequency, must be clearly indicated. High-power continuous-wave illumination (>200 mW) in semi-transparent samples can lead to thermal effects, especially for the illumination wavelengths outside the biological transparency window (600–800 nm). Such prolonged exposures could cause the local temperature to rise97, but the BFS caused by thermal effects is insignificant (at lower powers and for 530–800 nm laser wavelength) as it falls well below the measurement spectral precision.
When the investigation includes complex biological samples such as cells and tissues, statistical analysis is needed. Acquiring a large number of spectra on multiple samples is necessary to estimate the confidence of the results, considering the biological variability.
The optical layout chosen to acquire Brillouin spectra.
The optical configuration, including the scattering geometry, affects the final shape and position of the Brillouin spectra. The modifications are particularly relevant in microscopy arrangements using high NA objectives. In this case, the spread in the collected exchanged wave vector, q, can modify the position and the linewidth of the Brillouin peak. To extract the relevant parameters from the spectrum, the measure or the theoretical evaluation of the q-spread has to be used to de-convolve the raw data.
The use of different Brillouin spectrometers.
The spectral resolution and contrast of the spectrometer/interferometer can modify the measured spectra in terms of the SNR and line shape. For isotropic samples, the deconvolution with the response function of the spectrometers has to be used to obtain the corrected values of and . In general, to account for the spread in the collected exchanged wave vector and the finite spectral resolution of the spectrometer, the spectrum of a material with a negligible intrinsic broadening, with an expected Brillouin peak approximal to a Dirac delta function (for example, vitreous silica), can be used for the direct evaluation of the Brillouin set-up response.
Data deposition
Although the standard unit in the reporting of BFS has been proposed15, the standardization in the data treatment and the creation of data repositories remains a challenge in the bio-Brillouin field. To date, Brillouin spectroscopy applied to biological samples is still not used routinely and most laboratories put together home-built set-ups in which the frequency response, the spectral resolution and the contrast are characteristics of each individual system. Additionally, different models are currently used to fit the Brillouin spectra. The importance of the standardization of the data treatment and the creation of a data repository was recently highlighted in the community, and is one of the main goals for the newly founded Bio-Brillouin Society. The Bio-Brillouin Society is working to build the first repository for raw Brillouin data acquired on simple samples by the different set-ups. The required information is the wavelength of the laser beam, the scattering geometry and the used set-up. Moreover, the laser power, the sample condition (such as temperature and hydration) and the microscope objective should be specified in order to obtain comparable data.
Limitations and optimizations
The first obstacle towards wide adoption of Brillouin microscopy is the slow acquisition time compared with other microscopy techniques. Spontaneous Brillouin microscopy is now close to the fundamental limit set by the weak interaction strength between incident photons and thermal phonons inside the medium; further improvements in imaging speed can be achieved with multiplexing speed37,39. In non-linear Brillouin scattering, much stronger Brillouin signals are available because two light beams drive material phonon excitation which, in turn, scatters light efficiently. Non-linear Brillouin technologies are established in material research and several laboratories have now translated stimulated Brillouin scattering (SBS) to biomedicine with frequency-domain or time-domain approaches. However, current SBS spectrometers have not provided dramatic speed-up for biomedical measurements so far because of small interaction volumes and photodamage limits as well as the lack of suitable sources for SBS in biological media. Improving on SBS performance represents a great opportunity for the field to grow.
Instrumentation, cost, size and complexity of operation are also a significant hurdle for widespread adoption. The currently used spontaneous Brillouin microscopes cost more than US$100,000 to run, occupy an optical table and are sophisticated instruments to operate at peak performance; SBS microscopes may be even more costly and complex to optimize and operate given the double-access alignment requirements. On this front, several companies are trying to overcome the issue and individual laboratories are aiming to make intuitive software and alignment protocol instructions more readily available and accessible to non-experts84,133. Advances in nanofabrication and optical fibre technology can help reduce the complexity and footprint of Brillouin systems further, as recently demonstrated for fibre-integrated Brillouin imaging206 and chip-based laser light filtering207.
Brillouin microscopy is currently under-utilized as a label-free technique providing mechanical contrast. The scientific reason behind this is probably the unconventional nature of the mechanical information compared with gold-standard biomechanical measurements. The high-frequency longitudinal modulus46 occupies a different scale of magnitude values compared with the traditional Young’s modulus, which can be confusing for end users and can generate scepticism. Interestingly, this challenge has spurred high-quality research in the past decade, focusing on the fundamental relation between traditional and Brillouin mechanical signatures that goes beyond the empirical correlations used in early studies117,120; the search for the biological determinants of Brillouin mechanical signatures; and a direct connection between Brillouin signature and end-user diagnostic applications, which in some fields, for example ophthalmology, has provided remarkable success in the clinic132,208.
Brillouin microscopy is highly directional, probing the axis of the modulus in the same direction of the light (in a traditional epi-detection configuration). On the one hand, this means that uniquely Brillouin microscopy can characterize the full elastic modulus tensor in anisotropic materials25,134,135; on the other, in anisotropic biological materials, mapping only one axis — as in the majority of Brillouin biological studies demonstrated to date — does not provide a comprehensive characterization and may actually provide confusing results. This is only a technical challenge for the field; once the speed of the instruments grows, being able to create multiple maps at different angles will become feasible within practical experimental settings.
Outlook
Exploiting the optical advantage
Brillouin microscopy’s biggest advantage, which it shares with other forms of optical elastography48, is that it is an optical technique that permits the quantitative exploration of material properties inside cells, tissues and organisms. This is an option not previously available with other more traditional methods to measure mechanical properties in biology209. Whereas Brillouin microscopy exploits the inherent, thermally driven density fluctuations, other optical elastography techniques externally apply a known, often oscillatory deformation at given frequencies. This has important consequences. The option to choose excitation frequencies means that the mechanical response can be tested in a shear mode and on timescales that are closer to those commonly used in non-optical techniques, such as AFM nanoindentation (0.1–10 kHz), which is far away from the gigahertz frequencies relevant in Brillouin microscopy. On the other hand, gigahertz frequencies translate to higher spatial, subcellular resolution, whereas other optical techniques often struggle to achieve even cellular resolution. It is this high spatial resolution that permits the investigation of intracellular mechanical heterogeneities. In particular, nuclear mechanics measured inside intact cells is hardly possible with other means, and very important in the context of cell migration in confined spaces where nuclear deformability is the rate-limiting factor. Also, the dynamic mechanical changes of immune and cancer cells during transendothelial and interstitial migration in situ can now be quantified with important implications for better understanding of the immune response and cancer metastasis. As a new and emerging topic, the increased deformability of cancer cells seems not only relevant for migration but also as an immune escape mechanism. In this context, Brillouin microscopy will be an important tool to study the mechanical interaction between cancer and immune cells at the immune synapse.
Brillouin microscopy for mechanosensing
In the wider context of biological application, the use of Brillouin microscopy will increasingly move away from demonstrating its general applicability for all kinds of cells, tissues and organisms and towards answering biologically relevant questions in morphogenesis, development and pathology. An exciting example could be investigating the role of mechanosensing during spinal cord repair. The use of zebrafish larvae is beneficial in this context because of their high optical transparency, which is particularly important for Brillouin microscopy, and because they can actually repair an injury to their spinal cord within days52. Insight into these mechanisms could help cure spinal cord injuries also in mammals, which is currently not possible. Of course, the likely mode of cells mechanically testing their environment is a shear deformation on timescales of milliseconds to seconds, which renders Brillouin microscopy seemingly less ideal for studies into mechanosensing. Also, the correlation frequently observed between the longitudinal modulus measured with Brillouin microscopy and the Young’s modulus obtained with AFM nanoindentation probably holds only in the same type of material. What if the structural basis for the mechanical properties changes within tissues, for example from cell-body dominated to extracellular matrix dominated? Even though not yet demonstrated, it might be possible to address these challenges by using machine learning together with employing other measurement modalities simultaneously, such as fluorescence or Raman imaging in order to obtain sufficient additional information about the tissues. Transferring this learning, it might be possible to reliably infer AFM-type shear modulus information at timescales of seconds from Brillouin frequency maps obtained in areas not accessible to AFM.
Quantifying emergent biophysical phenomena
Finally, it might also be time to emancipate the field of Brillouin microscopy from the attempt to provide the kind of information conventionally obtained by other contact-based methods (such as AFM nanoindentation). After all, there should be significance in the fact that tissues are naturally excited at gigahertz frequencies by thermal excitation. It seems unlikely that this aspect had been excluded from evolutionary optimization and is not somehow utilized by cells. Looking at the amplitudes of the density fluctuations involved, which are on the scale of nanometres, it is only natural to suggest that Brillouin microscopy might be sensitive to intermolecular interactions. How strongly molecules interact with each other is a very relevant question across all of molecular biology, particularly biomolecular condensates167. Certain proteins condense to form liquid droplets inside cells. This is a biologically regulated process, likely to sequester these proteins away from other reactions going on in the cell, or to locally increase their concentration beyond a critical point to enable certain reactions. The interaction of these molecules must be so specific that only certain proteins form condensates, but without proceeding to crystallization. The molecular grammar involved in these condensation phenomena is currently being deciphered. This is an area where Brillouin microscopy, with its sensitivity for intermolecular interactions, could be the label-free, quantitative and direct measurement technique needed to take this field forward. Brillouin microscopy has already demonstrated that it can detect and analyse the properties of protein condensates inside living cells44. Interestingly, mutations in some of the proteins known to form biomolecular condensates, such as fused in sarcoma proteins, are linked to neurodegenerative disorders such as amyotrophic lateral sclerosis210. Mutations also cause an irreversible phase transition of the condensates from a liquid to a solid state. At present, it is not clear whether, and if so how, this change in material properties of the condensates is causally linked to disease pathogenesis, but considering the plaques found in the end states of the disease, it seems sensible to look for such a connection. Brillouin microscopy will have a role to play in this quest. More generally, with the molecular level of biology starting to be very well investigated, there is the question of how the genomic, proteomic, transcriptomic molecular information is used and how this level of information can be conceptionally transferred to larger scales. Some aspect of the solution will involve the self-organization of proteins and so on at supramolecular, subcellular scales. Again, Brillouin microscopy seems ideally suited to contribute to the quantitative exploration of these emergent properties and to drive the paradigm shift in biology from the biochemistry of molecules to the next level. In summary, the future of Brillouin microscopy looks bright.
Supplementary Material
Acknowledgements
The authors thank R. Schlüßler, S. Möllmert, K. Kim, T. Beck, J. Czarske, K. Elsayad, R. Prevedel, G. Antonacci, P. Török, F. Palombo, M. Mattarelli and T. Dehoux for many stimulating discussions. I.K. is supported by the Australian Research Council Centre of Excellence in Optical Microcombs for Breakthrough Science (CE230100006) and the Australian Research Council Centre of Excellence in Quantum Biotechnology (CE230100021). J.Z. is supported by the Eunice Kennedy Shriver National Institute of Child Health and Human Development, National Institutes of Health (NIH) (K25HD097288, R21HD112663). G.S is supported by the National Science Foundation (DBI-1942003) and NIH (R21CA258008, R01EY028666, R01EY030063). A.B. acknowledges the support of the Israel Science Foundation (grant no. 2576/21). S.C. is supported by European Union — NextGenerationEU under the Italian Ministry of University and Research (MUR) National Innovation Ecosystem grant ECS00000041 — VITALITY — CUP B43C22000470005 and PRIN 2022 PNRR P2022RH4HH.
Glossary
- Abbe diffraction limit
This determines the size of the spot to which the light can be focused based on the law of diffraction. It is approximately proportional to half of the light wavelength, setting the lower limit to the resolution of a classical optical system
- Beat signal
The wave pattern formed via interference of two waves with slightly different frequencies
- Bragg angle
The angle between the incident light beam and the diffracting grating plane
- Brillouin frequency shift
(BFS). The frequency shift that a light wave experiences undergoing Brillouin scattering process
- Elasticity tensor
A fourth-rank tensor describing the stress–strain relation in a linear elastic material. Most materials can be described as linear elastic under small deformation approximation
- Etalon
A device consisting of two reflecting glass plates, employed for measuring the light spectrum
- Longitudinal modulus
The ratio of axial stress to axial strain under confined uniaxial deformation (the object allowed to deform along one axis only, with possible changes in volume)
- Matrix pencil method
A computational technique based on diagonal matrices, particularly useful to estimate parameters of complex exponential signals buried in noise
- Mechanotransduction
The conversion of an external mechanical signal into intracellular biochemical signals
- Micromechanical properties
The elastic and viscous properties of materials at the microscale
- Rayleigh band
Refers to spectral signals at and near the frequency of the laser, including Rayleigh light scattering, quasi-elastic scattering and stray laser light
- Shear modulus
The ratio of shear stress to shear strain, serving as the measure of shear stiffness
- Voxel
A measurement of volume in a structure to be imaged
- Young’s modulus
A measure of the ratio of applied stress to axial strain under unconfined deformation (the object is free to deform in all directions while its volume is conserved)
Footnotes
Competing interests
The authors declare no competing interests.
Peer review information Nature Reviews Methods Primers thanks Giuseppe Antonacci, Robert Prevedel, Seok Hyun Yun and Vladislav V. Yakovlev for their contribution to the peer review of this work.
Additional information
Supplementary information The online version contains supplementary material available at https://doi.org/10.1038/s43586-023-00286-z.
Related links
Bio-Brillouin Society: https://www.biobrillouin.org/
References
- 1.Wolff J, Das Gesetz der Transformation der Knochen [German] (Hirschwald, 1892).
- 2.Ingber D Mechanobiology and diseases of mechanotransduction. Ann. Med. 35, 564–577 (2003). [DOI] [PubMed] [Google Scholar]
- 3.Keller R Physical biology returns to morphogenesis. Science 338, 201–203 (2012). [DOI] [PubMed] [Google Scholar]
- 4.Baratchi S et al. Molecular sensors of blood flow in endothelial cells. Trends Mol. Med. 23, 850–868 (2017). [DOI] [PubMed] [Google Scholar]
- 5.Kalli M & Stylianopoulos T Defining the role of solid stress and matrix stiffness in cancer cell proliferation and metastasis. Front. Oncol. 8, 55 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Urbanska M et al. A comparison of microfluidic methods for high-throughput cell deformability measurements. Nat. Methods 17, 587–593 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Arbore C et al. Probing force in living cells with optical tweezers: from single-molecule mechanics to cell mechanotransduction. Biophys. Rev. 11, 765–782 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Wang H et al. Recent advances of optical tweezers-based dynamic force spectroscopy and mechanical measurement assays for live-cell mechanobiology. Front. Phys. 10, 1–11 (2022). [Google Scholar]
- 9.Staunton JR et al. Correlating confocal microscopy and atomic force indentation reveals metastatic cancer cells stiffen during invasion into collagen I matrices. Sci. Rep. 6, 19686 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Benitez R & Toca-herrera JL Looking at cell mechanics with atomic force microscopy: experiment and theory. Microsc. Res. Tech. 77, 947–958 (2014). [DOI] [PubMed] [Google Scholar]
- 11.Li Y et al. Ultrahigh-sensitive optical coherence elastography. Light Sci. Appl. 9, 58 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Scarcelli G & Yun SH Confocal Brillouin microscopy for three-dimensional mechanical imaging. Nat. Photonics 2, 39–43 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Poon C et al. Brillouin imaging for studies of micromechanics in biology and biomedicine: from current state-of-the-art to future clinical translation. J. Phys. Photonics 3, 012002 (2021). [Google Scholar]
- 14.Prevedel R et al. Brillouin microscopy: an emerging tool for mechanobiology. Nat. Methods 16, 969–977 (2019). [DOI] [PubMed] [Google Scholar]
- 15.Antonacci G et al. Recent progress and current opinions in Brillouin microscopy for life science applications. Biophys. Rev. 12, 615–624 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Matsuda O et al. Fundamentals of picosecond laser ultrasonics. Ultrasonics 56, 3–20 (2015). [DOI] [PubMed] [Google Scholar]
- 17.Tomoda M et al. Time-domain Brillouin imaging of sound velocity and refractive index using automated angle scanning. Photoacoustics 31, 100486 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Gusev V & Ruello P Advances in applications of time-domain Brillouin scattering for nanoscale imaging. Appl. Phys. Rev. 5, 031101 (2018). [Google Scholar]
- 19.Conrad C et al. Mechanical characterization of 3D ovarian cancer nodules using Brillouin confocal microscopy. Cell. Mol. Bioeng. 12, 215–226 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Bevilacqua C et al. Imaging mechanical properties of sub-micron ECM in live zebrafish using Brillouin microscopy. Biomed. Opt. Express 10, 1420–1431 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Scarcelli G et al. In vivo biomechanical mapping of normal and keratoconus corneas. JAMA Ophthalmol. 133, 480–482 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Scarcelli G et al. Noncontact three-dimensional mapping of intracellular hydromechanical properties by Brillouin microscopy. Nat. Methods 12, 1132–1134 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Fiore A, Bevilacqua C & Scarcelli G Direct three-dimensional measurement of refractive index via dual photon-phonon scattering. Phys. Rev. Lett. 122, 103901 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Chan CJ, Bevilacqua C & Prevedel R Mechanical mapping of mammalian follicle development using Brillouin microscopy. Commun. Biol. 4, 1133 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Palombo F et al. Biomechanics of fibrous proteins of the extracellular matrix studied by Brillouin scattering. J. Royal Soc. Interface 11, 20140739 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Bottani CE & Fioretto D Brillouin scattering of phonons in complex materials. Adv. Phys.: X 3, 1467281 (2018). [Google Scholar]
- 27.Merklein M et al. 100 years of Brillouin scattering: historical and future perspectives. Appl. Phys. Rev. 9, 041306 (2022). [Google Scholar]
- 28.Kim M et al. Shear Brillouin light scattering microscope. Opt. Express 24, 319–328 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Fioretto D et al. Cauchy relation in relaxing liquids. J. Chem. Phys. 128, 214502 (2008). [DOI] [PubMed] [Google Scholar]
- 30.Cusack S & Miller A Determination of the elastic constants of collagen by Brillouin light scattering. J. Mol. Biol. 135, 39–51 (1979). [DOI] [PubMed] [Google Scholar]
- 31.Koski KJ et al. Non-invasive determination of the complete elastic moduli of spider silks. Nat. Mater. 12, 262–267 (2013). [DOI] [PubMed] [Google Scholar]
- 32.Remer I et al. High-sensitivity and high-specificity biomechanical imaging by stimulated Brillouin scattering microscopy. Nat. Methods 17, 913–916 (2020). [DOI] [PubMed] [Google Scholar]
- 33.Ballmann CW et al. Stimulated Brillouin scattering microscopic imaging. Sci. Rep. 5, 18139 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Mock R, Hillebrands B & Sandercock R Construction and performance of a Brillouin scattering set-up using a triple-pass tandem Fabry–Perot interferometer. J. Phys. E Sci. Instruments 20, 656 (1987). [Google Scholar]
- 35.Shirasaki M Large angular dispersion by a virtually imaged phased array and its application to a wavelength demultiplexer. Opt. Lett. 21, 366–368 (1996). [DOI] [PubMed] [Google Scholar]
- 36.Alunni Cardinali M et al. Brillouin–Raman microspectroscopy for the morpho-mechanical imaging of human lamellar bone. J. R. Soc. Interface 19, 20210642 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Bevilacqua C et al. High-resolution line-scan Brillouin microscopy for live imaging of mechanical properties during embryo development. Nat. Methods 20, 755–760 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Margueritat J et al. High-frequency mechanical properties of tumors measured by Brillouin light scattering. Phys. Rev. Lett. 122, 018101 (2019). [DOI] [PubMed] [Google Scholar]
- 39.Zhang J et al. Rapid biomechanical imaging at low irradiation level via dual line-scanning Brillouin microscopy. Nat. Methods 20, 677–681 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Mahajan V et al. Mapping tumor spheroid mechanics in dependence of 3D microenvironment stiffness and degradability by Brillouin microscopy. Cancers 13, 5549 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Karampatzakis A et al. Probing the internal micromechanical properties of Pseudomonas aeruginosa biofilms by Brillouin imaging. NPJ Biofilms Microbiomes 3, 20 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Zhang J et al. Nuclear mechanics within intact cells is regulated by cytoskeletal network and internal nanostructures. Small 16, 1907688 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Bacete L et al. THESEUS1 modulates cell wall stiffness and abscisic acid production in Arabidopsis thaliana. Proc. Natl Acad. Sci. USA 119, e2119258119 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Antonacci G et al. Background-deflection Brillouin microscopy reveals altered biomechanics of intracellular stress granules by ALS protein FUS. Commun. Biol. 1, 139 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Fasciani A et al. MLL4-associated condensates counterbalance Polycomb-mediated nuclear mechanical stress in Kabuki syndrome. Nat. Genet. 52, 1397–1411 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Schlüßler R et al. Correlative all-optical quantification of mass density and mechanics of subcellular compartments with fluorescence specificity. eLife 11, e68490 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Shao P et al. Spatially-resolved Brillouin spectroscopy reveals biomechanical abnormalities in mild to advanced keratoconus in vivo. Sci. Rep. 9, 7467 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Kennedy BF, Wijesinghe P & Sampson DD The emergence of optical elastography in biomedicine. Nat. Photonics 11, 215–221 (2017). [Google Scholar]
- 49.Olie CS et al. Cytoskeletal disorganization underlies PABPN1-mediated myogenic disability. Sci. Rep. 10, 17621 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Frittoli E et al. Tissue fluidification promotes a cGAS–STING cytosolic DNA response in invasive breast cancer. Nat. Mater. 22, 644–655 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Pukhlyakova E et al. β-Catenin-dependent mechanotransduction dates back to the common ancestor of Cnidaria and Bilateria. Proc. Natl Acad. Sci. USA 115, 6231–6236 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Schlüßler R et al. Mechanical mapping of spinal cord growth and repair in living zebrafish larvae by Brillouin imaging. Biophys. J. 115, 911–923 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Bailey M et al. Viscoelastic properties of biopolymer hydrogels determined by Brillouin spectroscopy: a probe of tissue micromechanics. Sci. Adv. 6, eabc1937 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Polonchuk L et al. Towards engineering heart tissues from bioprinted cardiac spheroids. Biofabrication 13, 045009 (2021). [DOI] [PubMed] [Google Scholar]
- 55.Pagano S et al. Bio-mechanical characterization of a CAD/CAM PMMA resin for digital removable prostheses. Dental Mater. 37, e118–e130 (2021). [DOI] [PubMed] [Google Scholar]
- 56.Lainović T et al. Micromechanical imaging of dentin with Brillouin microscopy. Acta Biomaterialia 105, 214–222 (2020). [DOI] [PubMed] [Google Scholar]
- 57.Mahmodi H et al. Mechanical mapping of bioprinted hydrogel models by Brillouin microscopy. Bioprinting 23, e00151 (2021). [Google Scholar]
- 58.Mattarelli M, Vassalli M & Caponi S Relevant length scales in Brillouin imaging of biomaterials: the interplay between phonons propagation and light focalization. ACS Photonics 7, 2319–2328 (2020). [Google Scholar]
- 59.Passeri AA et al. Size and environment: the effect of phonon localization on micro-Brillouin imaging. Biomater. Adv. 147, 213341 (2023). [DOI] [PubMed] [Google Scholar]
- 60.Caponi S, Fioretto D & Mattarelli M On the actual spatial resolution of Brillouin imaging. Opt. Lett. 45, 1063–1066 (2020). [DOI] [PubMed] [Google Scholar]
- 61.Mattarelli M et al. Vibration spectroscopy of weakly interacting mesoscopic colloids. Soft Matter 8, 4235–4243 (2012). [Google Scholar]
- 62.Still T et al. Eigenvibrations of submicrometer colloidal spheres. J. Phys. Chem. Lett. 1, 2440–2444 (2010). [Google Scholar]
- 63.Gomopoulos N et al. Out-of-plane longitudinal elastic modulus of supported polymer thin films. Macromolecules 42, 7164–7167 (2009). [Google Scholar]
- 64.Kabakova I, Scarcelli G & Yun S-H in Semiconductors and Semimetals Ch. 15 (eds Eggleton BJ, Steel MJ & Poulton CG) 313–348 (Elsevier, 2022). [Google Scholar]
- 65.Remer I & Bilenca A High-speed stimulated Brillouin scattering spectroscopy at 780 nm. APL Photonics 1, 061301 (2016). [Google Scholar]
- 66.Shaashoua R et al. Brillouin gain microscopy. Preprint at 10.48550/arXiv.2306.00889 (2023). [DOI]
- 67.Chow DM & Yun S-H Pulsed stimulated Brillouin microscopy. Opt. Express 31, 19818–19827 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Lindsay S, Anderson M & Sandercock J Construction and alignment of a high performance multipass vernier tandem Fabry–Perot interferometer. Rev. Sci. Instrum. 52, 1478–1486 (1981). [Google Scholar]
- 69.Dil JG Brillouin scattering in condensed matter. Rep. Prog. Phys. 45, 285–334 (1982). [Google Scholar]
- 70.Hickman GD et al. Aircraft laser sensing of sound velocity in water: Brillouin scattering. Remote sensing Environ. 36, 165–178 (1991). [Google Scholar]
- 71.Harley R et al. Phonons and the elastic moduli of collagen and muscle. Nature 267, 285–287 (1977). [DOI] [PubMed] [Google Scholar]
- 72.Randall JT & Vaughan JM Brillouin scattering in systems of biological significance. Philos. Trans. Royal Soc. Lond. Series A, Math. Phys. Sci. 293, 341–348 (1979). [Google Scholar]
- 73.Rad MA et al. Micromechanical characterisation of 3D bioprinted neural cell models using Brillouin microspectroscopy. Bioprinting 25, e00179 (2022). [Google Scholar]
- 74.Vaughan J & Randall J Brillouin scattering, density and elastic properties of the lens and cornea of the eye. Nature 284, 489–491 (1980). [DOI] [PubMed] [Google Scholar]
- 75.Randall J & Vaughan J The measurement and interpretation of Brillouin scattering in the lens of the eye. Proc. Royal Soc. Lond. B: Biol. Sci. 214, 449–470 (1982). [DOI] [PubMed] [Google Scholar]
- 76.Mercatelli R et al. Morpho-mechanics of human collagen superstructures revealed by all-optical correlative micro-spectroscopies. Commun. Biol. 2, 117 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Lee S et al. A Brillouin scattering study of the hydration of Li- and Na-DNA films. Biopolymers: Original Res. Biomolecules 26, 1637–1665 (1987). [DOI] [PubMed] [Google Scholar]
- 78.Scarponi F et al. High-performance versatile setup for simultaneous Brillouin–Raman microspectroscopy. Phys. Rev. X 7, 031015 (2017). [Google Scholar]
- 79.Itoh S-I, Yamana T & Kojima S Quick measurement of Brillouin spectra of glass-forming material trimethylene glycol by angular dispersion-type Fabry–Perot interferometer system. Japanese J.Appl. Phys. 35, 2879 (1996). [Google Scholar]
- 80.Cardinali MA et al. Mechano-chemistry of human femoral diaphysis revealed by correlative Brillouin–Raman microspectroscopy. Sci. Rep. 10, 17341 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Cardinali MA et al. Brillouin micro-spectroscopy of subchondral, trabecular bone and articular cartilage of the human femoral head. Biomed. Opt. Express 10, 2606–2611 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Koski KJ & Yarger JL Brillouin imaging. Appl. Phys. Lett. 87, 061903 (2005). [Google Scholar]
- 83.Scarcelli G & Yun SH Multistage VIPA etalons for high-extinction parallel Brillouin spectroscopy. Opt. Express 19, 10913–10922 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Berghaus KV, Yun SH & Scarcelli G High speed sub-GHz spectrometer for Brillouin scattering analysis. J. Vis. Exp. 106, e53468 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Antonacci G et al. Elastic suppression in Brillouin imaging by destructive interference. Appl. Phys. Lett. 107, 061102 (2015). [Google Scholar]
- 86.Fiore A et al. High-extinction virtually imaged phased array-based Brillouin spectroscopy of turbid biological media. Appl. Phys. Lett. 108, 203701 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Edrei E, Gather MC & Scarcelli G Integration of spectral coronagraphy within VIPA-based spectrometers for high extinction Brillouin imaging. Opt. Express 25, 6895–6903 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Nikolić M & Scarcelli G Long-term Brillouin imaging of live cells with reduced absorption-mediated damage at 660 nm wavelength. Biomed. Optics Exp. 10, 1567–1580 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Raghunathan R et al. Evaluating biomechanical properties of murine embryos using Brillouin microscopy and optical coherence tomography. J. Biomed. Opt. 22, 086013 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Zhang J et al. Tissue biomechanics during cranial neural tube closure measured by Brillouin microscopy and optical coherence tomography. Birth Defects Res. 111, 991–998 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Huisken J et al. Optical sectioning deep inside live embryos by selective plane illumination microscopy. Science 305, 1007–1009 (2004). [DOI] [PubMed] [Google Scholar]
- 92.Stelzer EH et al. Light sheet fluorescence microscopy. Nat. Rev. Methods Primers 1, 73 (2021). [Google Scholar]
- 93.Zhang J et al. Line-scanning Brillouin microscopy for rapid non-invasive mechanical imaging. Sci. Rep. 6, 35398 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Remer I, Cohen L & Bilenca A High-speed continuous-wave stimulated Brillouin scattering spectrometer for material analysis. J. Vis. Exp. 127, e55527 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Faris GW, Jusinski LE & Hickman AP High-resolution stimulated Brillouin gain spectroscopy in glasses and crystals. J. Opt. Soc. Am. B 10, 587–599 (1993). [Google Scholar]
- 96.Ratanaphruks K, Tandy Grubbs W & MacPhail RA cw stimulated Brillouin gain spectroscopy of liquids. Chem. Phys. Lett. 182, 371–378 (1991). [Google Scholar]
- 97.Taylor MA et al. Heterodyne Brillouin microscopy for biomechanical imaging. Biomed. Opt. Exp. 12, 6259–6268 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Shaashoua R et al. Enhancing biomechanical stimulated Brillouin scattering imaging with physics-driven model selection. Preprint at 10.48550/arXiv.2306.00910 (2023). [DOI]
- 99.Shaashoua R & Bilenca A Aperture-induced spectral effects in stimulated Brillouin scattering microscopy. Appl. Phys. Lett. 122, 143702 (2023). [Google Scholar]
- 100.Li T et al. Quantum-enhanced stimulated Brillouin scattering spectroscopy and imaging. Optica 9, 959–964 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Yang F et al. Pulsed stimulated Brillouin microscopy enables high-sensitivity mechanical imaging of live and fragile biological specimens. Nat. Methods 20, 1971–1979 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Ballmann CW et al. Impulsive Brillouin microscopy. Optica 4, 124–128 (2017). [Google Scholar]
- 103.Li J et al. High-speed non-contact measurement of elasto-optic coefficient via laser-induced phonons. Appl. Phys. Lett. 121, 251102 (2022). [Google Scholar]
- 104.Krug B et al. Nonlinear microscopy using impulsive stimulated Brillouin scattering for high-speed elastography. Opt. Express 30, 4748–4758 (2022). [DOI] [PubMed] [Google Scholar]
- 105.Krug B, Koukourakis N & Czarske JW Impulsive stimulated Brillouin microscopy for non-contact, fast mechanical investigations of hydrogels. Opt. Express 27, 26910–26923 (2019). [DOI] [PubMed] [Google Scholar]
- 106.Maznev AA, Nelson KA & Rogers JA Optical heterodyne detection of laser-induced gratings. Opt. Lett. 23, 1319–1321 (1998). [DOI] [PubMed] [Google Scholar]
- 107.Kinoshita S et al. New high-resolution phonon spectroscopy using impulsive stimulated Brillouin scattering. Rev. Sci. Instrum. 64, 3384–3393 (1993). [Google Scholar]
- 108.Nelson KA & Fayer MD Laser induced phonons: a probe of intermolecular interactions in molecular solids. J. Chem. Phys. 72, 5202–5218 (2008). [Google Scholar]
- 109.Erdogan T & Mizrahi V Thin-film filters come of age. Photonics Spectra 37, 94–100 (2003). [Google Scholar]
- 110.Meng Z, Traverso AJ & Yakovlev VV Background clean-up in Brillouin microspectroscopy of scattering medium. Opt. Express 22, 5410–5415 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Lepert G et al. Assessing corneal biomechanics with Brillouin spectro-microscopy. Faraday Discuss 187, 415–428 (2016). [DOI] [PubMed] [Google Scholar]
- 112.Berghaus K et al. High-finesse sub-GHz-resolution spectrometer employing VIPA etalons of different dispersion. Opt. Lett. 40, 4436–4439 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113.Correa N et al. Image analysis applied to Brillouin images of tissue-mimicking collagen gelatins. Biomed. Opt. Express 10, 1329–1338 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Hauck N et al. PNIPAAm microgels with defined network architecture as temperature sensors in optical stretchers. Mater. Adv. 3, 6179–6190 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Girardo S et al. Standardized microgel beads as elastic cell mechanical probes. J. Mater. Chem. B 6, 6245–6261 (2018). [DOI] [PubMed] [Google Scholar]
- 116.Hauck N et al. Droplet-assisted microfluidic fabrication and characterization of multifunctional polysaccharide microgels formed by multicomponent reactions. Polymers 10, 1055 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Wu P-J et al. Water content, not stiffness, dominates Brillouin spectroscopy measurements in hydrated materials. Nat. Methods 15, 561–562 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Zhao & Vanderwal JJ Brillouin scattering study of gelatin gel. Polym. Gels Netw. 5, 23–36 (1997). [Google Scholar]
- 119.Johnson DL Elastodynamics of gels. J. Chem. Phys. 77, 1531–1539 (1982). [Google Scholar]
- 120.Scarcelli G & Yun SH Reply to ‘Water content, not stiffness, dominates Brillouin spectroscopy measurements in hydrated materials’. Nat. Methods 15, 562–563 (2018). [DOI] [PubMed] [Google Scholar]
- 121.Shao P et al. Effects of corneal hydration on Brillouin microscopy in vivo. Invest. Ophthalmol. Vis. Sci. 59, 3020–3027 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Scarcelli G et al. Brillouin microscopy of collagen crosslinking: noncontact depth-dependent analysis of corneal elastic modulus. Invest. Ophthalmol. Vis. Sci. 54, 1418–1425 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Scarcelli G, Kim & Yun, S. H. In vivo measurement of age-related stiffening in the crystalline lens by Brillouin optical microscopy. Biophys. J. 101, 1539–1545 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Scarcelli G, Pineda R & Yun SH Brillouin optical microscopy for corneal biomechanics. Invest. Ophthalmol. Vis. Sci. 53, 185–190 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 125.Gouveia RM et al. Assessment of corneal substrate biomechanics and its effect on epithelial stem cell maintenance and differentiation. Nat. Commun. 10, 1496 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Weber IP et al. The role of cell body density in ruminant retina mechanics assessed by atomic force and Brillouin microscopy. Phys. Biol. 14, 065006 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 127.Amini R et al. Amoeboid-like migration ensures correct horizontal cell layer formation in the developing vertebrate retina. eLife 11, e76408 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Soans KG et al. Collective cell migration during optic cup formation features changing cell–matrix interactions linked to matrix topology. Curr. Biol. 32, 4817–4831.e9 (2022). [DOI] [PubMed] [Google Scholar]
- 129.Sánchez-Iranzo H et al. A 3D Brillouin microscopy dataset of the in-vivo zebrafish eye. Data Brief 30, 105427 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Ryu S et al. Label-free histological imaging of tissues using Brillouin light scattering contrast. Biomed. Opt. Express 12, 1437–1448 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Riquelme-Guzmán C et al. In vivo assessment of mechanical properties during axolotl development and regeneration using confocal Brillouin microscopy. Open Biol. 12, 220078 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132.Zhang H et al. Motion tracking Brillouin microscopy evaluation of normal, keratoconic, and post-laser vision correction corneas. Am. J. Ophthalmol. 254, 128–140 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133.Zhang J & Scarcelli G Mapping mechanical properties of biological materials via an add-on Brillouin module to confocal microscopes. Nat. Protocols 16, 1251–1275 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 134.Webb JN et al. Detecting mechanical anisotropy of the cornea using Brillouin microscopy. Transl. Vis. Science Technol. 9, 26–26 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 135.Eltony AM, Shao P & Yun SH Measuring mechanical anisotropy of the cornea with Brillouin microscopy. Nat. Commun 13, 1354 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136.Griffith LG & Swartz MA Capturing complex 3D tissue physiology in vitro. Nat. Rev. Mol. Cell Biol. 7, 211–224 (2006). [DOI] [PubMed] [Google Scholar]
- 137.Nikolić M, Scarcelli G & Tanner K Multimodal microscale mechanical mapping of cancer cells in complex microenvironments. Biophys. J. 121, 3586–3599 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Bakhshandeh S et al. Optical quantification of intracellular mass density and cell mechanics in 3D mechanical confinement. Soft Matter 17, 853–862 (2021). [DOI] [PubMed] [Google Scholar]
- 139.Mattana S et al. Non-contact mechanical and chemical analysis of single living cells by microspectroscopic techniques. Light Sci. Appl. 7, 17139–17139 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Barer R Interference microscopy and mass determination. Nature 169, 366–367 (1952). [DOI] [PubMed] [Google Scholar]
- 141.Kim K et al. Diffraction optical tomography using a quantitative phase imaging unit. Opt. Lett. 39, 6935–6938 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142.Montrose CJ, Solovyev VA & Litovitz TA Brillouin scattering and relaxation in liquids. J Acoust. Soc. Am. 43, 117–130 (2005). [Google Scholar]
- 143.Antonacci G et al. Spectral broadening in Brillouin imaging. Appl. Phys. Lett. 103, 221105 (2013). [Google Scholar]
- 144.Mattarelli M et al. Disentanglement of multiple scattering contribution in Brillouin microscopy. ACS Photonics 9, 2087–2091 (2022). [Google Scholar]
- 145.Lasch P Spectral pre-processing for biomedical vibrational spectroscopy and microspectroscopic imaging. Chemom. Intell. Lab. Syst. 117, 100–114 (2012). [Google Scholar]
- 146.Xiang Y et al. Multivariate analysis of Brillouin imaging data by supervised and unsupervised learning. J. Biophotonics 14, e202000508 (2021). [DOI] [PubMed] [Google Scholar]
- 147.Xiang Y, Foreman MR & Torok P SNR enhancement in Brillouin microspectroscopy using spectrum reconstruction. Biomed. Opt. Express 11, 1020–1031 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.Li J et al. Sensitive impulsive stimulated Brillouin spectroscopy by an adaptive noise-suppression matrix pencil. Opt. Express 30, 29598–29610 (2022). [DOI] [PubMed] [Google Scholar]
- 149.Torok P & Foreman MR Precision and informational limits in inelastic optical spectroscopy. Sci. Rep. 9, 6140 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 150.Antonacci G et al. Quantification of plaque stiffness by Brillouin microscopy in experimental thin cap fibroatheroma. J. R. Soc. Interface 12, 20150843 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Elsayad K Spectral phasor analysis for Brillouin microspectroscopy. Front. Phys. 7, 62 (2019). [Google Scholar]
- 152.Fioretto D, Caponi S & Palombo F Brillouin–Raman mapping of natural fibers with spectral moment analysis. Biomed. Opt. Express 10, 1469–1474 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Jun L et al. Minimum volume simplex analysis: a fast algorithm for linear hyperspectral unmixing. IEEE Trans. Geosci. Remote Sensing 53, 5067–5082 (2015). [Google Scholar]
- 154.Palombo F et al. Hyperspectral analysis applied to micro-Brillouin maps of amyloid-β plaques in Alzheimer’s disease brains. Analyst 143, 6095–6102 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 155.Bradbury P et al. Timothy grass pollen induces spatial reorganisation of F-actin and loss of junctional integrity in respiratory cells. Inflammation 45, 1209–1223 (2022). [DOI] [PubMed] [Google Scholar]
- 156.Krafft C et al. Disease recognition by infrared and Raman spectroscopy. J. Biophotonics 2, 13–28 (2009). [DOI] [PubMed] [Google Scholar]
- 157.Bao G & Suresh S Cell and molecular mechanics of biological materials. Nat. Mater. 2, 715–725 (2003). [DOI] [PubMed] [Google Scholar]
- 158.Discher DE, Janmey P & Wang YL Tissue cells feel and respond to the stiffness of their substrate. Science 310, 1139–1143 (2005). [DOI] [PubMed] [Google Scholar]
- 159.Janmey PA & McCulloch CA Cell mechanics: integrating cell responses to mechanical stimuli. Annu. Rev. Biomed. Eng. 9, 1–34 (2007). [DOI] [PubMed] [Google Scholar]
- 160.Bershadsky A, Kozlov M & Geiger B Adhesion-mediated mechanosensitivity: a time to experiment, and a time to theorize. Curr. Opin. Cell Biol. 18, 472–481 (2006). [DOI] [PubMed] [Google Scholar]
- 161.Hu SH et al. Intracellular stress tomography reveals stress focusing and structural anisotropy in cytoskeleton of living cells. Am. J. Physiol. Cell Physiol. 285, C1082–C1090 (2003). [DOI] [PubMed] [Google Scholar]
- 162.Ingber DE Cellular mechanotransduction: putting all the pieces together again. FASEB J. 20, 811–827 (2006). [DOI] [PubMed] [Google Scholar]
- 163.Antonacci G & Braakman S Biomechanics of subcellular structures by non-invasive Brillouin microscopy. Sci. Rep. 6, 37217 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 164.Coppola S et al. Quantifying cellular forces and biomechanical properties by correlative micropillar traction force and Brillouin microscopy. Biomed. Opt. Express 10, 2202–2212 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 165.Elsayad K et al. Mapping the subcellular mechanical properties of live cells in tissues with fluorescence emission–Brillouin imaging. Sci. Signal. 9, rs5 (2016). [DOI] [PubMed] [Google Scholar]
- 166.Zhang J et al. Brillouin flow cytometry for label-free mechanical phenotyping of the nucleus. Lab Chip 17, 663–670 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Abuhattum S et al. Adipose cells and tissues soften with lipid accumulation while in diabetes adipose tissue stiffens. Sci. Rep. 12, 10325 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 168.Krachmer JH, Feder RS & Belin MW Keratoconus and related noninflammatory corneal thinning disorders. Surv. Ophthalmol. 28, 293–322 (1984). [DOI] [PubMed] [Google Scholar]
- 169.Jun AS et al. Subnormal cytokine profile in the tear fluid of keratoconus patients. PLoS ONE 6, e16437 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 170.Rabinowitz Y Ectasia after laser in situ keratomileusis. Curr. Opin. Ophthalmol. 17, 421–427 (2006). [DOI] [PubMed] [Google Scholar]
- 171.Binder PS et al. Keratoconus and corneal ectasia after LASIK. J. Refract. Surg. 21, 749–752 (2005). [DOI] [PubMed] [Google Scholar]
- 172.Roy AS & Dupps WJ Jr Patient-specific computational modeling of keratoconus progression and differential responses to collagen cross-linking. Invest. Ophthalmol. Vis. Sci. 52, 9174–9187 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 173.Meek K et al. Changes in collagen orientation and distribution in keratoconus corneas. Invest. ophthalmol. Vis. Sci. 46, 1948–2004 (2005). [DOI] [PubMed] [Google Scholar]
- 174.Morishige N et al. Second-harmonic imaging microscopy of normal human and keratoconus cornea. Invest. Ophthalmol. Vis. Sci. 48, 1087–1181 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 175.Zimmermann DR et al. Comparative studies of collagens in normal and keratoconus corneas. Exp. Eye Res. 46, 431–442 (1988). [DOI] [PubMed] [Google Scholar]
- 176.Petsche SJ & Pinsky PM The role of 3-D collagen organization in stromal elasticity: a model based on X-ray diffraction data and second harmonic-generated images. Biomech. Model. Mechanobiol. 12, 1101–1113 (2014). [DOI] [PubMed] [Google Scholar]
- 177.Zhang H et al. Motion-tracking Brillouin microscopy for in-vivo corneal biomechanics mapping. Biomed. Opt. Express 13, 6196–6210 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 178.Ambekar Y et al. Multimodal quantitative optical elastography of the crystalline lens with optical coherence elastography and Brillouin microscopy. Biomed. Opt. Express 11, 2041–2051 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 179.Besner S et al. In vivo Brillouin analysis of the aging crystalline lens. Invest. Ophthalmol. Vis. Sci. 57, 5093–5100 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Liu W & Wang Z Current understanding of the biomechanics of ventricular tissues in heart failure. Bioengineering 7, 2 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 181.Villalba-Orero M et al. Assessment of myocardial viscoelasticity with Brillouin spectroscopy in myocardial infarction and aortic stenosis models. Sci. Rep. 11, 21369 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Libby P Mechanisms of acute coronary syndromes. N. Engl. J. Med. 369, 883–884 (2013). [DOI] [PubMed] [Google Scholar]
- 183.VIRMANI R et al. Pathology of the thin-cap fibroatheroma. J. Interv. Cardiol. 16, 267–272 (2003). [DOI] [PubMed] [Google Scholar]
- 184.Mathieu V et al. Micro-Brillouin scattering measurements in mature and newly formed bone tissue surrounding an implant. J. Biomech. Eng. 133, 021006 (2011). [DOI] [PubMed] [Google Scholar]
- 185.Matsukawa M et al. Application of a micro-Brillouin scattering technique to characterize bone in the GHz range. Ultrasonics 54, 1155–1161 (2014). [DOI] [PubMed] [Google Scholar]
- 186.Akilbekova D et al. Brillouin spectroscopy and radiography for assessment of viscoelastic and regenerative properties of mammalian bones. J. Biomed. Opt. 23, 097004 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 187.Wu P-J et al. Detection of proteoglycan loss from articular cartilage using Brillouin microscopy, with applications to osteoarthritis. Biomed. Opt. Express 10, 2457–2466 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 188.Guck J et al. Optical deformability as an inherent cell marker for testing malignant transformation and metastatic competence. Biophys. J. 88, 3689–3698 (2005). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 189.Cross SE et al. Nanomechanical analysis of cells from cancer patients. Nat. Nanotechnol. 2, 780–783 (2007). [DOI] [PubMed] [Google Scholar]
- 190.Soteriou D et al. Rapid single-cell physical phenotyping of mechanically dissociated tissue biopsies. Nat. Biomed. Eng. 7, 1392–1403 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 191.Plodinec M et al. The nanomechanical signature of breast cancer. Nat. Nanotechnol. 7, 757–765 (2012). [DOI] [PubMed] [Google Scholar]
- 192.Kumar S & Weaver VM Mechanics, malignancy, and metastasis: the force journey of a tumor cell. Cancer Metastasis Rev. 28, 113–127 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 193.Gensbittel V et al. Mechanical adaptability of tumor cells in metastasis. Dev. Cell 56, 164–179 (2021). [DOI] [PubMed] [Google Scholar]
- 194.Roberts AB et al. Tumor cell nuclei soften during transendothelial migration. J. Biomech. 121, 110400 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Cao X et al. A chemomechanical model for nuclear morphology and stresses during cell transendothelial migration. Biophys. J. 111, 1541–1552 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 196.Conrad C et al. Mechanical modulation of tumor nodules under flow. IEEE Trans. Biomed. Eng. 69, 294–301 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 197.Wisniewski E et al. Dorsoventral polarity directs cell responses to migration track geometries. Sci. Adv. 6, eaba6505 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 198.Koehl MAR Biomechanical approaches to morphogenesis. Seminars in Developmental Biology 1, 367–378 (1990). [Google Scholar]
- 199.Miller CJ & Davidson LA The interplay between cell signalling and mechanics in developmental processes. Nat. Rev. Genet. 14, 733–744 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 200.Heer NC & Martin AC Tension, contraction and tissue morphogenesis. Development 144, 4249–4260 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 201.Handler C, Scarcelli G & Zhang J Time-lapse mechanical imaging of neural tube closure in live embryo using Brillouin microscopy. Sci. Rep. 13, 263 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 202.Viola JM et al. Tubule jamming in the developing kidney creates cyclical mechanical stresses instructive to in vitro nephron formation. Preprint at bioRxiv 10.1101/2022.06.03.494718 (2022). [DOI] [Google Scholar]
- 203.Ambekar YS et al. Multimodal imaging system combining optical coherence tomography and Brillouin microscopy for neural tube imaging. Opt. Lett. 47, 1347–1350 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 204.Sun W et al. The bioprinting roadmap. Biofabrication 12, 022002 (2020). [DOI] [PubMed] [Google Scholar]
- 205.Vedadghavami A et al. Manufacturing of hydrogel biomaterials with controlled mechanical properties for tissue engineering applications. Acta Biomaterialia 62, 42–63 (2017). [DOI] [PubMed] [Google Scholar]
- 206.Xiang Y et al. Background-free fibre optic Brillouin probe for remote mapping of micromechanics. Biomed. Opt. Express 11, 6687–6698 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 207.Antonacci G, Elsayad K & Polli D On-Chip notch filter on a silicon nitride ring resonator for Brillouin spectroscopy. ACS Photonics 9, 772–777 (2022). [Google Scholar]
- 208.Randleman JB et al. Subclinical keratoconus detection and characterization using motion tracking Brillouin microscopy. Ophthalmology 10.1016/j.ophtha.2023.10.011 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 209.Wu P-H et al. A comparison of methods to assess cell mechanical properties. Nat. Methods 15, 491–498 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 210.Patel A et al. A liquid-to-solid phase transition of the ALS protein FUS accelerated by disease mutation. Cell 162, 1066–1077 (2015). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.